Page 1
OWNER’S MANUAL Model: F600B BARIATRIC FULL BODY PATIENT LIFT Span-America Medical Systems, Inc. Greenville, SC 29606 (800) 888-6752 Ensure the product has been assembled according to the instructions in this manual. All operators should read and understand the instructions for safe and proper operation of the patient lift.
Date of Purchase: the operator. That is due to it being a full electric lift. The F600B Date of Manufacture can lift the patient with ease from the fl oor to the bed. The 6 point...
Caregiver refers to the operator or person who is assisting with the transfer. A Span-America patient lift will provide years of service if it is properly maintained as any electric and/or mechanical equipment requires. Please Symbols used in this manual and on the product and their meanings: pay careful attention to the following important information regarding the care, maintenance, and operation of the patient lift.
• DO NOT replace any components of the lift without consulting with * F600B model shown Span-America and must follow proper instruction from Span when replacing any components. • DO NOT service any parts of the lift while in use with a patient.
Lift Weight Unloaded (kg) 1610.4 mm 389.7 (15.3) 126.0 (5.0) 26.2 (1.0) * Span-America is committed to continuous improvements of our products therefore the specifi cation, dimensions, and features listed above are for guidance only and are subject to change without prior notice.
Page 6
F600B Installation Step 4: To stand the mast upright, lift the mast upward 1” to clear the metal plate at the bottom of the mast. Step 1: Remove folded lift from box. Metal Plate Step 2: Remove the bottom plastic caps, then the bottom bolt, nut and washers on the side of the base bracket.
Page 7
Step 7: Install the round plastic cover and plastic washers to the boom. Step 10: Secure the upper end of the actuator to the bracket on the boom using a pin and retaining ring as follows. Step 8: Install the boom to the mast and align the holes. Then secure it by reinstalling the bolt, washer and nut.
Page 8
Power Control Unit Installation Operating Instructions Double check all assemblies for tightness and read operating instructions carefully prior to use. M5 SCREWS For optimum performance the lift should be transported and stored in following condition range: • -25°C to +5°C (-13°F to 41°F), and •...
Page 9
Power Control Unit Overview Release Pressed in Emergency Stop Button Emergency Stop Button Up / Down Button Up / Down Button Battery Pack Battery Pack Release Handle Release Handle LCD Display Panel LCD Display Panel LCD Display Panel Signs When the emergency stop button is released, the LCD Display Panel will show one of the four signs below.
Page 10
Operating Lift: Using Hand Control Warning! - Battery Low and Charging is Needed If the battery needs to be charged, the LCD LED Indicator Display Panel will show a blinking low battery Lifting * Green = In Use sign shown on the left either when the * Blank = Standby Lowering emergency stop button is released or when a...
Page 11
M (M920280) XL (M920940) L (M920281) XXL (M920941) 6 Point S-Bar S (M920942) XL (M920945) M (M920943) XXL (M920946) Disposable universal with head support (Span-America) Nonwoven, Disposable L (M920944) Hook up A Point 1 Hook up B Point 2 Sani Polyester...
2. Position user onto his/her side by rolling user towards you. • The operator of the lift shall inspect the Span lift before each use. Check all 3. Roll the sling in half. The handle on the back section sling should face outward bolts for tightness.
• REPLACE a sling when it show signs of deterioration. Deterioration of sling • Contact Span-America for “Guideline for Identifying Deteriorated Slings” if needed. Disinfecting of the lift and sling • Inspect slings prior to each use for contamination from previous use.
Check to make sure the actuator plug is properly connected and inserted all the way into the control box Patient lifts may not be returned unless the wrong lift is shipped in error by Span or the lift is heavily damaged or defective out of •...