LD Didactic CASSY Instruction Sheet

Power analyser
Hide thumbs Also See for CASSY:

Advertisement

Quick Links

I
-
+
+
1
A
U
-
AC:
700 V / 16 A
nicht ohne Maske
DC: 1000 V / 25 A
in Betrieb nehmen
CAT III
300 V
I
6
+
-
+
1
B
U
-
FUSE=FF 16 A
2
690 V AC
>30 kA
I
+
-
+
1
C
U
-
AC:
do not use
700 V / 16 A
DC: 1000 V / 25 A
without mask
CAT III
300 V
I
+
-
+
1
D
U
-
12
1 Safety notes
Please read first
This device has been designed for use in LD Didactic
experiments, and is therefore only intended for use in
dry rooms. If the device is used for other measurements,
the user is responsible for the risk assessment and must
take the necessary safety precautions.
All users should be instructed about the risk of electric
shock. The user of the measuring instrument must en-
sure compliance with the guidelines of the trade associa-
tion or other insurance carriers, country-specific require-
ments, as well as regulations for schools and universi-
ties.
The Power Analyser CASSY meets the following stan-
dards:
EC/EN61010-1-2001
Safety regulations for electrical equipment for
measurement, control, and laboratory use, Part 1:
General requirements
Rated voltage range: 700 V AC and 1000 V DC
CAT III 300 (see table of permitted types of mains).
Pollution degree 2.
9
10
3
13
A
B
C
D
7
13
4
OK
13
5
13
R
250 VAC
1 A
USB
14
8
11
CASSY POWER ANALYSER+
CASSY POWER ANALYSER+
Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved
Instruction sheet 721 100
Power Analyser CASSY (727 100)
Power Analyser CASSY Plus (727 110)
0
1
Measuring channels U and I
U
2
Mask
3
Display
V
4
Rotary selector and OK button
5
Cursor buttons
6
Start button
W
7
Channel selection
8
USB-C connector
X
9
Power switch
10
LEDs
11
Relay
Y
12
Potential equalisation connection
13
Output channels (only 727 110)
727 110
14
Output, function generator output (only 727 110)
Use the analyser and accessories only as speci-
fied in the operating instructions. Failure to follow
these instructions may affect the protective func-
tions provided by the analyser.
Warning!
A "Warning" indicates circumstances and actions that
represent one or more sources of risk to the user.
Warning!
To avoid electric shock and fire hazard:
Read this entire manual carefully before using the
analyser and accessories.
Read all instructions thoroughly.
Do not work alone.
Do not use the product near explosive gases,
vapours, or in humid or wet environments.
Use the product only as intended. Otherwise, the
protection offered by the product may be affected.
Only use safety leads that are approved for the
maximum values of the Power Analyser CASSY.
If measuring tips are used, keep your fingers be-
hind the finger guards on the measuring tips.
Inspect the analyser, voltage probe heads, mea-
surement leads, and accessories for any mechan-
ical damage before use, and replace them if nec-
essary. Check for cracks in the insulation (housing,
cable sheath) or if parts are missing. Pay special
attention to the insulation on the plugs.
Check the function of the measuring device by
measuring a known voltage.
© by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com

Advertisement

Table of Contents
loading

Summary of Contents for LD Didactic CASSY

  • Page 1 Pollution degree 2. • Check the function of the measuring device by measuring a known voltage. © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 2: Technical Specifications

    Fig. 1: Sampling rate: Max. 1,000,000 values/s per channel, for U and I, max. 500,000 values/s © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 3: Assembly Description

    The device must not be operated with an open mask. menu, area menu, or value list, but cannot activate it. © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 4: Operation

    Measurements recorded can be saved or deleted via the ◦ Tare (→ 0 ←) to zero the measured value menu (see section 5). © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 5 (e.g. zoom, line fitting, integrals). These are applied to the currently selected region of the plotted curve. © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 6: Meaning Of Leds

    OK, either until both ends of the range have In order for Power Analyser CASSY to be found more easily in been defined or a cursor key has been clicked to cancel the ac- WiFi, it has a name. The standard name for this is "PA-XXX", tion.
  • Page 7: Additional Settings

    , which can be entered instead of the point in time. IP address or the CASSY name in the address field of a Lab Doc. This PIN can be modified or disabled in Mobile Config 9.3 Clock visible (see section 9.1).
  • Page 8 8 / 14 Instruction sheet 721 100 Mobile Config configures a Power Analyser CASSY by editing its hidden configuration file MOBILE.CFG. Fig. 6: In the default operating state of the Power Analyser CASSY, dent groups in the form of a *.labm file, using the menu or MobileConfig.exe is located on its SD card and can be started...
  • Page 9 VNC Viewer app: Switch picture quality from automatic to high as otherwise no connection is established. In addition, the Power Analyser CASSY Web App displays the measurement data, tables, diagrams, and evaluations of a Power Analyser Fig. 9: CASSY in up to three browsers.
  • Page 10: Firmware Updates

    Fig. 12: Attention The safety leads used must be rated for the expected voltage value. © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 11 The safety leads used must be rated for the expected voltage and current values. © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 12 The safety leads values. used must be rated for the expected voltage and current values. © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 13 Table of possible assignments. Tab. 1: © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...
  • Page 14 0 ... ±10 0 ... ±10 Table with setting ranges for the function generator. Tab. 2: © by LD DIDACTIC GmbH · Leyboldstraße 1 · D-50354 Hürth · www.ld-didactic.com Tel: +49-2233-604-0 · Fax: +49-2233-604-222 · Email: info@ld-didactic.de · Technical alterations reserved...