Download Print this page

M-system M2RS1 Instruction Manual

Rtd transmitter

Advertisement

Quick Links

INSTRUCTION MANUAL
RTD TRANSMITTER
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System's Sales Office or representatives.
PACKAGE INCLUDES:
Signal conditioner (body + base socket) ...................... (1)
MODEL NO.
Confirm that the model number described on the product is
exactly what you ordered.
INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
CONFORMITY WITH EC DIRECTIVES
• This equipment is suitable for Pollution Degree 2 and In-
stallation Category II. Reinforced insulation (signal input
or output to power input: 300V) and basic insulation (sig-
nal input to output: 300V) are maintained. Prior to instal-
lation, check that the insulation class of this unit satisfies
the system requirements.
• The equipment must be mounted inside a suitable ire
enclosure.
• Altitude up to 2000 meters
• The equipment must be installed such that appropriate
clearance and creepage distances are maintained to con-
form to CE requirements. Failure to observe these re-
quirements may invalidate the CE conformance.
POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
100 – 240V rating: 85 – 264V, 47 – 66 Hz, approx. 3 – 5VA
24V DC rating: 24V ±10%, approx. 3W
11 – 27V DC rating: 11 – 27V, approx. 3W
110V DC rating: 85 – 150V, approx. 3W
GENERAL PRECAUTIONS
• Before you remove the unit from its base socket or mount
it, turn off the power supply and input signal for safety.
ENVIRONMENT
• Indoor use
• When heavy dust or metal particles are present in the air,
install the unit inside proper housing with sufficient ven-
tilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
• Be sure that the ventilation slits are not covered with ca-
bles, etc.
(1 mA sensing current)
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
MODEL
WIRING
• Do not install cables (power supply, input and output)
close to noise sources (relay drive cable, high frequency
line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.
• With voltage output, do not leave the output terminals
shortcircuited for a long time. The unit is designed to en-
dure it without breakdown, however, it may shorten ap-
propriate life duration.
COMPONENT IDENTIFICATION
Body
Fixing Screw
Zero Adj.
Span Adj.
INSTALLATION
Loosen the ixing screw at the front of the unit in order to
separate the body from the base socket.
DIN RAIL MOUNTING
Set the base socket so that
its DIN rail adaptor is at
the bottom.
Position the
upper hook at the rear side
of base socket on the DIN
rail and push in the lower.
When removing the socket,
push down the DIN rail
adaptor utilizing a minus
screwdriver and pull.
WALL MOUNTING
Refer to "EXTERNAL DI-
MENSIONS."
M2RS1
M2RS1
Base Socket
Connection
Diagram
Specifications
DIN Rail
35mm wide
Spring Loaded
DIN Rail Adaptor
EM-5127
P. 1 / 2

Advertisement

loading

Summary of Contents for M-system M2RS1

  • Page 1 BEFORE USE ..• Do not install cables (power supply, input and output) close to noise sources (relay drive cable, high frequency Thank you for choosing M-System. Before use, please check line, etc.). contents of the package you received as outlined below.
  • Page 2 21.5 (.85) 10 (.39) 84 (3.31) 15 (.59) 114 (4.49) [4 (.16)] 23 (.91) • When mounting, no extra space is needed between units. 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN EM-5127 P. 2 / 2 Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp...