User's Manual
BH027C11 EN C
panametrics.com

Advertisement

Table of Contents
loading

Summary of Contents for Panametrics PanaFlow HT

  • Page 1 PanaFlow™ HT User’s Manual BH027C11 EN C panametrics.com...
  • Page 3 User’s Manual BH027C11 EN C March 2024 panametrics.com Copyright 2024 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective...
  • Page 4 [no content intended for this page]...
  • Page 5 Preface Services Panametrics provides customers with an experienced staff of customer support personnel ready to respond to technical inquiries, as well as other remote and on-site support needs. To complement our broad portfolio of industry-leading solutions, we offer several types of flexible and scalable support services including: Training, Product Repairs, Service Agreements and more.
  • Page 6 Environmental Compliance Waste Electrical and Electronic Equipment (WEEE) Directive Panametrics is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative, directive 2012/19/EU. The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment.
  • Page 7: Table Of Contents

    2.3.1.1 PanaFlow HT Transmitter Identification ........
  • Page 8 Wiring......................37 PanaFlow HT User’s Manual...
  • Page 9: Chapter 1. Introduction

    Introduction Overview Thank you for purchasing the PanaFlow HT ultrasonic flowmeter. PanaFlow HT is a wetted ultrasonic flow meter that is available with SIL certification by design to give you confidence in your flow measurement and to provide reliable flow meter operation for both safety and process control systems. In addition to the peace of mind that SIL...
  • Page 10: Sil Application

    Chapter 1. Introduction SIL Application The PanaFlow HT with appropriate flow meter selection can be a SIL2 ultrasonic flowmeter (sensor) with the capability of providing a SIL3 system in a redundant design configuration. The PanaFlow HT system is IEC61508 certified (when selection) through a complete design validation from a third-party organization. By achieving a third-party certification, we have proven the required design rigor through the product safety lifecycle, and the implementation of functional safety management.
  • Page 11: Chapter 2. Installation

    Panametrics Customer Care immediately for assistance. Please note that your PanaFlow HT system may come in one of six common configurations as shown below or a custom design system. Also, the electronics may be in a separate box from the meter body.
  • Page 12 Chapter 2. Installation Figure 4: 2P1R Configuration Figure 5: 1P2R Configuration Figure 6: 1P3R Configuration PanaFlow HT User’s Manual...
  • Page 13: Identification

    Figure 8: 2P4R Configuration 2.2.1 Identification The PanaFlow HT meter has two separate labels for identification, depending on configuration. 2.2.1.1 PanaFlow HT Transmitter Identification ThePanaFlow HT is supplied with both a serial number label and a certification label for identification of the instrument.
  • Page 14: Meter Body Identification

    Chapter 2. Installation Figure 9: PanaFlow HT labels (stainless steel enclosure) 2.2.1.2 Meter Body Identification The PanaFlow HT meter will have a label for the assembly per figure below. Figure 10: Label for the assembly 2.2.2 Transport of PanaFlow HT Meter Body Below indicates the proper way to cinch the lifting straps to the flowmeter.
  • Page 15: Site Considerations

    2.3.1.1 Transducer location For a given fluid and pipe, the PanaFlow HT's accuracy depends on the location and alignment of the transducers. In addition to accessibility, when planning for transducer location, adhere to the following guidelines: •...
  • Page 16: Xmt1000 Electronics Mounting & Location

    For compliance with the European Union's Low Voltage Directive, this unit requires an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit. PanaFlow HT User’s Manual...
  • Page 17 / must be upwards as the measurement method requires the pipe to be full. See below figure for mounting the XMT1000 electronics Figure 12: XMT1000 Enclosure clearances (ref. dwg. 712-2164) PanaFlow HT User’s Manual...
  • Page 18: Making Electrical Connections

    Digital outputs (totalizer, frequency, calibration) • Modbus or Service port • Hart or Foundation Field Ports (if applicable) • Additional analog input or output (if applicable) • Additional SIL analog output (If applicable) • Transducer (flying leads) • Line power PanaFlow HT User’s Manual...
  • Page 19: Chapter 3. Programming

    Some menus are restricted to only those users with the proper passcodes. 3.1.1 HMI Features Figure 13: PanaFlow HT HMI The six keys on the magnetic keypad are used to program the PanaFlow HT: Key Symbol Key Name Functions...
  • Page 20: Indicator Lights

    Measurement View 3.3.1 Measurement View On power up, the PanaFlow HT meter shows the following screens: • Panametrics Logo screen • Meter Initialization screens • Power-on self-tests and results •...
  • Page 21: Changing Display Format

