Mirus for H2/H3 Graingage User's Guide
13.
Tap the check icon
EM Moisture Recalibrate
When recalibrating an EM Moisture curve, the slope of the curve is not changed. The
entire curve shifts to match the new sample. Fine-tuning a curve is often done later in
the season as additional moisture samples are available.
CAUTION: Test Weight Recalibration must be completed before following the
steps in this section.
To adjust an M2.0 moisture calibration,
1.
Open Mirus.
2.
Go to Setup > H2 or H3 GrainGage > Moisture Curves.
3.
Select the EM moisture curve you want to adjust.
4.
Tap the edit icon
5.
Tap Recalibrate.
5. Chamber Calibration
to save the new M2.0 calibration file.
.
91