Edaphic Scientific MS 08 Operating Instructions Manual

Edaphic Scientific MS 08 Operating Instructions Manual

Multi-sensor measuring instrument for h2s/h2o2/o2/o3/h2/ph/temperature

Advertisement

Quick Links

Multi-sensor Measuring Instrument MS 08
for H
S/H
O
/O
/O
/H
/pH/temperature
2
2
2
2
3
2
- Operating Instructions -
Edaphic Scientific Pty Ltd
www.edaphic.com.au
info@edaphic.com.au

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MS 08 and is the answer not in the manual?

Questions and answers

Summary of Contents for Edaphic Scientific MS 08

  • Page 1 Multi-sensor Measuring Instrument MS 08 for H /pH/temperature - Operating Instructions - Edaphic Scientific Pty Ltd www.edaphic.com.au info@edaphic.com.au...
  • Page 2: Table Of Contents

    Multi-sensor Measuring System _________________________________________________________________________________________________ Contents Preface Structure and components of the measuring system, system equipment Operation General operating instructions Putting into operation Description of the menu items of the measuring program 3.3.1 Key description 3.3.2 Description of the menus Accessories and function extension Maintenance Description of the H micro-sensors special features...
  • Page 3: Preface

    Multi-sensor Measuring System _________________________________________________________________________________________________ Preface The battery-operated Multi-sensor Measuring System has been developed for the determination of the analytes H S/sulphide resp. total sulphide amount, dissolved oxygen, dissolved hydrogen and ozone. Additionally some further sensors have been integrated into the measuring system depending on the analyte which has to be determined.
  • Page 4: Structure And Components Of The Measuring System, System Equipment

    Multi-sensor Measuring System _________________________________________________________________________________________________ Structure and components, system equipment The Multi-sensor measuring system MS 08 can be delivered with the following sensor combinations: S-sensor with combined pH/temperature-sensor (for measurements within the range 0-8,5 pH) Oxygen sensor with temperature sensor Ozone sensor with temperature sensor...
  • Page 5 Multi-sensor Measuring System _________________________________________________________________________________________________ In the first line from above is visible the analyte, which has to be determined (ozone, oxygen, total Sulphide, H S), or alternatively the measured sensor current (instead of the analytes name "current" is visible). Next to it at the right edge of the display the actual measured temperature is shown in "Degrees Celsius -°C".
  • Page 6: Operation

    Multi-sensor Measuring System _________________________________________________________________________________________________ Operation General Operating Instructions The Multi-sensor Measuring System is operated only by means of the numeric keyboard. After switching on the instrument starts with the last used sensor data and adjustments (e.g. as in fig. 2). The operation of the measuring device is menu driven and self explaining in essential parts.
  • Page 7: Description Of The Menus

    Multi-sensor Measuring System _________________________________________________________________________________________________ Switching on of the measuring device is realized by pressing the yellow key "On/Off". The measuring signals of the sensors are renewed at the display every 1-2 seconds. Please take note, that the measuring device uses the last stored data for the calculation and display (sensor, analyte, calibration data, ...).
  • Page 8 Multi-sensor Measuring System _________________________________________________________________________________________________ To reach these menus, please press after switching on the measuring device ("On/Off") the key "Menu". By means of the arrows  or  or  or  you will reach the requested menu. For adopting the adjustments you have to choose "Special"...
  • Page 9 Multi-sensor Measuring System _________________________________________________________________________________________________ ordered. Therefore for every sensor exists a specific equation with the variable factors a . The values for a you will also find at the sensors calibration 0...3 0...3 sheet (below, right in the formula). Feed in these coefficients in the designated lines and confirm every a coefficient with "Enter".
  • Page 10: Accessories And Function Extension

    Multi-sensor Measuring System _________________________________________________________________________________________________ Rinse the pH sensor carefully with distilled water and dive in the sensor after- wards into the second buffer. Wait again for the adjustment of "mV". Correct by means of the arrows  and  the pH until the value has reached the pH from the bottle with the buffer (please consider the temperature).
  • Page 11: Maintenance

    Multi-sensor Measuring System _________________________________________________________________________________________________ Protocoll: none Maintenance To save battery capacity you should switch on the device only, if you want to measure. Take care, that the batteries have enough capacity before starting with your measurement. Please read again the chap- ter 3.1 general operating instructions.
  • Page 12 Multi-sensor Measuring System _________________________________________________________________________________________________ A P P E N D I X ______________________________________________________________________________________________...