Page 1
Operating instructions Water sampler MAXX SP 5 - Zone 2 II 3G EEX nC/R/ic IIB T3 Gc SP5 Zone 2 May 2015...
Page 2
R e m a r k MAXX GmbH is not liable for possible faults in this documentation. Any liability for direct or consequential damages in connection with the delivery or the use of this device is excluded as far as it is legal.
Should you require further information or should particular problems occur which are not described in a sufficient manner in this manual, help can be requested through your local MAXX office or representative. We have examined the document for compliance with the described hardware and software.
Remark is an important information on the product itself, the handling of the product or the respective part of the manual to which particular attention should be paid. Manual SP 5 Zone 2 Page 5...
Equipment category 3 Equipment group Application areas: Equipment category Potentially explosive mixtures of air and gases (G) Category 1 Zone 0, 1 or 2 Category 2 Zone 1 or 2 Category 3 Zone 2 Page 6 Manual SP 5 Zone 2...
Remark: It is not allowed to dispose of the instrument using municipal waste disposal services. Please talk to your contact person at MAXX GmbH if you have to dispose of a unit. Manual SP 5 Zone 2 Page 7...
The tightness of the housing should be regular tested. The test interval must be set by the operator. Usually it is between 1 to 3 years. Page 8 Manual SP 5 Zone 2...
Installation Characteristics The device consists of the following component parts: Protective top (can be opened) Upper part of housing Back lit LC display Keyboard Flow-restricting protective enclosure Lower part of housing – cooled sample compartment Heating (explosion- proof) Distributor with pneumatic drive Bottles Fixing bar...
The signal inputs (e.g. for flow meter) are located on the „fold-out“ plate and can be connected according to the plan of terminal connections. Danger By connecting external signals the explosion protection can be put at risk. If necessary, use separation barriers! Page 12 Manual SP 5 Zone 2...
Signal inputs Installation Troubleshooting If a fault develops in your sampler, we can only provide prompt assistance if you give us the type and serial number of the apparatus concerned. You will find these data on the type plate. The more precise your description of the fault is, the better our fault diagnosis will be. Symptom Possible cause Action...
Bottle tray for 12 x 2,9 L bottles 0060034 2,9 L PE bottle 0060035 Cap for PE bottle 0900569 Distributor Version with glass bottles: 0030013 2,0 L glass bottle 0060161 Cap for glass bottle Page 14 Manual SP 5 Zone 2...
Installation For all fixed site samplers: Order number Designation 0030004 350 ml Duran 50 glass replacement metering vessel 0900012 Suction hose with screw connection, length 5 metres 0069304 Suction hose, running metre 0050025 V2A stainless steel sinker weight, 180 mm long 0900014 Extraction unit 0030009...
Connections resp. relays are in the terminal box Sample inlet: Hose connection at top of metering unit. Union nut R 3/4", hose ID 9 mm Sample outlet: Hose connection at metering unit. Silicone hose 12x2; is only slid on Page 16 Manual SP 5 Zone 2...
Installation Data sheet Data sheet Housing Double-walled stainless steel (material 1.4301) with 40 mm insulation and 2 lockable doors. Upper door with plexiglass window. Protective heating in upper part. Protective top which can be opened for connection and maintenance works. Self-contained Automatic cooling and heating of the sample compartment at +4°...
Operation Sampling modes Graphical representation of sampling modes TIME-FLOW-EVENT Flow graph Time- A sample is extracted in proportional fixed time intervals (e.g. every 10 minutes) with a fixed sample volume (e.g. 50 ml). A sample is extracted in variable time intervals Flow- (depending on the flow (Q)) dependent...
Page 20
If an event is longer than the programmed bottle filling time, two or more bottles will be filled for this event depending on the programmed bottle filling time. Page 20 Manual SP 5 Zone 2...
Operation Vacuum sampling system Vacuum metering vessel with pinch-valve - pneumatic pinch-valve which closes the dosing hose below the metering vessel. - no sensor, thus almost no wear. - easy disassembly of metering vessel for cleaning. - the sample volume can be adjusted between 20 and 350 ml by displacing the filling tube.
Page 22
Diaphragm pump (manufacturer: KNF-Neuberger / order No. 0091671) Please see the attached operating instructions from KNF-Neuberger. Solenoid valves (manufacturer: Bürkert / order No. 0091645) Please see the attached operating instructions from Bürkert. Page 22 Manual SP 5 Zone 2...
