1.5. Safety Precautions

The CAREvent® CA Handheld Resuscitator is designed to provide
emergency ventilatory support to patients su ering from
respiratory and/or cardiac arrest.
The CAREvent® CA Handheld Resuscitator is intended for use by
suitably trained and qualifi ed personnel. The following precautions
should always be observed:
1.
When not in use, always turn o the cylinder.
2. Never allow oil or grease to come into contact with any part of
the cylinder, regulator or resuscitator.
3. Do not disassemble any part of the resuscitator except where
described in this manual, as any unauthorized disassembly
will invalidate the warranty.
4. After use, always ensure that all components are cleaned in
accordance with the instructions provided in this manual.
5. Ensure that all components are reassembled correctly and
that all items are replaced in the carrying case.
6. After use, always ensure that a full air or oxygen cylinder
is attached before returning the unit to its normal storage
position.
7.
Ensure that a new sealing washer is used every time you attach
the regulator to the cylinder.
8. It is recommended that an alternative
the patient be available in case of gas supply failure.
2. Pre Use Function Check

2.1. Set Up

Along with the contents of the shipping cartons you will require
the following items to enable you to undertake the pre-use
functional check:
1.
Full oxygen cylinder
2. Oxygen regulator with a 50 PSI 9/16 DISS outlet. The regulator
must be able to output a minimum of 120 L/min at no less than
45 PSI (3.1 Bar).
3. Calibrated Test Lung (supplied with ventilator)
CAREvent® CA - User Manual
means of ventilating
7