Shock and Vibration........................15 Compliance and Certifications....................16 Safety Compliance Standards..................16 Electromagnetic Compatibility..................16 CE Compliance ......................17 Product Certifications and Declarations................. 17 Environmental Management................... 17 Definitions Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. The following characteristic specifications describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
Noise Figure 1. DC Voltage Noise, Typical 300 V 100 V 100 m 10 V 100 mV 10 m 100 µ 10 µ 1 µ 100 n 10 n 1 μ 10 μ 100 μ 100 n 10 m 100 m Aperture Time (Seconds) With input shorted, Normal DC Noise Rejection, and Auto Zero ON.
Noise Figure 2. PXIe-4080 Resistance Noise, Typical 100 k ≥10 MΩ Range 1 MΩ Range 10 k 100 kΩ Range 10 kΩ Range 1 kΩ Range 100 Ω Range 100 m 10 m 100 µ 1 μ 10 μ 100 μ...
Table 4. AC Voltage Accuracy ± (% of reading + % of range), 2 Years, T ± 10 °C, extcal ± 5 °C selfcal Range Peak 1 Hz to >40 Hz to >20 kHz to >50 kHz >100 kHz (rms) Voltage 40 Hz 20 kHz...
AC Input Corresponding Minimum Maximum Accuracy Voltage Isolated Digitizer Peak-to-Peak Peak-to-Peak Range Range Signal Signal Amplitude Amplitude 50 mV 100 mV 5 mV 200 mV Refer to the PXIe_CLK100 accuracy 500 mV 50 mV of the chassis. 10 V 500 mV 20 V 50 V 100 V...
Page 13
Figure 4. PXIe-4080 Voltage Waveform Noise, Typical 300 V Range 100 V Range 10 V Range 100 m 1 V Range 100 mV Range 10 m 100 µ 10 µ 1 µ 100 n 10 k 100 k 10 M Sample Rate (Samples per Second) With input shorted.
Figure 5. PXIe-4080 Current Waveform Noise, Typical 1 A Range 200 mA Range 20 mA Range 100 μ 10 μ 1 μ 100 n 10 n 10 k 100 k 10 M Sample Rate (Samples per Second) With input open.
HI SENSE 300 VDC or AC LO SENSE 300 VDC or AC When this fuse symbol is marked on a device, take proper precautions. Fuse This icon denotes a warning advising you to take precautions Hazardous Voltage to avoid electrical shock. Timing Mode Trigger Latency...
Compliance and Certifications Electromagnetic interference can adversely affect the measurement Caution accuracy of this product. The input terminals of this device are not protected for electromagnetic interference. As a result, this device may experience reduced measurement accuracy or other temporary performance degradation when connected cables are routed in an environment with radiated or conducted radio frequency electromagnetic interference.
Page 19
For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your patents.txt ni.com/patents software, the file on your media, or the National Instruments Patent Notice at . You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer ni.com/legal/export-compliance...