Advertisement

Quick Links

 2003 - IMR Environmental Equipment, Inc
May 2003
INTRODUCTION _______________________________________________________________ 3
SAFETY INSTRUCTIONS _______________________________________________________ 3
1
ANALYSER LAYOUT AND FEATURES ________________________________________ 4
1.1
Instrument Features and Keypad ________________________________________________ 4
1.2
Standard Probe Configuration __________________________________________________ 5
1.3
Analyser Connections __________________________________________________________ 6
2.
SAFETY WARNING ________________________________________________________ 7
3.
FIRST TIME USE __________________________________________________________ 7
3.1
Installing Batteries ____________________________________________________________ 7
4.
NORMAL START UP SEQUENCE ____________________________________________ 8
4.1
Every Time You Use the analyser ________________________________________________ 8
4.2
Automatic Calibration _________________________________________________________ 9
4.2.1
Quick Autozero I ___________________________________________________________________ 9
4.2.2
Pressure and Temperature Mode (PTM)_________________________________________________ 10
4.3
Changing the Main Display ____________________________________________________ 11
4.4
Setting Inlet Temperature _____________________________________________________ 12
4.5
Sampling the Flue Gas ________________________________________________________ 13
4.6
Taking a Pressure Reading ____________________________________________________ 14
4.7
Regular Checks During Sampling _______________________________________________ 14
4.8
Normal Shutdown Sequence ___________________________________________________ 15
5.
MOVING THROUGH THE MENU ___________________________________________ 16
5.1
Basic Operation ______________________________________________________________ 17
5.2
Menu Options and Settings ______________________________________________________ 18
5.2.1
Main Menu Options ________________________________________________________________ 18
5.2.2
Set-up menu options ________________________________________________________________ 22
5.3
Setting and Changing Numbers __________________________________________________ 24
6.
PRINTING INFORMATION_________________________________________________ 25
6.1
Printing a 'Live' Test _________________________________________________________ 25
6.2
Standard Printout ____________________________________________________________ 26
7.
STORING AND RETRIEVING DATA _________________________________________ 27
7.1
Storing a 'Live' Test __________________________________________________________ 27
7.2
Viewing and Printing a 'Stored' Test ____________________________________________ 28
7.3
Clearing Logged Data _________________________________________________________ 29
8.
MAINTENANCE __________________________________________________________ 30
8.1
Emptying and Cleaning the In-line Water Trap ___________________________________ 30
8.2
Changing the Particle Filter____________________________________________________ 31
9.
PROBLEM SOLVING ______________________________________________________ 32
10.
ANNUAL RE-CALIBRATION _____________________________________________ 32
IMR 1100- Operator Manual
IMR1100
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 1100 and is the answer not in the manual?

Questions and answers

Summary of Contents for IMR 1100

  • Page 1: Table Of Contents

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 INTRODUCTION _______________________________________________________________ 3 SAFETY INSTRUCTIONS _______________________________________________________ 3 ANALYSER LAYOUT AND FEATURES ________________________________________ 4 Instrument Features and Keypad ________________________________________________ 4 Standard Probe Configuration __________________________________________________ 5 Analyser Connections __________________________________________________________ 6 SAFETY WARNING ________________________________________________________ 7...
  • Page 2 OXYGEN REFERENCE AND CONVERTING TO mg/m ____________________________ 38 ELECTROMAGNETIC COMPATIBILITY (EMC) STATEMENT ____________________ 39 PROTECTION AGAINST ELECTRIC SHOCK ____________________________________ 40 WARRANTY __________________________________________________________________ 41 IMR__________________________________________________________________________ 43 IMR® reserves the right to adopt technical modifications without prior notice. IMR 1100- Operator Manual...
  • Page 3: Introduction

    The dust filter must be cleaned/replaced when dirty. The Condensation trap must be checked and the condensed water removed if necessary. IMR® or an authorized service facility must re-calibrate the IMR® 1100 once a year to ensure the accuracy and performance SAFETY INSTRUCTIONS Please make sure that you read this section carefully for use of your new combustion gas analyzer.
  • Page 4: Analyser Layout And Features

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 ANALYSER LAYOUT AND FEATURES Instrument Features and Keypad Infra-red printer emitter Display Keypad Particle filter Analyser Water trap Connections See section 1.3 ON / OFF MENU Turns analyser ON and OFF...
  • Page 5: Standard Probe Configuration

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Standard Probe Configuration 120 mm 240 mm 3 m neoprene hose Fit probe connector to end of tube using barbed end IMR 1100- Operator Manual...
  • Page 6: Analyser Connections

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Analyser Connections Flue (Tf) Flue temperature Inlet (Ti) Socket Mains Inlet temperature adapter socket socket Pressure inlets P1 & P2 Water trap drain plug Note! Take care when inserting the temperature probes as the pins are polarised. Insert with the smaller pin (+) to the right.
  • Page 7: Safety Warning

