Adjusting Image Display And Camera Capture - Hitachi HT7800 User Manual

Table of Contents

Advertisement

Adjusting image display and camera capture

To acquire images with high quality use the MAIN camera!
The camera display, white/black values and
gamma can be changed. If "Auto LUT" is checked,
the values change automatically.
A
A
• Find your preferred "capture parameter" (D).
"Maximize" is imaging the whole camera chip
dimension. Default is very useful. Set 1 will
give more contrast (dedicated to HUS). Set 2
and 3 can be set up by any user.
The camera exposure and gain can be
changed for the live mode. "R" is to reset.
• Increase the brightness in order to place the
spectra in the middle of the window (A).
• Set the standard averaging to 4-6 images and set
the capture exposure between 200-800 ms (B).
• Set the capture gain (C) to 2-6 (under "image"
window).
D
C
B
10