KDS Legato 100 Series USE R’ S MANU A L KDS LEGATO 100 SERIES w w w . k d s c i e n t i f i c . c o m Publication 5617-006 Rev 2.0...
Page 2
KDS Legato 100 Series Syringe Pump Series Model No.: KDS Legato 100 Series - Catalog # 7881xx KDS Legato 950 series- Catalog # 7889xx To which this declaration relates, is in conformity with the applicable EC Directives, harmonized standards, and...
Use only the specified line cord for this product and make sure line cord is certified for country of use. The operating voltage range for the KDS Legato 100 Series is 12-30 VDC. The universal power supply operating voltage range is 100-240 VAC, 50-60 Hz.
Warranty KD Scientific warranties this instrument for a period of one year from date of purchase. At its option, KD Scientific will repair or replace the unit if it is found to be defective as to workmanship or materials. This warranty does not extend to damage resulting from misuse, neglect or abuse, normal wear and tear, or accident.
22.6 x 19.05 x 15 cm (9 x 7.5 x 5 in) 22.6 x 19.05 x 15 cm (9 x 7.5 x 5 in) Weight 2.66 kg (5.9 lbs) 2.66 kg (5.9 lbs) 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Product Overview Product Overview The KD Scientific Legato 100 Series is the next For limited laboratory space the Legato 100 Series can generation of syringe pumps. The series offers be placed on its side to reduce the footprint by 55%.
The pump’s alarm configurability includes alarms for near end of run (user selectable), completion of run, power up, keypad clicks and stall detection. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Product Overview LEGATO 100 SERIES TOUCHSCREEN BUTTON REFERENCE The following buttons are part of the KDS Legato 100 Series software: Settings: Access the settings menu used to adjust Page Down: Used to page down in a display list. the following parameters: Audible Alarms, Pusher Block Force, Display Brightness, etc.
Backspace Key Decimal point Accept the value and go back up to the previous screen. Cancel and go back to the previous screen. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Legato 100 Series series is 100 – 240 pump or computer. VAC, 50-60 Hz. 3. Turn on the main power switch. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
NOTE: For infuse/withdraw models, make sure the plunger is secured in the retaining bracket and the barrel of the syringe is secured in the retaining bracket. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Configuring the KDS Legato 100 Series Configuring the KDS Legato 100 Series The KDS Legato 100 Series program is easy to set up. Select Mode Select Syringe Manufacturer Size The Legato 100 Series is easily configured. When the unit is powered up, the Configuration Screen General will be displayed.
Press to enter the total change the syringe volume units (µl, CAUTION: BE SURE TO LOWER THE FORCE VALUE WHEN USING GLASS SYRINGES. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
3. In the Units Selection Display select the volume and the time base. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Press to select target time Press to select the maximum volume for the syringe size selected 4. In the Units Selection Display select the volume units. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
The pump has basic parameters that should be configured during initial start-up. Once these parameters are entered they probably won’t be changed. To enter the set-up screen press 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Stall – Motor stalls Near End of Run – When the volume dispensed is reached for the specified % Mute All – No audible sounds 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Set-up screen. Cancel the changes made by pressing and go back to the Set-up screen. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Pump Password screen. Enter your password and press the “ACCEPT” button. ✐ NOTE: If you forget your password contact KD Scientific Technical Support at 508-429-6809. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Set-up screen. Cancel the changes made by pressing and go back to the Set-up screen. Press here for custom display brightness level 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Press Here and go back to the Set-up screen. Press here to select Baud Rate 508.429.6809 508.429.6809 508.893.0160 508.893.0160 phone phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com info@kdscientific.com www.kdscientific.com www.kdscientific.com e-mail e-mail Publication 5617-006-REV-2.0...
Set-up screen. Cancel the changes made by pressing and go back to the Set-up screen. Press here to reset the maintenance counter 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Total volume Press to go to Fast forward Press to delivered Set-up screen Start/Stop Fast reverse the pump Press to go to Configuration Screen 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
✐ Press Here NOTE: The fast forward and fast reverse keys are not active (grayed out) when the pump is running. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Step in the Program, but will not review the Program in its entirety to verify that it is logical and can be fully executed 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Step Definition Screen showing the list of Steps currently defined for the Program 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
New Program Press View programs up a page View the programs down a page Custom Program Views up the program Views down the program 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Page 34
Programs can be edited and saved under a new name, allowing them to be simply edited to fit another application. Steps can be copied, moved, inserted or deleted. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Operating the KDS Legato 100 Series Setting A Delay You can set the KDS Legato 100 Series to run after a delay. This delayed run function allows you to set up all the parameters for operation, then have the pump start unattended after a specified amount of time.
Operating the KDS Legato 100 Series Creating a Custom New Program If none of the Programs stored on the KDS Legato 100, 111 or 180 Series match your requirements, you can create a new Program and store the Program for use.
