SCIEX Dry Pump MSR 90
OPERATOR GUIDE
Original instructions
RUO-IDV-05-13982-A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MSR 90 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SCIEX MSR 90

  • Page 1 SCIEX Dry Pump MSR 90 OPERATOR GUIDE Original instructions RUO-IDV-05-13982-A...
  • Page 2 Trademarks and/or registered trademarks mentioned herein, including associated logos, are the property of AB Sciex Pte. Ltd., or their respective owners, in the United States and/or certain other countries (see sciex.com/trademarks). For Research Use Only. Not for use in Diagnostic Procedures AB Sciex™...
  • Page 3: Table Of Contents

    Maintenance and service..........34 Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A...
  • Page 4 7.6. Service..............35 7.6.1. Return the equipment or components for service (SCIEX)....35 Fault finding.
  • Page 5 Figure 5: MSR 90 performance curve at 250 Hz....... . .
  • Page 6: Safety And Compliance

    The safety symbols on the products show the areas where care and attention is necessary. The safety symbols that we use on the product or in the product documentation have the following meanings: SCIEX Dry Pump MSR 90 Operator Guide Page 6 of 40...
  • Page 7 The equipment must be discarded carefully. Obey local and national regula- tions for disposal of this equipment. Warning - Use protective equipment Use appropriate protective equipment for the task. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 7 of 40...
  • Page 8: General Description

    The MSR 90 is a dry, multi-stage roots vacuum pump that offers high pumping speed in a compact form. The pump is connected by a power cord and is designed for use on clean duty applications.
  • Page 9: Figure 1 Pump Features

    Mains circuit breaker Mains circuit breaker Mains connector port Mains connector port Exhaust port Exhaust port Protective earth stud Protective earth stud Lifting eye positions Lifting eye positions Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 9 of 40...
  • Page 10: Control Panel

    Table: Auxiliary load currents, the output will shut down to protect the pump controller. Refer to Table: Recommended mating connector for the recommended mating plugs. SCIEX Dry Pump MSR 90 Operator Guide Page 10 of 40 RUO-IDV-05-13982-A...
  • Page 11: Logic Interface

    For logic interface data refer to Logic interface data on page 26. 2.4. Auto-run The auto‑run setting configures the pump to start automatically when the power is switched on, without any customer intervention. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 11 of 40...
  • Page 12: Pump Controller Configuration

    ▪ remotely, with the serial communications or digital and analogue process control (parallel), through the 15-way D-type interface connector. Refer to Connection for remote control and monitoring on page 21. SCIEX Dry Pump MSR 90 Operator Guide Page 12 of 40 RUO-IDV-05-13982-A...
  • Page 13: Technical Data

    Overall dimensions (L x W x H) 494 x 217 x 301 mm Maximum tilt angle for operation ± 5° Maximum tilt angle for transport 30° Nominal rotational speed 15000 rpm Outlet connection NW25 Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 13 of 40...
  • Page 14: Performance

    *The pumping speed is measured at standard ambient temperature and pressure. At higher temperatures or different ambient pressures pump performance may be compromised and the rotational speed may drop below nominal. SCIEX Dry Pump MSR 90 Operator Guide Page 14 of 40...
  • Page 15: Materials Exposed To Gases Pumped

    1 m from the pump in free space Uncertainty, K 2.5 dB(A) Product tested to ISO standard 3744:2010 Figure 5 MSR 90 performance curve at 250 Hz Pumping speed with gas ballast Pumping speed with gas ballast Power with gas ballast Power with gas ballast 3.3.
  • Page 16: Figure 6 Speed Vs Power Graph

    10 A, 250 V a.c. rms Europe 220 V 10 A, 250 V a.c. rms Japan 200 V 10 A, 250 V a.c. rms Figure 6 Speed vs Power graph SCIEX Dry Pump MSR 90 Operator Guide Page 16 of 40 RUO-IDV-05-13982-A...
  • Page 17: Installation

    Possible hazards on the dry pumping system include electricity, process chemicals, and Fomblin® (PFPE) oil: ▪ Contact us or see sciex.com/support/request-support more information for advice or assistance on installation. ▪ Do not remove the temporary covers from the system inlet and exhaust until ready to connect.
  • Page 18: Mechanical Installation

    Push-pull hazard with strain injury. Push or pull the pump system only for short distance and over flat surfaces. Lift the system if the floor is uneven or has obstacles. SCIEX Dry Pump MSR 90 Operator Guide Page 18 of 40...
  • Page 19 The pressure rating of the bellows must be higher than the highest pressure generated in the system. We recommend that you use the manufacturer's bellows. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 19 of 40...
  • Page 20: Leak Test The System

    (ground) faults. We recommend you attach a protective earth (ground) conductor (with a cross sectional area of 2.5 mm /14 AWG) to the protective earth (ground) stud. SCIEX Dry Pump MSR 90 Operator Guide Page 20 of 40 RUO-IDV-05-13982-A...
  • Page 21: Connection For Remote Control And Monitoring

