® TheraFlo AP LAL Power Unit Service Manual Table of Contents General Information Page A. Introduction B. Safety Precautions C. Warnings Maintenance - Level 1 A. Introduction Power Unit A. Part Identification Overview B. General Cleaning Instructions C. Clean / Replace Air Filter D.
Page 3
® TheraFlo AP LAL Power Unit Service Manual Table of Contents Repair - Level 2 A. Introduction B. Tools Required Power Unit A. Part Identification Overview B. Replacing the Top Case C. Replacing the Control PCBA D. Replacing the Blower E.
General Information A. Introduction The TheraFlo AP Power Unit can be matched with the TheraFlo AP 7” M-Set and with the TheraFlo 9” M- Set. This Carilex Service Manual provides repair and maintenance instructions for the TheraFlo AP Low Air Loss Power Unit.
Risk of Electric Shock: DO NOT open back cover. This device is NOT user serviceable. This device should only be opened by qualified personnel approved by Carilex. Refer all service to your local Carilex authorized dealer. Oxygen Equipment: Explosion risk if used in the presence of flammable anesthetics.
Page 6
® TheraFlo AP LAL Power Unit Service Manual IPX0: Don’t immerse power unit in any liquid or spray any liquids directly on the power unit. Electrical: Do not insert items into any opening of the power unit. This could short internal components, which could cause fire or electrical shock.
® TheraFlo AP LAL Power Unit Service Manual Maintenance - Level 1 A. Introduction The procedures described in the Maintenance - Level 1 section of this manual can be done by anyone on staff carefully following the exact instructions. Document #:M-1-S2002-QP01 Revision: 1.1...
® TheraFlo AP LAL Power Unit Service Manual General Cleaning Instructions ROUTINE CLEANING WHILE IN USE BY A SINGLE PATIENT: Always unplug power unit before cleaning. Power unit can be cleaned by wiping down with damp cloth using soap and water or mild neutral detergent. Never spray cleaners or liquids directly on power unit.
® TheraFlo AP LAL Power Unit Service Manual Clean/Replace Air Filter 1. Remove the filter from rear chassis. 2. Check for excess dirt or dust. The filter should be cleaned with mild soap and water or, if necessary, replaced with a Genuine replacement filter. 3.
® TheraFlo AP LAL Power Unit Service Manual Control Panel Sticker Replacement 1. To Remove Control Panel Sticker The control panel sticker is held in place by a self-adhesive backing. Use a small, thin, flat bladed screwdriver to pry up an edge and gently remove control panel sticker. Control Panel Sticker 2.
® TheraFlo AP LAL Power Unit Service Manual Feet Bumper Replacement 1. To Remove Feet Bumpers The feet bumpers are held in place by a self-adhesive backing. Use a small, flat bladed screwdriver to pry up an edge and gently remove it. Feet Bumper 2.
® TheraFlo AP LAL Power Unit Service Manual Fuse or Fuse Cap Replacement 1. Push in and turn counter clockwise one quarter turn then the fuse cap will pop out. You can remove the fuse. 2. To replace, insert fuse to fuse cap, push in and turn one quarter clockwise. 3.
® TheraFlo AP LAL Power Unit Service Manual Hanging Hook Replacement 1. Remove the 8x screws as shown. (4X screws for each hanger hook). 2. Reverse procedure to install new hanger hook Hook (Right) Hook (Left) Document #:M-1-S2002-QP01 Revision: 1.1...
® TheraFlo AP LAL Power Unit Service Manual Testing - Level 1 A. Introduction The procedures described in the Testing - Level 1 section of this manual can be done by anyone on staff carefully following the exact instructions. B. Tools Required Manometer.........................
Repair - Level 2 section of this manual. If you are not authorized to do repairs please contact Carilex or your local distributor to receive a return authorization number and the details of where you can send the unit for servicing.
Be sure to put a new Do Not Open sticker in place and fill out a Product Service Record Form (found at the end of this manual) and fax/mail to Carilex. FPC connector is rather delicate and fragile. Be sure to handle carefully, always insert the connection of FPC connector perpendicular to the PCBA.
TheraFlo LAL Power Unit Service Manual C. Replacing the Control PCBA 1. Follow the steps in “Replacing the Top Case” to remove the top case. 2. Take out the screws that hold the control PCBA in place. Remove the PCBA and replace it with a new one.
TheraFlo LAL Power Unit Service Manual D. Replacing the Blower 1. Follow the steps in “Replacing the Top Case” to remove the top case. Cut the cable ties that bind all the wires together. 3. Loosen and remove the wire plug. 4.
TheraFlo LAL Power Unit Service Manual E. Replacing the Power PCBA 1. Follow the steps in “Replacing the Top Case” to remove the top case. Remove 10X connectors from the PCBA. Unscrew the 6x screws that hold the board in place. Remove the board and replace it with a new one.
TheraFlo LAL Power Unit Service Manual F. Replacing the Transformer 1. Follow the steps in “Replacing the Top Case” to remove the top case. 2. Cut the cable ties that bind all the wires together. 3. Loosen and remove the 2X transformer wire connectors from the power PCBA. 4.
TheraFlo LAL Power Unit Service Manual G. Replacing the Battery 1. Follow the steps in “Replacing the Top Case” to remove the top case. 2. Follow the steps in “Replacing the blower” to remove the blower. Cut the cable ties that fixed the battery on battery PCB. Disconnect the 2-pin battery wire from the power PCBA and remove the battery pack.
TheraFlo LAL Power Unit Service Manual H. Replacing the Valve 1. Follow the steps in “Replacing the Top Case” to remove the top case. 2. Refer to section F to remove the blower. 3. Remove FPC connector from power PCBA. 4.
TheraFlo LAL Power Unit Service Manual I. Replacing the Bottom Case Follow the steps in “Replacing the Hanging Hooks” to remove the Hanging Hooks from the bottom case. Follow the steps in “Replacing the Top Case” to remove the top case. Follow the steps in “Replacing the Blower”...
® TheraFlo AP LAL Power Unit Service Manual Testing Function Test FUNCTIONAL TEST Pass Step Directions Plug AC power, chose 7” or 9”, confirm with Power Button, verify orange standby indicator is blink after 2 minutes (valve fix position) the indicator is lit. Press Power Button, Verify green ON indicator is lit and blower spins up (air is flowing out connectors).
Troubleshooting can be carried out by any trained maintenance personnel. If your system cannot be repaired using the instructions in the Level 1 sections of this manual and you are not a manufacture-authorised service person, please contact Carilex Medical Inc regarding repairs.
® TheraFlo AP LAL Power Unit Service Manual Power Unit Excessive Excessive Airflow from Troubleshooting Lights Airflow Noise Vibration Power Unit from from from Power Stops Power Power Power Unit Intermittently Unit Unit Unit Check that the wall socket is working ...
Page 29
NO.____________ Old: Result: □PASS □NG New: Old: New: For Carilex QA Division use only: 類別: A: 硬體設計類 B:軟體設計類 C:機構設計類 D:製造品質 E:檢驗品質 F: 供應商品質 G:其他 是否建議開立 ECR □No. □ Yes, ECR/ECN# 是否建議開立 Customer Complain □No(Consumables/ User error/ others_________). □Yes(Abnormal function/ Volume of abnormal quality/ others_________), CC NO.____________ For Carilex Sales Div.