Installation Manual Cable Junction Box PR 6130/08 Translation of the Original Installation Manual 9499 053 00800 Edition 4.1.0 01/24/2024 Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany Phone: +49.40.67960.303 Fax: +49.40.67960.383...
Page 2
Any information in this document is subject to change without notice and does not represent a commitment on the part of Minebea Intec unless legally prescribed. This product should only be operated/installed by trained and qualified personnel. In correspondence concerning this product, the type, name, and release number/serial number as well as all license numbers relating to the product have to be cited.
Cable Junction Box PR 6130/08 Table of contents Table of contents Introduction............................3 Read the manual................................. 3 This is what operating instructions look like ......................3 This is what lists look like............................3 This is what menu items and softkeys look like..................... 3 This is what the safety instructions look like......................
1 Introduction Cable Junction Box PR 6130/08 Introduction Read the manual Please read this manual carefully and completely before using the product. This manual is part of the product. Keep it in a safe and easily accessible location. This is what operating instructions look like 1.
Cable Junction Box PR 6130/08 1 Introduction NOTICE Warning of damage to property and/or the environment. NOTICE indicates that damage to property and/or the environment may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Note: User tips, useful information, and notes.
2 Safety instructions Cable Junction Box PR 6130/08 Safety instructions General notes CAUTION Warning of personal injury. The product was in perfect condition with regard to safety features when it left the factory. To maintain this condition and to ensure safe operation, the user must follow the instructions and observe the warnings in this manual.
General information Repairs are subject to inspection and must be carried out at Minebea Intec. In case of defect or malfunction, please contact your local Minebea Intec dealer or service center for repair. When returning the device for repair, please include a precise and complete description of the problem.
Cable Junction Box PR 6130/08 3 Specifications Technical Data Protection classes in compliance with IEC 529 or DIN EN 60529 IP66: Dust-proof and leak-proof against strong water jets. Installation position Cable entry from below Quantity of load cells max. 8...
Cable Junction Box PR 6130/08 4 Installation and connection information 4 Installation and connection information General information Only use the cable glands provided by the manufacturer. Install the junction box so that the cable glands are on the bottom. Do not open the junction box when connected to the voltage supply.
Page 13
4 Installation and connection information Cable Junction Box PR 6130/08 SIGNAL - gray Terminal contacts - measuring voltage (load cells) Terminal contact - measuring voltage (connection cable) SUPPLY + Terminal contacts + supply voltage (load cells) Terminal contact + supply voltage (connection cable) Terminal contact + sense (connection cable –...
Cable Junction Box PR 6130/08 4 Installation and connection information Cable gland The cable glands each have one sealing ring. The M20 cable gland can be fitted with a reduction sealing ring if cables ≤8 mm are used (see Chapter 4.5).
4 Installation and connection information Cable Junction Box PR 6130/08 Note: It is essential that the following tightening torques are observed: M20: 4 Nm M25: 6 Nm 3. Screw the cap nut (1) onto the cable gland and tighten slightly.
Cable Junction Box PR 6130/08 5 Cable connections Cable connections General information Cable entry must be from below. Fit the wires and screen of the connection cable with wire end ferrules as per DIN 46228-1: 4× 0.75 mm (red, blue, white, black) 2×...
5 Cable connections Cable Junction Box PR 6130/08 Equipotential bonding conductor NOTICE It is especially important that the ground is connected correctly to the components and the cable junction box. You must also ground the device separately and ensure the power supply is properly shielded against the effects of lightning.
Repairs Repairs are subject to inspection and must be carried out at Minebea Intec. In case of defect or malfunction, please contact your local Minebea Intec dealer or service center for repair. When returning the device for repair, please include a precise and complete description of the problem.
Minebea Intec at a charge. This option of returning the product is intended to provide proper recycling or reuse in a manner that is collected separately from municipal waste.