Operation Spare parts list – vacuum sampling system Article No. Designation Component 0030004 Metering vessel 350 ml Dosing system 0030005 Metering vessel 500 ml (only standard Dosing system vacuum) 0060457 Black air filter (pneumatic unit) Dosing system 0069301 Silicone hose 12 x 2 Dosing system 0069401 O-ring 16 x 4, NBR...
Symptom Possible cause Action Chapter No function Solenoid valve is not Check solenoid Distributor activated valve Pneumatic line defective Check pneumatic line Wrong positioning of Wrong distributor type Check Distributor distributor selected programming Page 24 Manual SP 5 Zone 2...
Page 25
Programming instructions Software Version 5 Manual SP5 Zone 2 Page 25...
Terminate sleep mode (only portable Press for at least 5 Back key sec. samplers) Keep pressed when RESET / reset to factory settings Back key switching on (NOTE: all settings and data will be deleted!) Page 26 Manual SP 5 Zone 2...
NAVIGATION The sampler can be operated by means of the control unit. With the ARROW KEYS, the ENTER KEY and the BACK KEY you can move from one screen to another. An arrow on the display shows that there are further selection possibilities (see illustration).
Page 28
Selection of individual menu points In the selection menus, additional program settings are displayed. All the settings which have been activated by pressing the ENTER key or which are already active are marked with a symbol. Page 28 Manual SP 5 Zone 2...
MAIN MENU STRUCTURE Description of the main menu structure with submenu levels 2 and 3 PROGRAMS STATUS/STOP INFO PAUSE STOP START IMMEDIATELY DATE/TIME WEEKDAY/TIME CHANGE TIME FLOW DIGITAL FLOW ANALOGUE EVENT TIME EVENT DIGITAL EVENT ANALOG VARIABLE TIME EVENT BATCH SAMPLING MANUAL IN CURRENT BOTTLE SAMPLING...
Page 30
DATA MEMORY SAMPLING DATA TEMPERATURE DATA MEASURING DATA BOTTLE PROTOCOL BOTTLE ARCHIV Page 30 Manual SP 5 Zone 2...
Page 31
• PUMP DIAGNOSIS/ COMPONENTTEST • PINCH VALVE TEST • VALVE SYSTEM • DISTRIBUTOR • DIGITAL OUTPUTS OUTPUTS TEMPCARD LOWER HEATING COOLING UPPER HEATING DIGITAL INPUTS FLOW DIGITAL; EVENT DI3 DI4 DI5 DI6 DI7 DI8 (only if I/O board is installed) DI 9-12 ANALOG INPUTS ANALOG 1:...
Page 32
SLEEP MODE • ACTIVE (only portable devices) • INACTIVE PASSWORD • CHANGE PASSWORD • CHANGE PROGRAMS • CHANGE SETTINGS • PROGRAM STOP SERVICE Setting of base parameters (only for service technicians) Password protected Page 32 Manual SP 5 Zone 2...
Description of the displays with explanation DISPLAY DISPLAY EXPLANATION / FUNCTION PROGRAMS • INFO Display of program details Interruption of the running program (max. 120 • PAUSE min) STATUS/STOP • STOP Stop the current program or all programs IMMEDIATELY • Program start can be: •...
Page 34
DISPLAY OF VERSION • No. of startvalues • Software version of datalogger • Software version of Temp. Board Display of IP address (only if WEB-Board is installed) Default IP: 192.168.1.1 IP ADDRESS Default PORT: 47234 Page 34 Manual SP 5 Zone 2...
Page 35
SETTINGS DATE/TIME Setting of date/time • LANGUAGE Setting of the language • DISTRIBUTOR Selection of distributor type Setting of the maximum suction time (0-600 sec.) • MAX. SUCTION TIME Pre-purge = purge of suction hose • PRE-PURGE of the sample extraction (0 - 99,99 sec.) PRIOR Post-purge = active purging of the metering vessel •...
Page 36
- A password for changing programs can be entered • STOP PROGRAMS - A password for stopping programs can be entered Setting of base parameters (only to be done by a SERVICE service technician) (Password protected) Page 36 Manual SP 5 Zone 2...
Examples of programming Programming of a time-proportional sampling program Select PROGRAMS in the main menu Select CHANGE Select PROGRAM NO. 1 (out of 12). Programs No. 2-12 can be selected by pressing the left or right arrow key. Press ENTER to edit the program. Selection of the sampling mode TIME (Sampling is effected in fixed time intervals) Set the sampling interval(time interval between the single...
