Peracetic Acid SAM I-2020 0 to 5.00 PPM (mg/Liter)
Page 2
To Set Zero Displays and Troubleshooting 1. Press the ON/OFF key. E01: Light absorption too great (dirty optics) 2. The display will show “PAA”. E020 or E021: Too much light reaching detector 3. Insert the ZERO ampoule (supplied in Vacu-vials kit), flat end first, into the ®1 E022 or Battery Icon: Battery should be replaced...
Page 3
Menu Selection Setting Date and Time Upon initial start-up, the SAM will display “Set”, “dAtE”, and “YYYY”, then a 4 digit number. Proceed to Step 4 in the procedure below to set the date and time, or power the instrument off and on again to bypass this process. At any time that the time and/or data need to be reset, follow steps 1-6 of the procedure below.