Summary of Contents for SANI Membranes Vibro-I Series
Page 1
Please read all safety instructions carefully. This user manual is part of the product. Keep it in a safe place for future reference. Replacement manuals can be downloaded from our webpage at www.sanimembranes.com SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Vibro-I is fully drainable resulting in minimal product loss The Vibro-I can conform to GMP/FDA/EC regulations for and faster CIP cycles. materials in contact with food and other sanitary standards on request. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
This symbol denotes a possible danger with a low risk that moderate or minor injury may result if it is not avoided. ATTENTION This symbol denotes a danger with low risk of damage to property if not avoided. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
As an option, these sizes can also be supplied on wheels. The Vibro-I 20, 40 & 80 m can be supplied with an optional moveable base, suited for pallet lifters. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Page 7
The Vibro-I Cartridge can be directly lifted onto the Vibro-I drive frame by placing lifting eyes on top of two of the rods. When positioned on the motor assembly, the cartridge is mounted on the base plate using four bolts. SANI Membranes A/S www.sanimembranes.com...
Page 8
(4, 5). The ports are for the feed and retentate connections (5), and for possible vent and/or connection to instruments such as pressure transducer or temperature sensor (4). SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Page 9
Do not use the Vibro-I 2.5 & 5 m² Cartridges without the adaptor, as this could lead to an unstable system WARNING SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Page 10
Vibro-I units. Each manifold is then connected by flexible hoses to the 4 Cartridges. Installation details are available upon request. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
The Vibro-I is intended to filter media and can only be used with Vibro-I Cartridges from SANI Membranes. The Vibro-I must only be used together with a feed system with a built-in safety to protect the Vibro-I against over-pressure. If the feed system is able to deliver over-pressure, a CE approved safety valve set to the applicable maximum operating pressure of the Vibro-I system must be included before the Vibro-I inlet.
Personal safety equipment should always be worn when applicable (e.g. safety googles, safety gloves etc.). WARNING If the fluid system is leaking, liquid spill to the floor can cause a slipping hazard. CAUTION SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Device failure ATTENTION • Only use accessories, consumables and spare parts that are in technically perfect condition. The use of accessories, consumables and spare parts not approved by SANI Membranes is the sole responsibility of the operator. SANI Membranes A/S www.sanimembranes.com...
The Vibro-I Drive for 2.5, 5 & 10 m² units are supplied on wheels as standard, and can alternatively be supplied on machine feet. SANI Membranes A/S www.sanimembranes.com...
Place the cartridge onto the Vibro-I 2.5 & 5 Adaptor (align the nuts to the groves of the plate). Secure the adapter with 4 x M12 x 40mm screws. Tighten with Approx. 65 Nm. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Place the Vibro-I Cartridge onto the Vibro-I Drive Secure the Cartridge bottom plate with 4 x M16 x 50 mm bolts and M16 nuts (add washers on each side) Tighten the bolt with 100-120 Nm SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Repeat steps 1 – 3 for the next three cartridges – ensuring to orient them correctly. Install one cartridge at the time, and make sure it is securely fastened before moving on to the next. WARNING SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
10. Connect the top of each permeate line to the manifold using 10 mm tubing. Figure 2 Figure 3 SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
/ or preservative solutions that should be removed before first use. SANI Membranes recommend to flush the membrane for 30 minutes with clean hot water (50-55 °C). For the first 5-10 minutes lead both the retentate and the permeate streams to drain without any recirculation. Establish a positive transmembrane pressure to force liquid through the membrane’s porous layer.
TMP) for direct comparison. Although it is membrane and application dependent, a recovery of 80% or more of the initial water flux should be expected from a suitable cleaning regime. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Reduces fouling and enables high protein • transmission. Permeate This approach can be advantageous with other • configurations too. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
CIP system, degree of automation, temperature control system etc. Permeate Feed Retentate 80 m single pass 20 m mi 2 20 m single pass SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Further tightening or repeated tightening can result in reduced tension in the internal bolts in the top- and bottom assemblies which will cause these to come loose when the unit is vibrating. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Page 28
The four bolts mounting the cartridges to the drives must be re- tightened with 100 – 120 Nm on a regular basis. Preferably, once a quarter. Remember to tighten bolts in a cross pattern. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Page 29
Membrane fouling will take place more rapidly and result will be a drop in the flux. Please refer to previous section for trouble shooting of this problem. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Page 30
Flush the unit with 55 °C clean water without recirculating the permeate for 30 minutes to remove the protective glycerin layer from the new module. CIP clean the Vibro-I unit and restart the production. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Make sure the HP1 modules do not dry out while the cartridge is disassembled. For longer duration the HP1 modules may be stored in a suitable storage solution (see section 7.4 below). SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Vibro-I after a re-assembly. NOTE: The permeate outlets on the HP1 modules are secured with lock-nuts and assembled with spring loaded washers. These should not be disassembled or tightened without expert knowledge. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
Some examples of storage solutions are shown in the table. For long term storage it is recommended to replace the storage solution at regular intervals, for instance every 3-6 months. SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
0 - 4 bar(g) at 5 – 35 °C, 0 - 3 bar(g) at 5 – 55 °C and 0 - 1 bar(g) at up to 80 °C Vibration Motor 400 VAC, 50 Hz, 900 W SANI Membranes A/S www.sanimembranes.com Ryttermarken 8 | DK-3520 Farum | Denmark...
400 VAC, 50 Hz, 3,200 W 9. Conformity SANI Membranes are committed to develop and supply products that meet relevant regulatory standards and requirements set by governing bodies. For further compliance or safety information, please contact SANI Membranes customer support.