Sustainability Monitor User Manual Version 1.0 July 25, 2023 PO Box 7900, Wheaton, IL 60187 | 1-800-95-SPRAY www.spray.com www.sprayscan.com Spray.Analysis@spray.com The information contained in this user manual is confidential and may not be reproduced in whole or in part, in digital or hardcopy format, without the express written consent of Spraying Systems Company.
Page 2
SprayScan® mSM gives a visual readout and warns the user of any deviation in spray performance. With an installed database of all Spraying Systems Co. and TeeJet nozzles, parameters can be set to monitor specific nozzle performance. These signals can alert users of clogged or worn nozzles, or more serious downstream issues, both in real-time and historically.
Sustainability Monitor | User Manual v1.0 1 Introduction 1.1 Components The SprayScan® mSM Sustainability Monitor comes with the components shown in Figure 1, which includes the monitor and sensor modules, signal cable, universal power supply, and region-specific power cord. The monitor and sensor modules come with mounting brackets. The sensor module comes standard with liquid connection adapters with 3/8”...
Sustainability Monitor | User Manual v1.0 • Spraying Systems Co.® Sustainability Monitoring application 1.4 Sensor Module Specifications The sensor module has the following features: • Measurement ranges o Pressure 0 to 150 psi (0 to 10 bar) gauge pressure o Flow rate 0.1 to 5 GPM (0.4 to 20 LPM)
Sustainability Monitor | User Manual v1.0 When connecting the sensor module in the liquid feed line, take care to use a wrench on the flats of the liquid connections so that torque will not be applied directly to the sensor module case. Use thread sealant to prevent leaks at this connection.
Sustainability Monitor | User Manual v1.0 Figure 5: SprayScan® mSM Wiring Connections 2.4 Basic Operation & Home Screen Navigation At first startup, the sustainability monitor application defaults to a basic operation mode displaying real- time pressure, flow rate, and temperature measurements on the home screen. Touch any measurements to see historical measurements.
Sustainability Monitor | User Manual v1.0 3 Application Navigation The SprayScan® mSM home screen includes 4 options at the bottom of the screen. 3.1 Information Screen Figure 7: Displays the home screen and location of the Information Screen Icon. This screen tells the operator vital information regarding the device: •...
Sustainability Monitor | User Manual v1.0 Connecting to the cloud allows connection to the server, allowing the instrument to be remotely monitored in real time via our SprayScan Smart Monitoring software. This function is covered in more detail in Section 6 of the following pages. 3.3 Network Connection Figure 9: Displays the home screen and location of the Wi-Fi Connection Icon.
Sustainability Monitor | User Manual v1.0 4 Settings 4.1 Navigating the Settings Menu The Settings menu can be accessed by touching the gear icon on the home screen and entering pin code 7272. Figure 10: Settings menu navigation. 4.2 Settings Menu Details Here you’ll see a variety of options: 1.
Sustainability Monitor | User Manual v1.0 a. Adjusts the sample rate for the historical measurements. Additionally, sets the telemetry rate for the measurement transmission to the Smart Monitoring service (optional service). Default is 5 s. 6. Total Flow Reset a. Resets flow totalizer to 0. 7.
Sustainability Monitor | User Manual v1.0 Figure 11: settings defining Accessing Target Operation operation parameters. 5.2 Targeted Performance Operation After setting the Target Performance parameters, the home screen will change to show large, colorful visual status symbols. Green check mark means all measurements are with the 1 level alert threshold of the target value.
Sustainability Monitor | User Manual v1.0 Figure 13: Target vs Actual measurements (left) and historical flow measurements with flow total (right). 6 Smart Monitoring More to come soon! 7 Errors & Troubleshooting 7.1 Error Message at Startup When powering on the monitor, a notice may be displayed that says, “System Error: Tap for details”. There are several different kinds of errors which one may encounter.
Sustainability Monitor | User Manual v1.0 Sensors Not Connected The ADC is not receiving signals from the sensor module. Perform the following steps. • Check signal cable connection on monitor module. • Check signal cable connection on sensor module. • Power cycle the system.
Sustainability Monitor | User Manual v1.0 8 Acknowledgments The SprayScan® mSM by Spraying Systems Co.® is built on the Moddable SDK, which incorporates the following software: Moddable SDK Runtime Licensor: Moddable Tech, Inc. License: Commercial Link: https://github.com/Moddable-OpenSource/moddable/blob/public/licenses/readme.md Licensor: Marvell International Ltd and Kinoma, Inc.