    Take time to read this manual fully. Installing Batteries The IMR 1100 is supplied with 4 AA size alkaline batteries. These should be installed into the instrument as shown below and indicated on the back of the unit: See Appendix A - Main Parameters for the Battery Level Indicator.
  • Page 8: Normal Start Up Sequence

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 NORMAL START UP SEQUENCE Every Time You Use the analyser BEFORE SWITCH-ON CHECK THAT: • The particle filter is not dirty • The water trap and probe line are empty of water •...
  • Page 9: Automatic Calibration

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Automatic Calibration During this sequence the analyser pumps fresh air into the Oxygen and CO (if fitted) sensors to allow them to be set to 20.9 % and zero respectively.
  • Page 10: Main Display

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Once the time has reached zero or the sensors are with pre-determined limits the screen shows the selected fuel and analyser setting: MAIN DISPLAY 4.2.2 Pressure and Temperature Mode (PTM) It is possible to take temperature and pressure measurements without entering the ‘gas’...
  • Page 11: Changing The Main Display

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Changing the Main Display The Main Display can be configured to read three parameters at the same time. The text display and top numerical display are changed independently of the other two displays.
  • Page 12: Setting Inlet Temperature

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 The middle and bottom numerical displays are changed in the following pairs: Text display Units Description and CO % and ppm or Oxygen percentage and CO concentration mg/m and ∆T % and °C or °F...
  • Page 13: Sampling The Flue Gas

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Sampling the Flue Gas Once the automatic calibration procedure has been completed and the specific fuel has been selected (See MENU options) the probe can be inserted into the desired sampling point.
  • Page 14: Taking A Pressure Reading

    IMR1100 Taking a Pressure Reading An optional differential pressure sensor is available for the IMR 1100. With this option a flue draft measurement can be made or a differential pressure test made. To take a flue draft reading, connect the standard probe to the pressure sensor inlet P2 and insert the probe in the flue.
  • Page 15: Normal Shutdown Sequence

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Normal Shutdown Sequence DO THIS EVERY TIME YOU USE THE ANALYSER WARNING! THE PROBE WILL BE HOT FROM FLUE GASES Remove the probe from the flue and allow it to cool naturally. Do not immerse the probe in water, as this will be drawn into the analyser and damage the pump and sensors.
  • Page 16: Moving Through The Menu

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 MOVING THROUGH THE MENU The options in the menu system are in the following sequence: • Fuel selection • Pump / data hold feature • Pressure units • Oxygen reference •...
  • Page 17: Basic Operation

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Basic Operation From the MAIN DISPLAY press to access the MENU system. Menu option Menu mode indicator to scroll through the fuel options Press to accept the selection And move to next menu option...
  • Page 18: Menu Options And Settings

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Menu Options and Settings 5.2.1 Main Menu Options DESCRIPTION DISPLAYED FUEL Select the fuel being used by the boiler from list of standard fuels: Natural Gas Propane Butane Light Oil...
  • Page 19  2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 DESCRIPTION DISPLAYED RESET Resets the pressure readings to zero. To use this facility ensure that the pressure ports of the analyser are at atmospheric pressure, then press The display then asks for confirmation,...
  • Page 20: Temperature Units

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 DESCRIPTION DISPLAYED O2 REF Toxic gas measurements can be referenced to defined oxygen levels. Reference values can be set from 0-20% Oxygen Referencing is detailed in Appendix C. The value of the oxygen reference is shown on the top numerical display. The default value is set to 3%.
  • Page 21: Efficiency Calculation

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 DESCRIPTION DISPLAYED GAS UNITS Gas measurement unit can be changed from: Parts per million (ppm) Milligrams per meter cubed (mg/m DESCRIPTION DISPLAYED EFFICIENCY Efficiency can be selected for Gross CALCULATION or Net values.
  • Page 22: Set-Up Menu Options

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 5.2.2 Set-up menu options DESCRIPTION DISPLAYED SET UP Enter a sub menu to change the time, date and language setting. To accept the current set-up press to change the setting press enter the set up sub-menu: As with the main menu system the set up sub-menu system operates in the same way.
  • Page 23  2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 The next five options require a number to be set. See section 5.3 Setting and changing numbers. HOURS Input the time in hours. MINUTES Input the time in minutes. Input the year.
  • Page 24: Setting And Changing Numbers

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Setting and Changing Numbers Once the number changing sequence is initiated from one of the menu options i.e. hours, below the current setting a digit will appear. This digit on the second line of the display indicates that is the number to be changed.
  • Page 25: Printing Information

    (IR) link similar to a TV remote control. The IR emitter is positioned on the top of the IMR 1100 and the bottom of the printer. Ensure they are pointing at each other and within 1 meter, with no obstructions in the way. Data may be lost if transmission is interrupted.
  • Page 26: Standard Printout

