geko™ T-3 / WEBT3EN-R3 June 2023 / gekodevices.com
Instructions for Use
The geko™ device is a neuromuscular electrostimulation medical device and its intended use is:
• To increase blood circulation,
• For the prevention of venous thrombosis,
• For the prevention and treatment of oedema.
The device works by increasing blood circulation by stimulating the common peroneal nerve.
It is to be used on a single patient, in hospital, clinic and home environments. Apply to one or both legs as advised by your Healthcare Professional (e.g. Nurse
or Doctor). Patients should be trained in the positioning and operation of the device.
The geko device is intended for up to 24 hours continuous use, until it is no longer needed. It may be removed for short periods of time when washing or having
various tests (see warnings and precautions). Each device is single use and should be replaced every 24 hours, the device has an additional 6 hours run time.
Read this document fully before use and use only as instructed. Improper use of this device can be dangerous. Do not switch on unless
correctly attached to the patient.
If assistance is required or to report difficulties, unexpected operation or events, contact the manufacturer.
Visit the website gekodevices.com for further advice and usage tips.
Summary of Contents for Firstkind REBOOT geko PRO T-3
Page 1
geko™ T-3 / WEBT3EN-R3 June 2023 / gekodevices.com Instructions for Use The geko™ device is a neuromuscular electrostimulation medical device and its intended use is: • To increase blood circulation, • For the prevention of venous thrombosis, • For the prevention and treatment of oedema. The device works by increasing blood circulation by stimulating the common peroneal nerve.
Page 2
Fitting: Remove the film from geko device and place Cleaning: It should be applied to clean, dry skin. If there the marker line over the fibula head (round lump). is too much hair in the area it should be removed using Attach the short end round the front of the leg and the trimmers or clippers.
Page 3
How to find the Fibula Head & Alternative Positions geko™ T-3 / WEBT3EN-R3 June 2023 / gekodevices.com Location: There are three possible locations that will give a successful stimulation of the common peroneal nerve Fibula head: The fibula head A can (dorsiflexion).
Side Effects Skin Inflammation or Irritation: Do not apply over or near the following: In some cases, skin inflammation or irritation can develop in • sore, infected or inflamed areas, the contact area: either remove the device or re-attach in the •...
geko™ T-3 / WEBT3EN-R3 June 2023 / gekodevices.com Storage and operation When storing the geko device keep it in the protective foil pouch. It can be stored in a temperature range of -25°C to 40°C, but we recommend storing the device up to 30°C. Classification The device is internally powered by a non-replaceable CR2032 lithium ion coin cell battery.
Page 6
CH Authorised Representative MedEnvoy Switzerland, Unique device identifier Gotthardstrasse 28, 6302 Zug, Switzerland. Storage and transportation humidity range whilst within packaging Firstkind Limited Manufactured by Hawk House, Peregrine Business Park, See Instructions for Use Gomm Road, High Wycombe, HP13 7DL, United Kingdom,...