4.2. Cleaning the CAREvent® CA Handheld
Resuscitator and Accessories
Routine cleaning of the equipment should be undertaken to
maintain the equipment in a clean condition.
Reusable patient valve swivel housing and diaphram can be cleaned
using a mild soap solution and disinfected using a legally marketed
commercially available disinfectant, suitable for the application.
Single use patient valves, airway pressure gauges and masks should
be discarded after each patient use and replaced with a new unit.
All other components should be wiped clean with a mild soap
solution. Under no circumstances should the complete unit be
allowed to be soaked or immersed in cleaning solutions.
A
Resuscitator Body
B
Gas Supply Inlet
Disassembly of the resuscitator (fi g 3.)
CAREvent® CA - User Manual
PROCEDURE
User Manual
Visual
Chapter 4.1
Monthly
Checking
Regulator
work
Function
pressure
Check
limiting
pressure
Service
Level I
Manual
Level II
Service
Service
Manual
CRITERIA
Device in work
order, gas tanks are
full, no missing item
Within specifi cation
Within specifi cation
Meet product
specifi cations
SCHEDULE
BY
Month
User
Every 6
User
months
Every year
User
Every
Manufacturer/
2 years
Service Center
Pressure Relief
C
System
D
Manual Button
Patient Valve
E
Swivel Housing
F
Pressure Gauge
G
Facemask
13