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Standard Printout IMR 1100 The standard printout is shown below :- 26 – 06 – 00 14:27 SMOKE : NATURAL GAS 20.9 EFF% (G) ----- NET . ----- FLUE .
  • Page 27: Storing And Retrieving Data

    May 2003 IMR1100 STORING AND RETRIEVING DATA The IMR 1100 can store up to 99 combustion tests. Once stored, the data can be viewed on the display or downloaded to the infra-red printer. Options in the store menu are: • STORE - Allows data to be stored in memory.
  • Page 28: Viewing And Printing A 'Stored' Test

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Viewing and Printing a ‘Stored’ Test To retrieve a stored test from memory access the STORE menu by pressing Select VIEW using To indicate the data on the display is...
  • Page 29: Clearing Logged Data

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Clearing Logged Data To delete the data in stored memory press to enter the STORE MENU (as above) :- to select DELETE Indicates number of logs in memory to be deleted...
  • Page 30: Maintenance

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 MAINTENANCE Emptying and Cleaning the In-line Water Trap The in-line water trap should be checked and emptied on a regular basis. Water vapour will condense in the probe line, which may cause the water trap to fill suddenly if the probe is moved.
  • Page 31: Changing The Particle Filter

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Changing the Particle Filter This is a very important part of the analyser and should be changed regularly. It prevents dust and dirt particles from entering the pump and sensors that will cause damage.
  • Page 32: Problem Solving

    The following is a list of problems that may occur on the instrument through its operating life. If the cause of the fault is not easy to identify then we advise you contact IMR Inc. or an approved Service Agent. Fault symptom Causes •...
  • Page 33: Product Specification

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 11. PRODUCT SPECIFICATION Parameter Resolution Accuracy Range Temp Measurement Flue Temperature + 2°C (5.0°F) 0-600°C ° +0.3% reading 32-1112°F Inlet Temperature +1°C/F 0-100°C 1°C/F ±0.3% reading 32-212°F Temp (Net) + 2°C (5°F) 0-600°C...
  • Page 34: Appendices

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 APPENDICES A. MAIN PARAMETERS The following section gives details of the main display parameters. O2 : Oxygen reading in percentage (%) Tf : Temperature is measured by the flue gas probe in Centigrade or Fahrenheit.
  • Page 35  2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 CO/CO2 : The CO/CO 2 ratio is the ratio of measured CO divided by calculated CO 2 . It gives an indication of: • How good a gas sample the instrument is reading.
  • Page 36: Combustion Efficiency Calculation

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 COMBUSTION EFFICIENCY CALCULATION The efficiency calculation is based upon British Standard BS845. This identifies three sources of loss associated with fuel burning: Losses due to flue gasses: Dry Flue gas loss, Moisture and hydrogen,...
  • Page 37  2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 Calculated data: Tnet = Net Temperature % CO 2 content in flue gas % Dry Flue Gas losses % Wet losses % Unburned carbon loss % Efficiency Tnet = Flue Temperature - Inlet Temperature Dry flue gas loss % = 20.9 x K1 x (Tnet) / K2 x (20.9 - O 2 m)
  • Page 38: Oxygen Reference And Converting To Mg/M

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 OXYGEN REFERENCE AND CONVERTING TO mg/m Oxygen referencing is required by some regulations such as TA-LUFT. If a reference value is selected then the CO gas measurement will be displayed with the symbol (n) attached to the reading.
  • Page 39: Electromagnetic Compatibility (Emc) Statement

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 ELECTROMAGNETIC COMPATIBILITY (EMC) STATEMENT This product has been tested for compliance with the following generic standards: EN 50081-1 EN 50082-1 and is certified to be compliant Electromagnetic Compatibility The European Council Directive 89/336/EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and has an adequate level of immunity to enable it to be operated as intended.
  • Page 40: Protection Against Electric Shock

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 PROTECTION AGAINST ELECTRIC SHOCK (in accordance with EN61010 – 1 : 1993) This instrument is designated as Class III equipment and should only be connected to SELV circuits. The battery charger is designated as:...
  • Page 41: Warranty

    2. The IMR® warranty shall be granted on the factory new unit and shall commence on the date of the delivery of the original IMR® unit to the customer. It shall last for a period of twelve months regardless of the type and the intensity of use and regardless of any change of owner, which may occur during this warranty period.
  • Page 42  2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 d. The IMR® unit has been left in its original design and meets the original IMR® specifications. e. The fault is not due to external influences or use for a purpose other than the one for which it has been designed.
  • Page 43: Imr

     2003 - IMR Environmental Equipment, Inc May 2003 IMR1100 IMR® Environmental Equipment, Inc. 3634 Central Ave. Saint Petersburg, FL 33711 Toll free: 800-746-4467 Phone: 727-328-2818 Fax: 727-328-2826 Internet: www.imrusa.com Email: info@imrusa.com sales@imrusa.com technical_department@imrusa.com IMR 1100 Operator Manual...

Table of Contents