Page 2
Instrument. Opening Check Please check the instrument and appendix with the packing list when you first open the packing case. If anything does not match with the packing list, please contact with the vendor or the producer. www.labstac.com...
Page 3
The company will repair the instrument based on warranty description. A.C. power’s grounding should be reliable to safeguard against an electric shock. The 3-pin plug supplied with thermo-shaker’s power cable is a safety device that should be matched with a suitable grounded socket. www.labstac.com...
Page 4
Pull the connector plug from the jack at once in the following case, and contact the vendor. • There is some liquid flowing into the instrument; • Drenched or fire burned; • Abnormal operation: such as abnormal sound or smell; • Instrument dropping or outer shell damaged; • The function has obviously changed. www.labstac.com...
Page 5
If there are smutches on the instrument, clean them with cloth. Power off when cleaning the instrument. Do not drop the clean fluid in the well when cleaning. Corrosive clean fluid is strongly prohibited. www.labstac.com...
6. Low noise and stable working. 7. Conforms to CE safety standard 8. Customized blocks are available to satisfy experimental requirement. 9. Can be used for micro test tubes, PCR plates, deep-well plates and MTPs 10. Gentle, reliable mixing with long-life brushless motor www.labstac.com...
Temperature control range 10C~100C@ Room temp.<30C Timing range 1min ~ 99h59min Accuracy of the temperature <0.5C Temperature stability @100C:± 0.5C From 25 C TO 100C ^15min (The ambient Heating time temperature is 20°C~30C) ^20min (ET^20C:20C~0C) ^20min (ET^25C:RT~4C) Cooling time <20min (ET^30C:RT~10C) www.labstac.com...
Page 9
A-BLOCK: 96x0.2 ml F-BLOCK:24x <012mm tube B-BLOCK: 54 x 0.5ml G-BLOCK:12x15ml Standard Block Falcon C-BLOCK: 35 x 1.5ml H-BLOCK: 6x50ml Falcon D-BLOCK: 35 x 2.0ml E-BLOCK: 20 x 1.5ml + 15 x 0.5ml Fuse 250V 3.0A 05x20 Dimension (mm) 300(D)x225(W)x195(H) Net weight (kg) www.labstac.com...
This chapter mainly describes the instrument’s mechanical structure, the keyboard and functions of each key, as well as preparations before power on. Please learn this chapter well before the Thermo Shaker to be operated the first time. 1. Structure Description Standard Block Keyboard Power Connector Fuse Power Switch www.labstac.com...
Page 11
Thermo Shaker Incubator SHA71-1600 2. Keyboard and Display panel Display Panel: Segment in service Current Temperature Current Speed Remnant Time S1 100.0 1200 25:05 S1 100.0 1200 30:00 Setting Speed Setting Segment Setting Temperature Setting Time www.labstac.com...
Page 12
To stop the program in operation need to keep pressing the Start/Stop start/stop key for around 2 seconds. The indicator light is flickered when heating or cooling. The indicator Heating stays light on when keeps the reaching temperature. www.labstac.com...
When program finished, it will alarm a beep five times. Speed Shaking will be stopped while temperature kept at the setting value. S1 40.0 0 00:00 S1 40.0 800 10:00 M Time S1 40.0 0 00:00 S1 40.0 800 10:30M Start/Stop Start single temperature procedure www.labstac.com...
Page 14
S1, S2, S3, S4, S5. S1 OFF 1000 01:00 b) Press of Temp. till it displays “OFF” on the LCD to shut off the thermo function. Similarly, press of Speed or Time to shut off the shaking or timing function. www.labstac.com...
Page 15
It is linearly adjusted on 10 C, 40 C and 100 C. The temperature accuracy will be within ±0.5 C after temperature calibration. 2) Both the environmental and the block temperature should be lower than C when calibration. www.labstac.com...
Page 16
Practical temperature is behind “P:” which shows 20.5, and the program auto control the temperature to 10 C. At the meantime, the sign“ ” flickers P:10.1 ADJ ceaselessly. The value behind “AdjTemP” is calibration temperature. AdjTemP= 10.0 www.labstac.com...
Page 17
Program turns to interface for operation as the right chart. P:100.0 ADJ After temperature calibration, the temperature display is the same as the practical temperature of block. AdjTemP= 98.0 Press “Start/Stop” S1 30.0 0 00:00 S1 40.0 1000 2:00 www.labstac.com...
Page 18
Take out the screws, put out the block from the main instrument. c) Take another block, steadily lay it aside on the heating board. The block installment holes should aim at the holes in main instrument. www.labstac.com...
Page 19
Refer to the right chart b). 8. Plate fixing Press one long side of the plate into the base frame. Press another side till the spring locks the plate as the arrowhead in the right chart c) indicates. Make sure the plate is fastened. www.labstac.com...
alarm of beep. Temperature sensor broken “ERR” in the environmental temperature temperature is below display with the alarm of beep. Contact to the seller No hearing or Broken sensor Contact to the seller cooling broken TE module www.labstac.com...
Page 21
Contact with the Broken TE module seller “ERR” in the Engine blocked Contact with the speed display seller continuously with alarm of beep. “ERR” in the Engine unstuck Contact with the speed display seller intermittently with alarm of beep. www.labstac.com...
Page 22
0.5mm2 blue 0.75mm2 black 0.75mm2 red Lcd Display Power 16wire Keypad interface Power 1x5wire LCD interface 0.75mm2 red 0.75mm2 black TE Driver 0.5mm2 red FAN+ motor 0.5mm2 black FAN- Fan Driver Motor interface 0.2mm2 white 0.2mm2 white Sensor PT100 www.labstac.com...