(Confined Space/4-gas monitor)
The GX-2012 is designed to monitor: Lower
Explosive Limit (LEL), H

Normal Operation:

Press and hold the "Power" button, allow the
instrument approximately 10 minutes to
warm-up.

Fresh Air Calibration

Move to a clean air location, attach the zero filter to the end
of the inlet. Press and hold "Air" button, release when
prompted.

Perform a Bump-Test:

Instrument should be in normal operation mode.
Connect the gas standard to the inlet and allow the values to
stabilize, if values do not approximately equal printed gas
values consider recalibration.

Charging the Instrument:

A charging dock and AC power adapter are provided with
your rental. A red LED light on the dock indicates the instru-
ment is connected to power and is charging. Ensure the in-
strument is charged after daily use
Maxim
Environmental and Safety Inc.
sales@maximenvironmental.com
www.maximenvironmental.com
Maxim Quick Start Guide
GX-2012
S, CO and O
2
2
Mississauga: (905)670-1304
Ottawa: (613)224-4747
Toll Free: (888)285-2324

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GX-2012 and is the answer not in the manual?

Questions and answers

Summary of Contents for Maxim GX-2012

  • Page 1 Maxim Quick Start Guide GX-2012 (Confined Space/4-gas monitor) The GX-2012 is designed to monitor: Lower Explosive Limit (LEL), H S, CO and O Normal Operation: Press and hold the “Power” button, allow the instrument approximately 10 minutes to warm-up. Fresh Air Calibration Move to a clean air location, attach the zero filter to the end of the inlet.
  • Page 2 Do not allow the instrument to suck in water or soil as • this can lead to internal damages. Scan QR Code to access the PDF manual and other resources. Maxim Environmental and Safety Inc. Mississauga: (905)670-1304 sales@maximenvironmental.com Ottawa: (613)224-4747 www.maximenvironmental.com...

Table of Contents