    Errors/Warnings Totalizer Flow Info Display Format Velocity One Variable 0.000 Two Variable Ch1 Velocity Totalizer 0.000 Display Format Fwd Vol Total One Variable 0.000 Two Variable Rev Vol Total Totalizer 0.000 Figure 15: Changing Display Format PanaFlow HT User’s Manual...
  • Page 22: Selecting A Composite Measurement To Display

    Reynolds # On Board Temperature Inputs Composite Fluid Temperature Reynolds # On Board Temperature Pressure Inputs Diagnostics Composite Diagnostic Cycle Time Reynolds # On Board Temperature Inputs Diagnostics Figure 16: Selecting a Composite Measurement to Display PanaFlow HT User’s Manual...
  • Page 23: Selecting A Channel Measurement To Display

    Error Status Ch 1/Diagnostics ATw Amplitude Up ATw Amplitude Dn ATw SNR Up ATw SNR Dn Ch 1/Diagnostics ATw Gain Up ATw Gain Dn Wedge Soundspeed Wall Time Figure 17: Selecting a Channel Measurement to Display PanaFlow HT User’s Manual...
  • Page 24: Totalizer Display

    Press [] until the lock icon on the meter’s Measurement View display is highlighted, then press [ENTER]. In the Main Menu Scroll down and select [Program], then press [ENTER]. Scroll and select desired access level [Operator], then press [ENTER]. Enter the password Operator access level, and press [ENTER]. PanaFlow HT User’s Manual...
  • Page 25 Figure 19: Log-in Steps and Primary pages IMPORTANT: If the keypad has not been pressed for 5 minutes, the PanaFlow HT exits the Program and returns to displaying measurements. Because changes can only be retained after the user confirms them, the meter discards any unconfirmed configuration changes.
  • Page 26: Main Program

    Please refer to the XMT1000 User manual for detailed programming of the XMT1000 Electronics such as System Settings, Inputs/Outputs, Wetted Meter Programming, and Calibration. See chapter “Programming“ for instrument programming step-by-step instructions, or refer to appendix “Menu map” for the full menu map reference guide. PanaFlow HT User’s Manual...
  • Page 27: Chapter 4. Error Codes And Troubleshooting

    All the possible PanaFlow HT error code messages are discussed in this chapter, along with the possible causes and the recommended actions. When an error code is generated, it will appear in the lower left corner of the LCD screen, as discussed in Programming Chapter.
  • Page 28: Single Channel Error

    Viewing Channel Specific Error/Warnings To indicate the health of the meter, PanaFlow HT has built-in Error codes. The Channel specific errors are very critical in determining the corrective actions required. Figure 20 below shows the steps to view current channel specific errors/warnings.
  • Page 29 If the gain is greater than 35 dB, change the Transmit Voltage to “High” (Refer Programming Chapter). Refer to “Fluid and Pipe Problems” and “Transducer Problems” sections to correct any issues. Gather required diagnostics data before contacting Panametrics PanaFlow HT User’s Manual...
  • Page 30: Fluid And Pipe Problems

    The flow meter has not been The condition is not self-recovering been calibrated calibrated at the factory and will not automatically correct and hence not making itself. Contact Panametrics factory to measurements. Please get more information on the meter contact Panametrics setup factory...
  • Page 31: Fluid Problems

    • Pipe problems Read the following sections carefully to determine if the problem is related to the fluid or the pipe. If the instructions in this section fail to resolve the problem, contact Panametrics for assistance. 4.4.1 Fluid Problems Most fluid-related problems result from a failure to observe the flow meter system installation instructions, as described in Chapter: Installation.
  • Page 32: Transducer Problems

    If error persists after power cycle, contact Panametrics factory Fault: CPU registers have stuck bits. Try power S12: CPU CPU error cycling the meter. If error persists after power cycle, contact Panametrics factory PanaFlow HT User’s Manual...
  • Page 33 SIL Analog output and try power cycling the meter. If error persists after power cycle, contact Panametrics factory S30: Flash Save Failed Save to Flash Failed Error: Request to Save failed. Try again. If error persists, contact Panametrics factory. PanaFlow HT User’s Manual...
  • Page 34: Communication Errors (C-Errors)

    Transmitter Errors These errors are from the Transmitter subsystem. Should you encounter one of the Transmitter Errors, follow recommended actions as indicated in Table and contact Panametrics factory. Table 5: Transmitter Error Description and Recommended Actions Error Code Error Message...
  • Page 35: Option I/O Errors

    Description A1:AnalogCh(S2:3) Error! ADC Channel(S2:3) is not Analog input /RTD input is not working. If error responding persists after power cycle, contact Panametrics factory A2:AnalogCh (S2:4) Error! ADC Channel(S2:4) is not Analog input /RTD is not working. If error persists...
  • Page 36: 4.10 Diagnostics Data

