PCB Piezotronics EX357E91 Installation And Operating Manual
Charge output accel, with uht-12tm shear sensing crystal, 5 pc/g, +1200 f operation, int hardline cable with smaller 10-32 coax, with intrinsic safety approvals
Page 1
Charge output accel, with UHT-12TM shear sensing crystal, 5 pC/g, +1200 F operation, int hardline cable with smaller 10-32 coax, with intrinsic safety approvals Installation and Operating Manual For assistance with the operation of this product, contact the PCB Piezotronics, Inc. Toll-free: 716-684-0001 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com...
Page 2
Assistance is needed to safely operate equipment PCB Piezotronics is an ISO-9001 certified company whose Damage is visible or suspected calibration services are accredited by A2LA to ISO/IEC Equipment fails or malfunctions 17025, with full traceability to SI through N.I.S.T.
Page 3
CAUTION Refers to hazards that could damage the instrument. NOTE Indicates tips, recommendations and important information. The notes simplify processes and contain additional information on particular operating steps. The following symbols may be found on the equipment described in this manual: This symbol on the unit indicates that high voltage may be present.
Page 5
Component Name Hazardous Substances Lead (Pb) Mercury (Hg) Cadmium (Cd) Chromium VI Polybrominated Polybrominated Compounds Biphenyls (PBB) Diphenyl Ethers (Cr(VI)) (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
Page 6
General OPERATING GUIDE for use with PIEZOELECTRIC CHARGE MODE ACCELEROMETERS SPECIFICATION SHEET, INSTALLATION DRAWING AND CALIBRATION INFORMATION ENCLOSED PCB ASSUMES NO RESPONSIBILITY FOR DAMAGE CAUSED TO THIS PRODUCT AS A RESULT OF PROCEDURES THAT ARE INCONSISTENT WITH THIS OPERATING GUIDE. 1.0 INTRODUCTION Enclosed is a Specification Sheet, which lists the Congratulations on the purchase of a quality PCB charge...
Page 7
upon the coupling type or time constant of the external charge amplifying device. This characteristic may be precision-machined mounting surface with fixed or adjustable, depending on the charge amplifier. minimum finish µin (0.00016 Please refer to the appropriate charge amplifier recommended.
Page 8
METHOD 1 - Adhesive Mounting Base This method involves attaching a base to the test structure, then securing the sensor to the base. This allows for easy removal of the accelerometer. Also, since the bases are of a “hard-coated” aluminum design, they provide electrical isolation to eliminate ground loops and reduce electrical interference that may be present on the surface of the test object.
pins.) In many cases, it is also helpful to protect the Once each element is connected, allow a few minutes for solder joint with heat-shrink tubing or epoxy. the system to thermally stabilize. Place the switch on the charge amplifier in the OPERATE position and Note: If you do not have the experience or resources to proceed with the measurement.
For these reasons, it is recommended that a recalibration cycle be established for each accelerometer. This schedule is unique and is based on a variety of factors, such as extent of use, environmental conditions, accuracy requirements, trend information obtained from previous calibration records, contractual regulations, frequency of “cross-checking”...
Figure 11. Typical Test Accelerometer Response 7.3 PCB CALIBRATION PROCEDURE Numerous precautions are taken at PCB to insure accurate and repeatable results. This section provides a brief overview of the primary areas of concern. Since the Back-to-Back Calibration technique relies on Figure 12.
Page 12
along the axis of motion, as the mounting stiffness may be altered. The vibration at the test sensor sensing element may differ from the vibration at the reference standard due to a “cantilever” effect, seen in Figure 15. Figure 13. Stud Mounting For adhesive mount sensors, use a thin, stiff layer of temporary adhesive such as quick-bonding gel or superglue.
Page 13
Model Number CHARGE OUTPUT ACCELEROMETER Revision: A EX357E91 ECN #: 49874 Performance ENGLISH OPTIONAL VERSIONS Sensitivity(± 10 %) 5 pC/g .51 pC/(m/s²) Optional versions have identical specifications and accessories as listed for the standard Measurement Range ± 1000 g pk ±...
Page 14
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. EX357E91 WAS 357E91 49874 .656 [16.66] DIRECTION OF ACCELERATION...
Page 15
EU Declaration of Conformity PS199 In Accordance with ISO/IEC 17050 Manufacturer: PCB Piezotronics, Inc Authorized PCB Piezotronics Europe GmbH European Porschestrasse 20-30 3425 Walden Avenue Representative: 41836 Hückelhoven,Germany Depew, New York 14043 USA Certifies that type of equipment: High Temperature Accelerometer(s)
Page 16
Certificate history: Status: Current Issue No: 0 Date of Issue: 2019-02-04 Applicant: PCB Piezotronics, Incorporated. 3425 Walden Avenue, Depew, NY 14043-2495 United States of America Equipment: High Temperature Vibration Sensors: EX357 SERIES Optional accessory: Type of Protection: Intrinsic Safety 'ia'; Type of Protection "nA"...
Page 17
Page 2 of 4 Date of issue: 2019-02-04 Issue No: 0 Manufacturer: PCB Piezotronics, Incorporated. 3425 Walden Avenue, Depew, NY 14043-2495 United States of America Manufacturing locations: This certificate is issued as verification that a sample(s), representative of production, was assessed and tested and found to comply with the IEC Standard list below and that the manufacturer's quality system, relating to the Ex products covered by this certificate, was assessed and found to comply with the IECEx Quality system requirements.This certificate is granted subject to the conditions as set out in IECEx Scheme...
Page 18
IECEx Certificate of Conformity Certificate No.: IECEx ETL 19.0002X Page 3 of 4 Date of issue: 2019-02-04 Issue No: 0 EQUIPMENT: Equipment and systems covered by this Certificate are as follows: The EX357 Series products are High Temperature Vibration Sensors. There are various Models based on different construction combinations: family type, connector type, cable type, and/or cable length.
Page 19
IECEx Certificate of Conformity Certificate No.: IECEx ETL 19.0002X Page 4 of 4 Date of issue: 2019-02-04 Issue No: 0 Ex ia: The apparatus must only be connected to a certified associated intrinsically safe equipment. This combination must be compatible regarding intrinsic safety rules (see electrical parameters). The apparatus shall be connected according to drawing 69892.
Page 27
Report. This authorization also applies to multiple listee model(s) identified on the correlation page of the Listing Report. This document is the property of Intertek Testing Services and is not transferable. The certification mark(s) may be applied only at the location of the Party Authorized To Apply Mark. PCB Piezotronics, Inc. PCB Piezotronics, Inc. Applicant: Manufacturer:...
Page 28
AUTHORIZATION TO MARK Intrinsically Safe Apparatus And Associated Apparatus For Use In Class 1, 2, 3, Division 1, Hazardous (Classified) Locations [UL 913:2013 Ed.8+R:16Oct2015] Nonincendive Electrical Equipment For Use In Class I And II, Division 2 And Class III, Divisions 1 And 2 Hazardous (Classified) Locations [UL 121201:2017 Ed.9] Explosive Atmospheres –...