30 Pocket pH Meter PHMT-030-001 Introduction Thank you for choosing the pHscan30 pocket pH tester. This manual provides a step-by- step guide to help you operate the meter. Please carefully read the following instructions before use. Inserting the Batteries Take out the three “G13A”...
Page 2
Preparation of pH Standard Buffer Solutions The tester is packaged with three pH buffer packets required for calibration. • Open the pH 7.00 buffer packet, place the powder into a 250 ml volumetric flask. Pour distilled water up to the scale line, totaling 250 ml, and mix the solution until the reagent is completely dissolved.
Page 3
The tester will automatically display “CAL 4” or “CAL 10” (depending on the option selected in step 2). 6. Rinse the electrode with distilled water. Dip the electrode into the pH 4.01 or 10.01 standard buffer solution. Stir the tester gently. 7.
Page 4
Hold Function 1. During the measurement, press the HOLD key; the tester immediately freezes the currently displayed value, and the “HOLD” indicator appears on the display. 2. Press the key again to release the measured reading, allowing you to continue taking measurements.
Page 5
Check if the electrode is clogged, dirty, or broken. Incorrect buffer used or contaminated buffer solution: Use fresh pH buffer solutions for calibration. Electrode is broken: Replace the electrode module. Specifications Model pHScan 30 pH Range 0.00 – 14.00 pH Accuracy ±0.05 pH Calibration Points pH Buffer Options USA Standard (pH 4.01 / 7.00 / 10.01)