Page 1
MG-101 User’s Manual 910-115 E panametrics.com August 2022...
Page 3
MG-101 Field Moisture Calibration System User’s Manual 910-115 Rev. E August 2022 panametrics.com Copyright 2022 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective...
Page 5
Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary equipment. Examples include safety glasses, protective headgear, safety shoes, etc. Unauthorized Operation Make sure that unauthorized personnel cannot gain access to the operation of the equipment. MG-101 User’s Manual...
Page 6
The crossed-out wheeled bin symbol invites you to use those systems. If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration. Visit www.bakerhughesds.com/health-safety-and-environment-hse for take-back instructions and more information about this initiative. MG-101 User’s Manual...
Vent valve See the diagram and explanation in Figure 1 on page 2 for a description of how the MG-101 works. As shown in Figure 1 below, dry nitrogen enters at the dry gas inlet A, passes through the fixed internal pressure regulator, and splits into the following three streams: •...
Page 10
Vent Valve Gas Outlet Dry Down Dry Gas Dry Gas Valve Temp Second First Dilution 5μ Filter Dilution Flowmeter Flowmeter Wet Gas Wet Gas Diluted Flowmeter Flowmeter Dry Gas Saturator Inlet Bottle Pressure Regulator Figure 1: Flow Schematic MG-101 User’s Manual...
1.3.1 Operating Range While the driest dew/frost point temperature capability specified for the MG-101 is -75°C (-103°F), the actual limit is determined by the moisture content of the dry nitrogen source and by the ambient temperature. The driest dew/frost point temperature the MG-101 can produce is the dew/frost point temperature of the dry nitrogen source plus 25°C (45°F).
Page 12
Chapter 1. General Information [no content intended for this page] MG-101 User’s Manual...
IMPORTANT: Be sure you read and understand “Initial Considerations” on page 3 before proceeding. To prepare the MG-101 for operation, complete the following steps: Remove the back panel by removing the two small screws at the bottom (see Figure 2 below).
Page 14
Selection Valve Vent Valve Inlet Fitting Outlet Fitting Figure 3: MG-101 Front Panel Connect the nitrogen supply to the MG-101 inlet fitting via the stainless steel pressure regulator, using 1/4” OD stainless steel tubing (see Figure 4 below). ® Note: Swagelok fittings are preferred throughout the installation.
Page 15
Chapter 2. Setup and Operation Connect the calibration chamber to the MG-101 outlet fitting, using 1/4” OD SS tubing. Then, leak test the connection. Note: Panametrics Sample Cell #2830 is the preferred calibration chamber for M2 Type probes At the outlet of the calibration chamber, connect at least 5 ft (1.5 m) of 1/4” OD tubing as a vent to atmosphere.
Chapter 2. Setup and Operation Operating Procedure To operate the MG-101, refer to Figure 5 below and Figure 6 on page 9 to complete the following steps: Fully close all four rotameter valves by turning them clockwise. Set the WET/DRY selector valve to DRY.
Page 17
After the calibration system has reached equilibrium, set the valve on the WET/DRY rotameter to the WET position. Temperature Wet/Dry Gauge Selector Valve Wet Gas Wet Gas Diluted Dry Gas First Dilution Dry Gas Second Dilution Figure 6: Rotameters and Valves MG-101 User’s Manual...
Check the saturator bottle temperature on the temperature gauge. For each moisture calibration point you wish to generate, look at Data Tables 2 and 3 (supplied with the MG-101). Above the line which reads “Generated Dew Points (Deg C),” is a row of values headed “Ambient Temperature (Deg C).”...
After you have completed the calibration or other calibrated moisture operation at this dew/frost point temperature, either shut down the system (see “System Shutdown” on page 12), or repeat the steps in this section for the next calibration point. MG-101 User’s Manual...
(whichever is indicated in the table). IMPORTANT: Be sure to read the steel ball or glass ball (S or G), as indicated in the table. After equilibrium has been established, the MG-101 will be delivering a calibration gas with the selected dew/frost point temperature.
2.2.5 Preparation for Shipping To ship or to move the MG-101, refer to Figure 7 below and empty the saturator bottle as follows: Remove the rear panel by removing the two small screws at the bottom. Fully open the vent valve by turning it counter-clockwise.
Page 22
Chapter 2. Setup and Operation [no content intended for this page] MG-101 User’s Manual...
Dimensions (W x H x D) 12” x 18” x 6” (304.8 mm x 457.2 mm x 152.4 mm) ® Inlet and Outlet Connections 1/4” Swagelok tube fittings Note: See Figure 8 on page 16 for a dimensional drawing of the MG-101. MG-101 User’s Manual...
Appendix A. Supplemental Information Appendix A. Supplemental Information Formulas The moisture content generated by the MG-101, as determined by the water vapor pressure, is calculated with the following equation: --------------------------------------- ----------------------------------------------------------------------------- where: = the generated vapor pressure of water...
A.2.1 Example 1 Data Table 2 is provided with the MG-101 to enable you to use the field calibration system to check sensor probe calibration without calculating the value of p . Data Table 2 was compiled by calculating p and thus, the dew/frost point temperature by Equation 2 on the previous page.
Page 31
• If Panametrics determines that the damage is not covered under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
Page 32
Warranty [no content intended for this page] MG-101 User’s Manual...
Page 34
Customer Support Centers U.S.A. The Boston Center 1100 Technology Park Drive Billerica, MA 01821 U.S.A. Tel: 800 833 9438 (toll-free) 978 437 1000 E-mail: mstechsupport@bakerhughes.com Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland Tel: +353 (0)61 470200 E-mail: mstechsupport@bakerhughes.com Copyright 2022 Baker Hughes company.