Model 467A
GENERAL.
3-1.
3-2.
The Model 467A Power Amplifier/Supply can be
operated as a power amplifier or a power supply.
Controls on the front panel provide:
either operating mode and (2) selection of fixed or
variable amplifier gain and power supply ranges.
3-3.
DESCRIPTION
3-4.
Figure 3- 1 gives the description and function
of the 467A controls, indicators and connectors.
3-5.
OPERATING
3-6.
When operated as a power amplifier (non-invert­
ing), the Model 467A has four calibrated gain factors
(AMPLIFIER:
Xl, X2, X5, and X10) and provides a
maximum output power of 10 watts peak for input fre­
quencies between dc and 1 MHz. Accuracy of ±0.3%
from dc to 10 kHz is obtained when operating into a
load of 40 ohms or greater.
lated), the instrument can be operated in five ranges
of plus or minus output voltages (SUPPLY:
±4,
±10,
and ±20).
The amplifier gain, or the ± dc
output can be varied with the VAR. GAIN/VOLTS ADJ.
control. The input and output ter minals are dc isolated
from earth ground.
Connection of a grounding strap
between the power line ground
references both the INPUT and OUTPUT (� terminals
to earth ground.
3-7.

AMPLIFIER OPERATING INSTRUCTIONS.

a.
Set the AMPLIFIER/SUPPLY control to the
AMPLIFIER pOSition,
gain setting (VAR., Xl, X2, X5, or X10).
b.
Set the LINE switch to ON.
To assure maximum stability, allow
the instrument to warm up for 30
minutes.
SECTION
OPERAT IN G
( 1) selection of
O F
CONTROLS.
INSTRUCTIONS.
As a power supply (regu­
±l,
(-:1;:-)
(�
and
terminals
selecting the desired
II I
IN STRUCTIONS
Connect DC Voltmeter ( - hp - ModeI4 12A) to the
c.
467A OUTPUT terminals; adjust the ZERO
control to obtain a zero dc output.
d.
Connect the 467 A OUTPUT terminals to the de­
vice to which the amplified signal is being ap­
plied.
e.
Connect the Signal to be amplified to the 467 A
INPUT terminals.
3-8.
POWER SUPPLY OPERATING INSTRUCTIONS.
a.
Connect the 467 A OUTPUT terminals to the de­
vice to which the power supply is being applied.
b.
Set 467 A AMPLIFIER/SUPPLY control in the
SUPPLY range to the desired voltage settinf!;
(±l, ±2, ±4, ±10, or ±20).
c.
Set the LINE switch to ON.
±2,
To assure maximum stabilizatior.,
allow the instrument to warm up
for 30 minutes.
d.
Using the VAR. GAIN VOLTS ADJ. control, ad­
just the voltage to the desired level.
TheVAR. GAIN VOLTS ADJ. con ­
trol is used for both coarse and
fine adjustments.
is made after either limit of the
fine adjust is reached.
pOSition produces maximum negati ve
voltage.
maximum positive voltage .