PRODUCT SPECIFICATIONS
Item Number
Color
Wavelength of Laser & LED
Laser Power for Classification
Net Weight
Size
Number of Lasers
Number of LED
Power Adapter
Input
Operating Environment Temperature
Operating Environment Humidity
Operating Environment Pressure
Storage & Transportation Temperature
Storage & Transportation Humidity
Electrical Safety Classification
Life Time
The time required for the iRestore Essential to
warm from the minimum storage temperature
(-13°F/-25°C) between uses until the iRestore
is ready for its intended use when the ambient
temperature is 68°F (20°C).
The time required for the iRestore Essential to
cool from the maximum storage temperature
(158°F/70°C) between uses until the iRestore
is ready for its intended use when the ambient
temperature is 68°F (20°C).
Maximum Operating Time
Applied Part under the temperature
of 86°F (30°C), LED/Laser temperature reached
16
iRESTORE ESSENTIAL, ID-500
WHITE
655NM ± 10NM
<5MW, LASER CLASS 3R
1.03LBS
10.21 X 9.13 X 6.29 IN.
51
69
MODEL: E013-1D050240VU-W WHITE (AC100-
240VAC), 50/60HZ, 0.6A DC 5V, 2.4A
100 – 240V
41°F–104°F (5°C-40°C)
15%–90% RH
700 HPA TO 1060 HPA
-13°F–158°F (-25°C-70°C)
10%–90% RH
CLASS II, TYPE BF, APPLIED PART LED & LASER
NOT SUITABLE FOR USE IN THE PRESENCE OF A
FLAMMABLE ANESTHETIC MIXTURE
IP22: ENCLOSE EQUIPMENT WITHOUT PROTECTION
AGAINST INGRESS OF WATER
CONTINUOUS OPERATION
5 YEARS
0 MINUTES
0 MINUTES
25 MINUTES
104°F (40°C)