Models 8672S/86720A
General Information
Table 1-5. Recommended Test Equipment
Instrument
Critical Specifications
Recommended Mode!
Use*
Attenuator
6 dB (not critical)
Range: dc to 18 GHz
SWR: 1.5 maximum
HP 8491B
Option 006
P
Controller, HP-IB
HP-IB Compatible
HP 85F
mm
Diode Detector
Range: 0.01 to 18 GHz
SWR: 1.7 maximum
Sensitivity: >0.4 mV/g/Wl
HP 8472B
P, T
Frequency Counter
Range: 20 kHz to 18 GHz
Resolution: 10 Hz
HP 5343A
P, T
Mixer
0.01 to 2 GHz response
ZFM—150**
P, T
Oscilloscope
Bandwidth: 200 Mhz
Vertical Sensitivity: lOmV/div
Vertical Input: 500 ac or dc coupled
Delayed Sweep Mode: 20 ns/div
External Sweep Capability
HP1715A
P, A, T
Power Meter
Frequency Range: 0.01 to 18 GHz
Accuracy: ±0.1 dB
Compatible with Power Sensor
HP 436A
P, A, T
Power Sensor
Frequency Range: 0.01 to 18 GHz
Input Impedance: 500
Range: +17 to -25 dBm into 500
HP 8481A
P, A, T
Pulse Generator
Rate: 10 Hz to 4 MHz
Rise and Fall Times: >5 dBm
Output Impedance: 500
Output Level: 0 to 3.5V
Pulse Width: 100 ns to 2 ^/s
HP 8013B
P, A, T
Spectrum Analyzer
Frequency Range: 0.01 to 18 GHz
Frequency Span 50 kHz to 200 MHz/div
Resolution Bandwidth: 30 Hz to 300 kHz
Amplitude Range: +13 to -120 dBm
Vertical Sensitivity: 2 dB/div minimum
HP 8565A
P, T
Voltmeter, Digital
P, A, T
* P = Performance Testing, A = Adjustments, T = Troubleshooting
**Mini-Circuits Lab, 2625 East 14th Street, Brooklyn, NY 11235, Tel (212) 769-0200.