Metria pHscan 30 Quick Start Manual

Pocket ph meter

Advertisement

Quick Links

Introduction
Thank you for choosing the pHscan30 pocket pH tester. This manual provides a step-by-
step guide to help you operate the meter. Please carefully read the following instructions
before use.

Inserting the Batteries

Take out the three "G13A" button cell batteries from the packaging. Follow the steps
below to insert the batteries into the battery compartment.
• Remove the battery compartment lid. Insert the batteries into the battery
compartment, noting polarity.
• After installation, replace the battery compartment lid and turn it clockwise until it
is tight.

Prior to Use

Remove the electrode protective cap from the tester. If the electrode bulb dries out, soak
the electrode module in storage solution or tap water for at least 15 minutes. Do not use
distilled or deionized water, as it will shorten the sensor's life.
pHscan 30 Pocket pH Meter
PHMT-030-001
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the pHscan 30 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Metria pHscan 30

  • Page 1 30 Pocket pH Meter PHMT-030-001 Introduction Thank you for choosing the pHscan30 pocket pH tester. This manual provides a step-by- step guide to help you operate the meter. Please carefully read the following instructions before use. Inserting the Batteries Take out the three “G13A”...
  • Page 2 Preparation of pH Standard Buffer Solutions The tester is packaged with three pH buffer packets required for calibration. • Open the pH 7.00 buffer packet, place the powder into a 250 ml volumetric flask. Pour distilled water up to the scale line, totaling 250 ml, and mix the solution until the reagent is completely dissolved.
  • Page 3 The tester will automatically display “CAL 4” or “CAL 10” (depending on the option selected in step 2). 6. Rinse the electrode with distilled water. Dip the electrode into the pH 4.01 or 10.01 standard buffer solution. Stir the tester gently. 7.
  • Page 4 Hold Function 1. During the measurement, press the HOLD key; the tester immediately freezes the currently displayed value, and the “HOLD” indicator appears on the display. 2. Press the key again to release the measured reading, allowing you to continue taking measurements.
  • Page 5 Check if the electrode is clogged, dirty, or broken. Incorrect buffer used or contaminated buffer solution: Use fresh pH buffer solutions for calibration. Electrode is broken: Replace the electrode module. Specifications Model pHScan 30 pH Range 0.00 – 14.00 pH Accuracy ±0.05 pH Calibration Points pH Buffer Options USA Standard (pH 4.01 / 7.00 / 10.01)