    To determine the health of the meter, PanaFlow™ LC has built-in diagnostic parameters. Please refer to Table 7 below for diagnosing any problems with the system. If the meter shows errors and the diagnostics data indicate issues, fill in the User/Service record appendix before contacting Panametrics factory. Table 7: Diagnostic Parameter Description and Health Indicators...
  • Page 37 Signal quality of the >1000 <1000 Down downstream transducer Amplitude Up Signal amplitude of the >14 and <32 >32 or <14 upstream transducer Amplitude Down Signal amplitude of the >14 and <32 >32 or <14 downstream transducer PanaFlow HT User’s Manual...
  • Page 38 Chapter 4. Error Codes and Troubleshooting [no content intended for this page] PanaFlow HT User’s Manual...
  • Page 39: Chapter 5. Maintenance And Service

    Local requirements may or may not allow field replacement of any components in this flow metering system without a proper calibration of the entire system at an approved calibration facility. Check with your local Panametrics & Panametrics Flow meter representative to determine if field replacement of components is allowed.
  • Page 40 Chapter 5. Maintenance and Service [no content intended for this page] PanaFlow HT User’s Manual...
  • Page 41: Appendix A. Specification

    -82 ft/s to +82 ft/s (-25 m/s to +25 m/s) IEC61508 certified SIL2 certification with single set of electronics SIL3 achievable with redundant design system Consult Panametrics for proper ordering code when SIL certification is required if any questions arise. PanaFlow HT User’s Manual...
  • Page 42: Meter Body And Transducers

    Non-armored cable for use in conduit Flame retardant SWA armored cable (with cable glands) Fire resistant SWB armored cable (with cable glands) Flame retardant SWA armored cable (with cable glands) Electronics Refer to XMT1000 User Manual for specifications. PanaFlow HT User’s Manual...
  • Page 43: Appendix B. Digital Communications

    Appendix B. Digital Communications Appendix B. Digital Communications For information on how to communicate using Foundation Fieldbus, Modbus, HART, or Wireless HART, refer to the XMT1000 User Manual appendix section. PanaFlow HT User’s Manual...
  • Page 44 Appendix B. Digital Communications [no content intended for this page] PanaFlow HT User’s Manual...
  • Page 45 Invariable Flash Memory Test Failed 0x00001000 Invariable SRAM Memory Test Failed 0x00002000 Variable Memory Test Failed 0x00004000 FPGA Configuration Test Failed 0x00008000 Temperature Test Failed 0x00010000 Driver Fault 0x00020000 Watch-dog Test Failed 0x00040000 Analog Readback Failure 0x00080000 PanaFlow HT User’s Manual...
  • Page 46 MCU register access fail 0x00000400 System Command failed 0x00000800 Get GUI Node Fail 0x00001000 Node Memory Fail 0x00002000 Font API Initialize Fail 0x00004000 XML File Initialize Fail 0x00008000 Disconnect Std Dout 0x00010000 Aout(std) Out Of Range 0x00020000 PanaFlow HT User’s Manual...
  • Page 47 (S2:3) Input NotConnect! 0x00000200 (S2:4) Input NotConnect! 0x00000400 (S2:3) Ch OverRange Err! 0x00000800 (S2:4)Ch OverRange Err! 0x00001000 SerialNo Error! 0x00020000 Aout(S2:1)OutOfRange! 0x00800000 Aout(S2:2)OutOfRange! 0x01000000 Board Option Err! 0x20000000 (S2:3) Ch UnderRange! 0x40000000 (S2:4) Ch UnderRange! 0x80000000 PanaFlow HT User’s Manual...
  • Page 48 Appendix C. Error code bit-field representation [no content intended for this page] PanaFlow HT User’s Manual...
  • Page 49: Appendix C. Ce Mark Compliance

    Grounded using a cable gland. outside of jacket Power Armored 14 AWG 2 conductor Grounded using a cable gland. Note: If the PanaFlow HT, is wired as described above, the unit will comply with the EMC and LVD Directives. PanaFlow HT User’s Manual...
  • Page 50 Appendix D. CE Mark Compliance [no content intended for this page] PanaFlow HT User’s Manual...
  • Page 51 • If Panametrics determines that the damage is not covered under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
  • Page 52 Warranty [no content intended for this page] PanaFlow HT User's Manual...
  • Page 54 Shannon, County Clare Ireland Tel: +353 (0)61 470291 E-mail: panametricstechsupport@bakerhughes.com Scan here for Panametrics support and service information Copyright 2024 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective holders.