transmission latency in each direction (2x), and a receive-to-transmit mode (RX‑to‑TX) time. Data is resent if
not sent successfully. Each key press may transmit up to 8 data packets, depending on the number of
packets that need to be transmitted (that is, if there is only one packet to transmit, only one packet will be
transmitted).
Wireless Security Measures
The wireless signals are secured through device system design that includes the following:
The generator will encrypt its wireless communication.
Only one patient controller or clinician programmer may communicate with the generator at the same
time.
A unique key for each unit that is checked during each transmission.
Built-in pairing that specifies valid and legitimate pairing among units.
Proprietary authentication in addition to the pairing procedure specified in Bluetooth® Low Energy (LE)
wireless technology, which includes an element of proximity.
A proprietary algorithm that detects and prevents an unauthorized user from attempting to pair with
the generator.
Troubleshooting for Wireless and Coexistence Issues
If you experience issues with the wireless communication between the generator and the clinician
programmer or patient controller, try the following:
Decrease the distance between the devices.
Move the devices so they share line of sight.