Table of Contents SUBJECT PAGE # General Safety Summary Technical Specifications Features Keypad Functions Operating Instructions Power Switch Main Menu Syringe Diamter Entry Volume Entry Flowrate Entry Start the Pump Power Failure Mode Change or Review Volume Setting While Running Change or Review Rate Setting While Running Fast Forward/Fast Reverse Loading the Syringe...
General Safety Summary Please read the following safety precautions to ensure proper use of your syringe pump. To avoid potential hazards and product damage, use this product only as instructed in this manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Page 4
General Safety Summary Observe all Terminal Ratings Review the operating manual to learn the ratings on all connections. Orient Equipment Properly Do not position the equipment such that it is difficult to reach the disconnecting device. Place Product in Proper Environment Review the operating manual for guidelines for proper operating environments.
Cooling Conditions No special considerations Mechanical Stability No special considerations Protective Packaging No special considerations Earth Terminals No External connections required Removable Protective Means Supplier Name KD Scientific Inc. Address 84 October Hill Road, Holliston, MA 01746 Publication 5600-025 REV-D...
Features The Model 310 Plus is an infusion/withdrawal syringe pump designed to hold glass syringes [Hamilton] from 1 to 250 microliters. Setup and pump operation is extremely simple. A menu, displayed on an alphanumeric LCD “prompts” the operator to make the necessary selections using the keypad for choice of features and numerical entries.
Keypad Functions ↙ This key has two functions: Moves the display to the left. Causes the numerical entry to decrease. To change the displayed number by a single unit press and release the key quickly. If the key is pressed longer the number changes with increasing speed.
Operating Instructions POWER SWITCH The power switch is located at the right, rear corner of the pump. When the power is turned on the LCD will display the Volume or Rate setting. Press the select key to return to the main menu. MAIN MENU Pressing select repeatedly will always return the display to the main menu.
Operating Instructions VOLUME ENTRY The volume setting mode is selected from the main menu or is displayed automatically after changing the diameter. Display reads: xx.xx µl (Units displayed in µl) Use the arrow keys to enter the dispense volume required. When the correct dispense volume is displayed save this number with the select key.
Operating Instructions START THE PUMP Press the run/stop key to start the pump. A second press will stop the pump. If the pump is stopped during a dispense, the volume accumulator is not cleared - the dispense is paused. Restarting the pump with a second press of the run/stop key continues the dispense to the target volume.
Operating Instructions CHANGE OR REVIEW VOLUME SETTING WHILE RUNNING While the pump is running it is possible to look at the volume and flowrate settings without interrupting the operation of the pump. If required, the volume and flowrate settings can be changed while the pump continues to run. When the changes are entered the pump immediately changes to the new settings.
Loading the Syringe Raise the spring loaded syringe clamp and rotate away from the syringe barrel. Using the fast forward/fast reverse feature position the pusher block so that the syringe plunger and syringe barrel will fit into the syringe holder and pusher block. Place the syringe barrel in the V of the syringe holder, making sure that the flange of the syringe barrel is positioned between the two upright posts, and lower the syringe plunger knob into the pusher block.
Fuses The fuses are located in the power entry module on the rear panel. The linecord must be first removed to gain access to the fuse holder. FUSES 5x20mm, 250V~ slow blow, 0.1A Voltage Selector If it is necessary to change the input voltage selection, disconnect the line cord from the entry module on the rear panel.
Flowrate/Syringe Diameter Table TABLE 1 Syringe Min µ/min Max µ/min 700 Series 1700 Series Hamilton ID mm Hamilton ID mm 0.5 µl 0.001 0.726 0.103 1.0 µl 0.001 1.46 0.146 2.0 µl 0.01 2.907 0.206 5.0 µl 0.01 7.282 0.343 10.0 µl 0.01 14.56...
Model Numbers Legacy Model KDS 310 Plus Order Code 110 VAC 78-0311 Order Code 220 VAC with CE Mark 78-9311 Notes: KDS 310 Y suffix: Integrated footswitch capability. Publication 5600-025 REV-D...
Limited Warranty KD Scientific Inc. warrants to the first consumer purchaser, for a period of one year from the date of purchase that this unit, when shipped in its original container, will be free from defective workmanship and materials and agrees that it will, at its option, either repair or replace the defective unit.
Declaration of Conformity Manufacturer: KD Scientific Inc. 84 October Hill Rd. Holliston, MA 01746-1388 U.S.A. Phone: (508) 429-6809 We herewith declare that the following product: Product Name: Model 310 Plus Syringe Pump Model Numbers: KD310 Plus To which this declaration relates, is in conformity with the applicable...
Need help?
Do you have a question about the 310 Plus Series and is the answer not in the manual?