    To operate the pump with parallel or serial control, the 15-way D-type connector must be used (Figure: Pump features, (8)). Refer to Figure: Logic interface connections - analogue speed control for details of the logic interface pins. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 21 of 40...
  • Page 22: Commission The Pump

    Start the pump on page 31. to shut down the system; refer to Shut down the pump on page 31. SCIEX Dry Pump MSR 90 Operator Guide Page 22 of 40 RUO-IDV-05-13982-A...
  • Page 23: Operation

    European Electromagnetic Compatibility (EMC) requirements for EMC emissions. "Group 1" is defined as equipment which does not use RF energy as an intrinsic part of operation or process. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 23 of 40...
  • Page 24: Operational Modes

    In standby mode the pump operates at a reduced speed to improve the service life of the pump. 1. Push the standby button to select standby mode. The pump will initially run at factory default standby speed (70% of full speed). SCIEX Dry Pump MSR 90 Operator Guide Page 24 of 40 RUO-IDV-05-13982-A...
  • Page 25 The indicator will blink with every short push of the standby Standby speed increase in- speed increase button. The indicator will remain ON when dicator maximum standby speed has been reached. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 25 of 40...
  • Page 26: Logic Interface Data

    OFF (4.7 k pull up + diode to 12 V d.c.) > Normal speed ON (< 0.8 V d.c. sinking 10 mA) Maximum current rating 10 mA Maximum voltage rating 28.8 V d.c FAIL status output: SCIEX Dry Pump MSR 90 Operator Guide Page 26 of 40 RUO-IDV-05-13982-A...
  • Page 27 Connect to pin 2 (0 V) to enable remote con- trol via parallel or serial control modes. Normal – status output Logic low when the pump rotational speed is at normal speed or above. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 27 of 40...
  • Page 28: Parallel Control And Monitoring

    3 = Start/ Stop control input 4 = Standby enable input 14 = Remote enable input Start switch Start switch Standby switch (optional) Standby switch (optional) Pump logic interface Pump logic interface SCIEX Dry Pump MSR 90 Operator Guide Page 28 of 40 RUO-IDV-05-13982-A...
  • Page 29: Analogue Speed Control

    9 = Analogue speed control input: 0 - 10 V 11 = +10 V Analogue reference 14 = Remote enable input Analogue control switch Analogue control switch Start switch Start switch Pump logic interface Pump logic interface Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 29 of 40...
  • Page 30: Hardware Configuration

    (approximately 50% of the default run speed). ▪ The maximum running speed provided by the analogue speed control source is clamped by the maximum standby speed setting (100% of the default run speed). SCIEX Dry Pump MSR 90 Operator Guide Page 30 of 40...
  • Page 31: Start The Pump

    ▪ use the stop button in manual control mode. ▪ use the start/stop control input in parallel control mode. ▪ use stop command in serial control mode. 4. Isolate the mains supply. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 31 of 40...
  • Page 32: Restart The Pump

    The gas ballast should continue to be applied with the pump running, for at least 1 hour after the water vapour source has been removed to enable the pump to dry. SCIEX Dry Pump MSR 90 Operator Guide Page 32 of 40...
  • Page 33: Figure 10 Gas Ballast Open-Close

    Figure 10 Gas ballast open-close Position 0 (gas ballast is closed) Position 0 (gas ballast is closed) Position 1 (gas ballast is open) Position 1 (gas ballast is open) Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 33 of 40...
  • Page 34: Maintenance And Service

    7.5. Electrical safety check Test the earth continuity and the insulation resistance of the pump system in accordance with local regulations for the periodic test of electrical equipment. SCIEX Dry Pump MSR 90 Operator Guide Page 34 of 40 RUO-IDV-05-13982-A...
  • Page 35: Service

    7.6. Service 7.6.1. Return the equipment or components for service (SCIEX) When returning the pump for service or for any other reason, you must complete the required forms provided to you for Declaration of Contamination of Vacuum Equipment and Components. The competed form tells us if any substances found in the equipment are hazardous, which is important for the safety of our employees and all other people involved in the service of your equipment.
  • Page 36: Fault Finding

    Cause There is a leak in the system. Remedy Do the leak test of the pump system in accordance with the requirements specified in Table: Performance data. SCIEX Dry Pump MSR 90 Operator Guide Page 36 of 40 RUO-IDV-05-13982-A...
  • Page 37: Alarm Indicator Codes

    If it does, con- tact us or the supplier. s-s-s-s-s-L Serial control mode in- Re-activate the serial enable and send a serial com- terlock mand to clear the error code. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 37 of 40...
  • Page 38: Storage

    4. Store the pump in a clean and dry condition until required for use. 5. When necessary, prepare and install the pump. Refer to Installation on page 17. SCIEX Dry Pump MSR 90 Operator Guide Page 38 of 40 RUO-IDV-05-13982-A...
  • Page 39: Disposal

    Take particular care with: ▪ Fluoroelastomers which may have decomposed as the result of being subjected to high temperatures. ▪ Components which have been contaminated with dangerous process substances. Operator Guide SCIEX Dry Pump MSR 90 RUO-IDV-05-13982-A Page 39 of 40...
  • Page 40 sciex.com...