Sunje SPE-100 Instruction Manual

Advertisement

Quick Links

Instruction Manual
SPE-100(New)
It is recommended that only persons who have sufficient knowledge and experience such as system
designers and responsible persons deal with this product after carefully reading the product manual.
The matters of safety
1
* If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
* It may fail to function properly if someone drops or any misuse of the product.
* Please connect to Ground for precise measurement.
* 9V Battery only. (rectangular shaped battery)
Check the package contents
2
▶ Product compositions
The package includes the following products composition.
Body
Body
SPE-100 (New)
SPE-100 (New)
1ea
1ea
Ground Cable
SGC-PE-1-001
1ea
Adaptor
SAD-IC-1-001
1ea
01
Nomenclatures of parts
3
▶ Body
Eng
Analog Output
Power Adapter Jack
Button
Power
Peak / Reset
Record
Alligator Clip
Zero
SAC-PE-1-001
1ea
Banana Jack
SBJ-PE-1-001
1ea
Battery
1ea
Static Voltage Sensor
Back Lit
Graphic Display
Button
Function
Power On / Off
Switching to peak hold mode
This button is not supported in the current version.
Zeroing the Meter
02
▶ Static Voltage Sensor
1. Sensor Cap
The sensor cap is for adjust offset. Please cap securely in place over the end of the sensor.
Please remove sensor cap before measure.
2. Static Voltage Sensor
This is static voltage of charged object measuring sensor.
Please keep this sensor away from any particles.
3. Sensor Connector
The length of sensor cable is 60mm. (86mm with mounting parts)
How to use
4
* SPE-100(New) measures the magnitude and polarity of static charge
on objects or surfaces.
* Take a measure keeping distance 1inch(25.4mm) between the sensor
and the surface. At 2inch it is approximately 60% as sensitive.
* if the surface is a bare insulator (without a ground plane or metal
backing) that is charged, then the insulator will emit an electric field
(but will not have a well-defined voltage).
▶ How to use static voltage measurement instrument
Measuring Surface Voltages
1. To use, first ground the SPE-100(New) with the ground cable(SGC-PE-1-001)
2. With the sensor cap installed on the sensor, press the "Zero" pad.
3. Remove the sensor cap and hold the end of the sensor at a distance of one inch from the
surface being measured. At 1inch, the display shows voltage (+/-2%) of a conductive surface.
▶ How to operate
Power On
On power up, 0.0 will be displayed until the rotating shutter in the sensor reaches its
proper rotation speed, which typically takes no more than four seconds.
If [Probe Error] is displayed, verify that the probe is plugged in and that nothing is
obstructing the shutter's rotation.
To power off the meter, hold down the "Power" pad for one second.
03

Advertisement

loading
Need help?

Need help?

Do you have a question about the SPE-100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Sunje SPE-100

  • Page 1 Power Power On / Off Measuring Surface Voltages 1. To use, first ground the SPE-100(New) with the ground cable(SGC-PE-1-001) Peak / Reset Switching to peak hold mode 2. With the sensor cap installed on the sensor, press the "Zero" pad.
  • Page 2 120.31 20.00 Calibrated Distance SPE-100(New) is factory calibrated for a distance of 1inch (25.4mm) between the end of the sensor and the surface being measured. On conductive surfaces, the reading is approximately inversely proportional to the distance if at <1inch.