INTEGRA Biosciences COLDPLATE Installation Instructions

INTEGRA Biosciences COLDPLATE Installation Instructions

Advertisement

Quick Links

Integration of COLDPLATE and BIOSHAKE modules
Installation Instructions V00
Document Change History
Vers.
Status:
V00
Released 18.12.2023
Table of Contents
1
Introduction --------------------------------------------------------------------------------------------------------- 1
2
Scope of delivery -------------------------------------------------------------------------------------------------- 1
3
Initial setup of COLDPLATE and BIOSHAKE modules ---------------------------------------------- 2
3.1
Connect the power cable to the module -------------------------------------------------------------- 2
3.2
Install the ANSI/SLAS footprint adapter -------------------------------------------------------------- 2
4
Integration on ASSIST PLUS pipetting robot ----------------------------------------------------------- 2
4.1
Place the modules onto the ASSIST PLUS deck -------------------------------------------------- 2
4.2
Connect the module to the ASSIST PLUS ----------------------------------------------------------- 3
5
Integration on VIAFLO 96 or VIAFLO 384 ---------------------------------------------------------------- 3
5.1
Place the modules onto VIALFO 96 or VIAFLO 384 ---------------------------------------------- 3
6
Connect the module to the PC for controlling ---------------------------------------------------------- 4
6.1
Define module parameters via the PC ---------------------------------------------------------------- 4
7
Imprint----------------------------------------------------------------------------------------------------------------- 4
1
Introduction
This document describes integration of heating, cooling and shaking modules into the ASSIST PLUS
pipetting robot or VIAFLO 96 and VIAFLO 384.
The following modules are supported:
COLDPLATE on-deck module for cooling and heating, integration ready (part no. 4950)
BIOSHAKE 3000 on-deck module for shaking, integration ready (part no. 4951)
BIOSHAKE 3000-T on-deck module for heating and shaking, integration ready (part no. 4952)
Before starting with the installation, please carefully read the Operating Manual of the corresponding
module.
2
Scope of delivery
All modules are supplied with the following parts required for integration:
COLDPLATE (pre-assembled caps), BIOSHAKE 3000 or BIOSHAKE 3000-T (pre-assembled
clamping mechanism) with external power supply, power cable, RS-232 cable, certificate, and
operating manual
ANSI/SLAS footprint adapter kit (part no. 128152)
RS232-MINI DIN 8 cable
RS232-USB-A converter (Digitus)
Cable clips
Pedestal (50 mm+)
This document (Part no. 137890)
Please, verify the scope of delivery when unpacking the modules and check for potential
transportation damage.
137890_V00_Integration_COLDPLATE_BIOSHAKE.docx
Date:
Change:
Document issued.
1/4
18. December 2023

Advertisement

loading
Need help?

Need help?

Do you have a question about the COLDPLATE and is the answer not in the manual?

Questions and answers

Summary of Contents for INTEGRA Biosciences COLDPLATE

  • Page 1 This document describes integration of heating, cooling and shaking modules into the ASSIST PLUS pipetting robot or VIAFLO 96 and VIAFLO 384. The following modules are supported: COLDPLATE on-deck module for cooling and heating, integration ready (part no. 4950)  BIOSHAKE 3000 on-deck module for shaking, integration ready (part no. 4951) ...
  • Page 2 To avoid any labware collision with the pipette the pedestal can be optionally placed on position B or C on the 3-position universal deck of the ASSIST PLUS to raise the labware height. Compatibility table for placing of COLDPLATE and BIOSHAKE on ASSIST PLUS deck Position A...
  • Page 3 Connect the module to the ASSIST PLUS Connect the RS-232 communication cable with the short RS232-MINI DIN 8 adapter cable. Insert the MINI-DIN 8 poles plug into the socket labeled with AUX on the left side of the ASSIST PLUS. Integration on VIAFLO 96 or VIAFLO 384 Place the modules onto VIALFO 96 or VIAFLO 384 Place the module onto the plate holder position A, B...
  • Page 4 Imprint © 2023 INTEGRA Biosciences AG All rights relating to this documentation reserved. INTEGRA Biosciences AG especially reserves the rights of duplication, processing, translation and display format. Any reproduction of parts or the entire documentation, or storage thereof on electronic media, or processing or dissemination thereof, requires the prior written consent of INTEGRA Biosciences AG.