Page 38
Besides the standard programming there are several special program functions which can be activated selectively with each operating mode. In the menu „MORE SETTINGS“, you can find a list with all special functions available. Please find following a detailed description. Page 38 Manual SP 5 Zone 2...
SPECIAL PROGRAM FUNCTIONS Beside the standard programming features the unit also offers the following special functions: PROGRAMS CHANGE TIME/FLOW/EVENT MORE SETTINGS SPECIAL FUNCTIONS If "MORE SETTINGS" has been chosen, the following special functions are available depending on the single operating modes: PROGRAMMING OK When all the desired settings have been entered, and this function is selected all settings are...
Page 40
(e.g.in case of an exceedance of a limit value) andcan be activated or deactivated. For the event program the sampling interval and the bottle filling time have to be defined in hours and minutes (hh:mm). Page 40 Manual SP 5 Zone 2...
Page 41
Program run: As soon as there is an event signal, the distributor moves to the next empty bottle (is recorded in the memory as event bottle). The sample extraction is effected according to the set values as long as the signal is present. If the signal is active longer than the bottle filling time set, further bottles are filled.
Page 42
Remark: If the function „ABSOLUTE START TIME“ is activated /deactivated, also the first programmable input (PIN 40 to X5) is automaticallyactivated/deactivated! Page 42 Manual SP 5 Zone 2...
Page 43
BOTTLE FILLING PAUSE (SÜV program ) Setting range: between 0-10080 min. (7 days). This function enables a deferred sample extraction (filling pause) related to the bottles before the next bottle is filled. The pause is entered in minutes. Example: Sampler with 24-bottle distributor system. Bottle fill time: 2 hours.
(according to our example it would be bottle 2) and automatically changes to this position at the first requested sample extraction! WEEKDAY/TIME The program start is effected on the selected weekday and at the selected time in the format: day; hh:mm. Page 44 Manual SP 5 Zone 2...
Page 45
• PROGRAM ENDOPTIONS After having defined the start conditions, the program end can be set as follows: AFTER 1 RUN Program is terminated after 1 run. AFTER X RUNS Program is terminated after X runs. CONTINUOUS OPERATION The program is repeated indefinitely. DATE/TIME The program can be terminated at a certain date/time.
Page 46
STOP An active program can be stopped/aborted. If several programs are active all these programs can be stopped at the same time. Page 46 Manual SP 5 Zone 2...
FLOW-PROPORTIONAL SAMPLE EXTRACTION According to the output signal of your flow meter either the operating mode flow analog or flow digital can be selected in the program settings. FLOW ANALOG – set resp. calibrate - Under „SETTINGS“ ->„SAMPLER SETTINGS ->ANALOGSIGNAL“ the analog input can be set to the default values 4-20 or 0-20 mA or it can be adjusted/calibrated to the plant’s signal.
(the pump is running). During the bottle filling time of 2 hours there can thus be several pump cycles which activate the sample extraction, however, are only considered as 1 event. Page 48 Manual SP 5 Zone 2...
FREELY PROGRAMMABLE INPUTS: PROGR. INPUT ADDITIONAL INPUT 1 ADDITIONAL INPUT 2 ADDITIONAL INPUT 3 There is 1 programmable input in the basic version. With the option "I/O extension board" further 3 inputs are available. Each input can be programmed individually according to the following list: •...
MESSAGES – description - Description Text / meaning When program is started a contact is PROGRAM ACTIVE activated for the whole duration of the program PROGRAM TERMINATED Contact at program end Contact in case of an error ERROR ACTIVE Contact at each sample extraction SAMPLING ACTIVE Contact at each bottle change BOTTLE CHANGE...
Page 52
Date/time of a bottle change BOTTLE CHANGE Date/time of a sample extraction triggered by a SAMPLE EXTRACTION program BTLE CHANGE Bottle change triggered via a programmable REQUESTED (REMOTE) input (only possible if no program is active) Page 52 Manual SP 5 Zone 2...
Page 53
Date/time of the start of a power failure VOLTAGE LOSS START Date/time of the start of an event EVENT START Date/time ofthe end of an event END OF EVENT Date/time of the end of the sleep mode, only END OF SLEEP MODE possible at portable samplers - 1stvalue: CV of pair of electrodes 1 when detecting water...
Page 54
Values are logged when the rain weather pulse divider is activated or deactivated. Selection is RWA-DWA STATUS made per weekday from 00:00 – 23:59. (Only available at the Limburg sampler version) Page 54 Manual SP 5 Zone 2...