Set the standard source to an appropriate range;
5)
Set the output of standard source as the value given
in Table 10-3 and when the output stabilized , press
the 〔100%/START〕 key, and the LCD displays
'SAVE ', denoting that the calibrated point has
been stored;
6)
Press the key〔STEP/AUTO〕and the LCD displays
'CAL FS' , and repeat the operation of steps 5.
Voltage Measurement Calibration
1)
Press the key 〔 mA % / V 〕 to select voltage
measurement function , and the LCD displays
'INPUT '、'CAL 0'、'V ';
2)
Connect to the Instrument as shown in Figure 10-3;
3)
Set the standard source to an appropriate range;
4)
Set the output of standard source as the value given
in Table 10-3 and when the output stabilized , press
the key〔100%/START〕, and the LCD displays
'SAVE ', denoting that the calibrated point has been
stored;
NORMAL
AUX
Hi
Lo
Standard Source
(5520A)
5)
Press the key〔STEP/AUTO〕and the LCD displays
'CAL FS' , and repeat the operation of steps 4.

Section Ten Notice of the Manual

The present operation instruction is subject to
change without notice;
The content of the operation instruction is regarded
as correct. Whenever any user finds its mistakes,
omission, etc., he or she is requested to contact the
manufacturer;
The present manufacturer is not liable for any
accident and hazard arising from the customer
misuse or inadvertent operation;
The functions described in this operation instruction
should not be used as grounds to apply this product
to a particular purpose.
- 15 -
SCOPE
TRIG
OUT
Figure 10-3
https://tecpel.com

Advertisement

Table of Contents
loading

Related Products for TECPEL CL-420B

Table of Contents