Important safety information Important safety information Please read the operating instructions and maintenance instruc- tions carefully and completely prior to carrying out maintenance work at Testomat instruments. Observe the warning notices in these maintenance instructions and the operating instructions of the respective instrument. ...
General instructions Further documents Testomat instruments are plant components. Therefore, always ob- serve the documentation of the plant manufacturer. General instructions Regular maintenance is necessary to ensure trouble-free operation of the Testomat instrument. Regular visual inspections also increase operational reliability. Also refer to the notes in the operating instruc- tions! Fix errors by yourself.
Prior to dismantling, always write down a ® description of the error (failure effect). For Testomat 2000 ® Testomat ECO please download the respective checklist from our website www.heylanalysis.de.
General instructions Prior to carrying out maintenance work Carry out a visual inspection of the instrument: • Are the instrument doors closed properly? • Is the instrument heavily soiled? • Is there air inside the dosing hoses? • Are the hose connections of the dosing pump free of leaks? •...
Carrying out maintenance Carrying out maintenance Cleaning the measuring chamber and the sight- glass windows Proceed as follows: Switch off the instrument or press the "STANDBY" key. If required, remove any water from the measuring chamber: ➔ SERVICE I ➔ MANUAL OPERATION ➔ Drain chamber ...
Close the manually-operated valve of the supply line to Testomat ® 2000 ® Depressurise the lines of Testomat 2000 via the function: ➔ SERVICE ➔ MANUAL OPERATION ➔ Flush chamber Switch off the instrument and loosen the hose connections at the filter housing.
(Overflow detection Push locking (Required tool: Adjustment pin 2 mm, Art. no. 30990) ring upwards Testomat 2000 ® / ECO instruments have been factory set. Therefore, instrument adjustment during start-up is not required. All instruments from instrument number 222.899 and higher are equipped with a self-adjusting electronic circuit.
Carrying out maintenance Subsequently turn the potentiometer anti-clockwise by an approx. 1/8 turn. Check correct functioning of the overflow detection circuit via the key M = input valve: The "Analysis" LED must illuminate while wa- ter is overflowing. Press key M again for quitting. The LED must extinguish once the overflow has finished.
Carrying out maintenance Installing the interface cards SK910, UK910, ® RS910 (only Testomat 2000 Switch off the instrument: Open the upper housing cover: Insert the plug-in circuit board into the left-hand slot with the com- ponent side on the left (contact no.1 is at the top).
Carrying out maintenance Position and function of the fuses Description Error during failure Rate Art. No. Fuses ® Testomat 2000 Instrument not functioning T0,2A (115V) 31594 Instrument not functioning T0,1A (230V 31595 T1A (24V) 31592 T0,16A (230V) 31622 "Ff. 24V failure" dosing...
Page 13
Carrying out maintenance Description Error during failure Rate Art. No. Fuses ® Testomat ECO Instrument not functioning T0,16A (230V) 31622 T0,315A (115V) 31585 Instrument not functioning 31592 Instrument not functioning GS-M 5x20V 31582 In case of overload or short circuiting at the relay outputs, fuse F9 is triggered (if the power for the external consumers is supplied from terminals "l"...
Carrying out maintenance Repairing or replacing the dosing pump Maintenance rate NOTE ® The dosing pump DOSIClip is a high-precision piston dosing pump, which makes up to 400 000 pump strokes per year depending on the settings (a measurement all 10 minutes x 4 pump strokes every day). To ensure proper operation for many years, we recommend sending in the pump for servicing every 2-3 years.
Special function “Adjust mode” Checking dosing When measuring the hardness, it is checked whether the yellow val- ue is at least 50 below the limit value after each (except the first) dosing stroke. Each stroke usually results in the value being reduced by 80 to 100 digits.
Press the “M” (3) key to open the input valve IN. : the “IN” LED (6) illuminates. ® Only Testomat 2000 The “ANALYSIS” (7) LED must illuminate when the water flows through the overflow borehole of the measuring chamber (If this is not the case, adjust the overflow detection as described above).
Water consumption Water consumption Water consumption of the instruments is composed of the analysis- relevant volume (18.1 ml per analysis), of the measuring water ad- justment (150 – 200 ml per analysis), the water consumption for the internal flushing time 0.5 l (500 ml /minute). and the water consump- tion for flushing after analysis (approx.
Indicator consumption Indicator TH2025, TH2100, TH2250 Number of analysis and range with 500 ml Range in hours Num. of Water hardness (°dH) Analysis Interval (as a result of idle time and flush times/AUX) minutes TH2025 TH2100 TH2250 0,25 8333 1667 4444 8611 12778...
Valve block ® Testomat 2000 starts with an analysis cycle. If you have pro- grammed a flushing time, do not fill the measuring chamber until the output valve has closed (“OUT” LED extinguishes). Fill the measuring chamber with the standard solution until it over- flows.
Troubleshooting Troubleshooting Error message MST analysis A measuring fault analysis occurs when the water in the measuring chamber is too “light” after the second dosing. Check the following points if MST analysis is displayed: 1. The indicator’s expiry date has been exceeded. Remedy: ...
Page 21
Troubleshooting 5. The water pressure is insufficient. Water is supplied, but the water level inside the measuring chamber remains too high after closing the inlet valve. Remedy: The water level should be approx. 32 mm above the lower edge of the transparent measuring chamber block.
Water level correct Water level Switching on the real-time clock ® (only Testomat 2000 too low If the integrated clock no longer works (e.g. due to EMC faults), you can restart it. Switch off the instrument, ...
Troubleshooting ® Error message FST optics (Testomat 2000 Remedy The plug-in circuit board driver/receiver SE-T2000 (art. no. 40091) is defective. Replace the plug-in circuit board. There is an error at the optical component. The receiver is defec- tive. Replace the measuring chamber holder.
Troubleshooting Readjusting the current interface A signal of less than 20mA is output at the current interface. Remedy: Contact your customer service representative.
Spare parts lists ® Articles marked with an * can only be used for Testomat 2000 ® articles marked with an ** can only be used for Testomat 2000 ECO Unmarked components can be used for both instruments. Art. no Pressure controller Art.
® Check List Testomat 2000 Dear customers and service technicians, This check list cannot replace your expertise or extensive experience in fault resolution. It is intended to support fast and systematic error diagnosis and error documentation. This list does not claim to be complete. We are therefore always grateful for any advice and information you may be able to provide.
Page 32
Instrument settings Caution! Your settings may be deleted if repairs are carried out. Therefore, note down your instrument settings in the table below before sending the instrument to our service team for repairs. Please enclose a copy of the table with the instrument. If you have noted down the settings, they can be easily re-entered by your service staff once any repairs have been completed.
Page 33
Instrument settings Caution! Your settings may be deleted if repairs are carried out. Therefore, note down your instrument settings in the table below before sending the instrument to our service team for repairs. Please enclose a copy of the table with the instrument. If you have noted down the settings, they can be easily re-entered by your service staff once any repairs have been completed.