HP 8481H/82H
Operation
Operation
Environment
The operating environment for the Power Sensor should be as
follows:
Temperature
0° to 55° C
Relative humidity
< 95%
Altitude
< 4,572 metres (15,000 feet).
Operating Precautions
Before the Power Sensor is connected, the following precautions must
be observed.
Cautions
BEFORE CONNECTING THE POWER SENSOR TO ANOTHER
V
INSTRUMENT, ensure that the instrument and power meter are
connected to the protective earth ground.
Exceeding the energy and power levels shown in Table 1 may result
in damage to the power meter system.
To avoid internal damage to the Power Sensor, connect or disconnect
by turning the Type N RF connector only. Do not turn the Power
Sensor's plastic body.
The absolute maximum RF signal level that may be coupled to the
Power Sensor is:
Maximum Average Power
3.5W
Maximum Peak Power
100W
Maximum Energy Per Pulse
lOOW/^s.
Power Meter Calibrations
The procedure for calibrating one power meter may be different for
another power meter. Follow the calibration directions given in your
power meter manual.