Table 9. Guidance and Manufacturer's Declaration – Proximity Fields
Proximity Test
IEC 61000-4-3
IEC 60601 Test Level
385 MHz: 27 V/m @ 18 Hz pulse
modulation
450 MHz: 28 V/m @ FM
modulation
710 MHz, 745 MHz, 780 MHz:
9 V/m @ 217 Hz pulse
modulation
810 MHz, 870 MHz, 930 MHz:
28 V/m @ 18 Hz pulse
modulation
1720 MHz, 1845 MHz, 1970 MHz:
28 V/m @ 217 Hz pulse
modulation
2450 MHz: 28 V/m @ 217 Hz
pulse modulation
5240 MHz, 5500 MHz, 5785 MHz:
9 V/m @ 217 Hz pulse
modulation
Compliance Level
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m