Kofloc EX-201SC Instruction Manual

Thermal digital mass flow controller, rs-485 communications
Table of Contents

Advertisement

Quick Links

MJ400321A1
Thermal Digital Mass Flow Controller
EX-201SC
RS-485 Communications
Instruction Manual
0

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EX-201SC and is the answer not in the manual?

Questions and answers

Summary of Contents for Kofloc EX-201SC

  • Page 1 MJ400321A1 Thermal Digital Mass Flow Controller EX-201SC RS-485 Communications Instruction Manual...
  • Page 2: Table Of Contents

    MJ400321A1 Table of Contents Page 1.Foreword・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2 2.Wire Connection・・・・・・・・・・・・・・・・・・・・・・・・ 2 3.Basic Specifications・・・・・・・・・・・・・・・・・・・・・・・ 4 4.Message Structure・・・・・・・・・・・・・・・・・・・・・・・ 5 5.Details of Commands・・・・・・・・・・・・・・・・・・・・・・ 7 1) RMFS:Acquisition of the maximum full scale flow rate [significand]・・・・・ 2) RDPP:Acquisition of the flow rate decimal point position [number of decimal places]・・・・・・・・・ 3) RFRU:Acquisition of the flow rate unit・・・・・・・・・・・・・・・...
  • Page 3: 1.Foreword

    MJ400321A1 1. Foreword This document describes the specifications and handling of the RS-485 communications function installed as a standard feature in EX-201S. The wiring, installation and operating procedures, other than communications, are presented in a separate instruction manual. Prior to use, please read it also. 2.
  • Page 4 MJ400321A1 Valve open/close input (analog signal) Fully close (0V) Control (OPEN) Fully open (+5~24V) Controlled by flow rate setting Fully close Fully open voltage/current inputs (analog signals) Flow rate setting method “Analog (1)” Valve open/close input (analog signal) Fully close (0V) Control (OPEN) Fully open (+5~24V) Fully close...
  • Page 5: 3.Basic Specifications

    MJ400321A1 3. Basic Specifications Synchronization Start-stop ※ Transmission speed 9,600 bps Start bit 1 bit Data length 8 bits Stop bit 1 bit ※ Parity None Transmission system 3-wire half-duplex Communication – control circuit: Uninsulated Insulation Communication – power supply: Uninsulated By use of rotary switch SW1 and 2 on top of Communication ID setting equipment...
  • Page 6: 4.Message Structure

    MJ400321A1 4. Message Structure ・Command message The command message from the communication controller to the equipment should be structured as shown in the following example: Communication ID Command Data Check sum This equipment recognizes “STX” as the head of the message unconditionally.
  • Page 7 MJ400321A1 ・Response message When the communication ID that has been set to the equipment matches the communication ID specified by the command message, the equipment returns a response message as shown in the following example. Communication ID Command Exit code Data Check sum “%(25H)”...
  • Page 8: 5.Details Of Commands

    5. Details of Commands The commands that can be executed for the equipment will be described below. Please note that KOFLOC will bear no responsibility for motions resulting from the use of commands other than those specified here. 1)RMFS:Acquisition of the maximum full scale flow rate [significand] The maximum full scale flow rate [significand] is acquired.
  • Page 9: Rfrc:acquisition Of The Flow Rate Reference Temperature Condition

    MJ400321A1 4)RCFS:Acquisition of the full scale flow rate [significand] The full scale flow rate [significand] is acquired. The full scale flow rate [significand] refers to the significand of the full scale flow rate currently set to the equipment and is used to calculate the full scale flow rate currently set to the equipment together with the flow rate decimal point position [number of decimal places] and flow rate unit.
  • Page 10: Rpgt:acquisition Of The Calibration Gas Type

    MJ400321A1 8)RPGT:Acquisition of the calibration gas type The calibration gas type used for equipment calibration is acquired. Data Command: None Response: Decimal 1 digit (fixed length 1 byte) 1:N (Nitrogen) 9)RCGT:Acquisition of the gas type The gas type that has been set is acquired. Data Command: None Response: Decimal 1 digit (fixed length 1 byte)
  • Page 11: Rccf:acquisition Of The Cf Value

    MJ400321A1 11)RCCF:Acquisition of the CF value The CF value that has been set is acquired. The CF value is used for the CF conversion mode. This is expressed always based on N (nitrogen) (1000) regardless of the calibration gas types. Data Command: None Response: Decimal 4 digits (fixed length 4 bytes)
  • Page 12: Rcfr:acquisition Of The Instantaneous Flow Rate [Significand]

    MJ400321A1 15)RCFR:Acquisition of the instantaneous flow rate [significand] The instantaneous flow rate [significand] is acquired. This is used to calculate the instantaneous flow rate together with the flow rate decimal point position [number of decimal places] and flow rate unit. Data Command: None Response: Decimal 4 digits (fixed length 4 bytes)
  • Page 13: Rcvs:acquisition Of The Valve Status

    MJ400321A1 20)RCVS:Acquisition of the valve status The present valve status is acquired. Data Command: None Response: Decimal 1 digit (fixed length 1 byte) 0:Fully opened, 1:Controlled, 2:Fully closed, 3:50%opened 21)RSFD:Acquisition of the set flow rate [significand] (digital) The set flow rate [significand] (digital) that has been set is acquired. This is used to calculate the set flow rate (digital) together with the flow rate decimal point position [number of decimal places] and flow rate unit.
  • Page 14: Ralm:acquisition Of The Alarm Occurrence Status

    MJ400321A1 24)RALM:Acquisition of the alarm occurrence status The alarm occurrence status is acquired. Data Command: None Response: Decimal 1 digit (fixed length 1 byte) 0:No alarm (normal) 1:Alarm present (Sensor error) 2:Alarm present (valve overheat) 4:Alarm present (set value storage circuit error) When several alarms are present, their values are added.
  • Page 15: Wlfd:setting Of The 0 - 0.5% Range Flow Rate Indication

    MJ400321A1 28)WLFD:Setting of the 0 – 0.5% range flow rate indication Set the 0 – 0.5% range flow rate indication. Data Command: Decimal 1 digit (fixed length 1 byte) 0:0 – 0.5% range flow rate indication used 1:0 – 0.5% range flow rate indication not used (Forced indication of “0”) Response: None 29)RAZS:Acquisition of the auto zero...
  • Page 16: 6.Flow Rate Expression・・・・・・・・・・・・・・・・・・・・・・15

    MJ400321A1 6. Flow Rate Expression The maximum full scale flow rate, full scale flow rate, instantaneous flow rate, set flow rate (digital) and set flow rate are expressed by a combination of the significand and flow rate decimal point position [number of decimal places] and flow rate unit. Note that flow rate decimal point position [number of decimal places] and flow rate unit are used commonly for each flow rate and cannot be changed.
  • Page 17 MJ400321A1 URL:http:// www.kofloc.co.jp...

Table of Contents