Page 37
(e.g. Constant Rate, Ramp, etc.) you will be taken to the Step Definition screen for the selected Program. 9. Complete the parameter entry applicable to the 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Step options and to properly use the Advanced Options commands. The information that follows is organized into two sections: • Pump Operation Profiles • Advanced Options (Commands and I/O) 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Operating the KDS Legato 100 Series PUMP OPERATION PROFILES The KDS Legato 100 Series Existing Programs and all User-Defined Programs use Pump Operation Profiles as part of the Programs controlling the pump. These profiles are combined with a syringe selection and any Commands or Input/Output triggers to create a complete Program.
‘ENTR’ of time over which the ramp will be calculated 6. Repeat Steps 3 to 5 for the End Rate by selecting During a Ramp Rate profile, the KDS Legato 100 Series End Rate button and continuing data entry. varies the flow rate linearly between the starting rate and ending rate over the full time period entered.
Repeat command is set to repeat Step 2 one time, the system will loop back to Step 2 and execute both Steps 2 and 3 one additional time. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Program Steps Menu screen. 4. Choose Advanced Options to enter the Set Advanced Options screen. 5. Choose Stop to insert the Stop Command into the Program. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
(High or Low). You may then use the trigger command to change the level as desired for your Program. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
• User Press: Sets the pump to wait for manual input from the user at the touchscreen (screen touch). 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Accept to save your changes and return to the File Options screen. 6. Choose ‘BACK’ to return to the Mode Selection screen. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
File Options screen and display a confirmation that the Program was copied. 6. Choose to return to the Mode ‘BACK’ Selection screen. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Program Select from the Program Main screen on the destination pump, then browse the available Programs to verify that the exported Program is shown in the menu. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Program is now saved at the pump. ✐ NOTE: The KD Scientific Syringe Pump Communicator program, provided on the Legato CD, can be used to simplify the import / export process for programmable pumps. Please refer to the Syringe Pump Communicator manual for instructions.
4. Set up the Port Settings as shown below and click OK. This section of the KDS Legato 100 Series manual describes the control of the pump using an external computer device. If communicating via the USB Port, see Appendix C for one time installation instructions of USB Driver.
<lf>[##]<prompt> is the response unless otherwise noted. Pump Chain Commands KDS Legato 100 commands are entered via the USB Error messages port with a terminal program such as HyperTerm, or by Error messages are displayed if the entered command using a controlling computer.
Query response: header. <lf>[##:]#%<cr> <lf>[##]<prompt> The last line shows the number of files and the total number of steps being used on the disk. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Page 54
Typing ‘syrm ?’ displays a list of manufacturers with their associated 3-letter code. Typing ‘syrm {code} ?’ shows a list of syringe sizes associated with the specified manufacturer. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
• Flag six is the target reached status. It will be “T” if the target time or volume was reached and “.” if not. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
ERROR 217- Motor has stalled. ERROR 218- Program needs a rate. ERROR 219- Program needs infuse and withdraw rates and a target volume. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Page 60
NOTICE- Copying a program. NOTICE- Verify rates and target. NOTICE- Withdrawal complete, press to clear alarm. NOTICE- Infusion complete, press to clear alarm. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
26.594 mm 85.1297 nl/min 88.4040 ml/min * Not available on Legato 101 and 111. These pumps work with 0.5 µl to 10 ml syringes 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Page 63
3 ml 8.585 mm 3.99042 nl/min 4.14394 ml/min 5 ml 11.989 mm 7.78230 nl/min 8.08163 ml/min 10 ml 14.427 mm 11.2692 nl/min 11.7027 ml/min 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
- Data + Data USB VIRTUAL COMMPORT DRIVER INSTALLATION When you connect the KDS Legato 100 Series pump to a computer via USB for the first time, Windows will seek to install a driver for communication. The following section details the installation of the Virtual CommPort Driver supplied with the KDS Legato 100 Series Pump.
Page 66
5. If a warning message regarding the driver publisher is displayed, choose Install this driver software anyway. 6. Click Close after the software has been successfully installed. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
“downstream” ✐ NOTE: Each pump in the daisy-chain must have a unique address. The first pump must be set to address 00. 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...
Appendices APPENDIX D: MAINTENANCE Maintenance KDS Legato 100 Series pumps require no special To clean the exterior surfaces, use a lint-free cloth to maintenance other than keeping them clean by avoiding remove loose dust. Use care to avoid scratching the clear accidental spills of pumped material.
Appendices Upgrading Legato 100 Series Software ✐ NOTE: Prior to upgrading, users will need to install the KDS Legato 100 Series USB driver as well as the bootloader driver file (KDS Legato 100 Series-Bootloader.inf). Reference “USB Virtual Comport Driver Installation”.
Additionally, make sure that the guide rods and lead screw are properly lubricated. ✐ The KDS Legato 100 Series Pump will notify you of a power failure with a NOTE: It is recommended that you save message when power is restored your programs periodically by uploading them to a PC.
Adapter D Sub 15 to Terminal Block 780225 Footswitch with Phono Jack Plug 788314 Adapter for 25 ml and 50 ml Hamilton Gastight Syringe 788315 Hex Key 508.429.6809 508.893.0160 phone KD Scientific Legato 100 Series Syringe Pump User’s Manual info@kdscientific.com www.kdscientific.com e-mail Publication 5617-006-REV-2.0...