Table of Contents

Advertisement

Quick Links

SCIEX Dry Pump MSR 130
OPERATOR GUIDE
Original instructions
RUO-IDV-05-14943-A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MSR 130 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SCIEX MSR 130

  • Page 1 SCIEX Dry Pump MSR 130 OPERATOR GUIDE Original instructions RUO-IDV-05-14943-A...
  • Page 2 Trademarks and/or registered trademarks mentioned herein, including associated logos, are the property of AB Sciex Pte. Ltd., or their respective owners, in the United States and/or certain other countries (see sciex.com/trademarks). For Research Use Only. Not for use in Diagnostic Procedures AB Sciex™...
  • Page 3: Table Of Contents

    6.13. Gas ballast connection........... . 32 Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A...
  • Page 4 7.5. Electrical safety check............34 7.6. Return the equipment or components for service (SCIEX)......35 Fault finding.
  • Page 5 Figure 11: Gas ballast connection..........33 Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A...
  • Page 6: Safety And Compliance

    Information about properties or instructions for an action which, if ignored, will cause damage to the equipment. We reserve the right to change the design and the stated data. The illustrations are not binding. SCIEX Dry Pump MSR 130 Operator Guide Page 6 of 40 RUO-IDV-05-14943-A...
  • Page 7: Safety Symbols

    Symbol - Protective earth Identifies an electrical equipment earth (ground) terminal. Warning - Use protective equipment Risk of injury. Use appropriate Personal Protective Equipment (PPE) when performing the task. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 7 of 40...
  • Page 8: General Description

    The MSR 130 is a dry, multi-stage roots vacuum pump that offers high pumping speed in a compact form. The pump is connected by a power cord and is designed for use on clean duty applications.
  • Page 9: Figure 1 Pump Features

    Mains circuit breaker Mains circuit breaker Mains connector port Mains connector port Protective earth stud Protective earth stud 15-way D-type connector 15-way D-type connector Auxiliary connector Auxiliary connector Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 9 of 40...
  • Page 10: Interfaces And Control Panel

    The auxiliary connector is regulated to 24 V d.c. to control the accessories. Refer to Table: Auxiliary load currents, the output will shut down to protect the pump controller. Refer to Table: Recommended mating plugs for the recommended mating plugs. SCIEX Dry Pump MSR 130 Operator Guide Page 10 of 40 RUO-IDV-05-14943-A...
  • Page 11: Logic Interface

    For logic interface data refer to Logic interface data. 2.2.3. Auto-run The auto‑run setting configures the pump to start automatically when the power is switched on, without any customer intervention. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 11 of 40...
  • Page 12: Pump Controller

    ▪ Remotely, with the serial communications or the digital and analogue process control (parallel). The control is connected to the pump through the 15‑way D‑type interface connector. Refer to Figure: Pump features Connection for remote control and monitoring. SCIEX Dry Pump MSR 130 Operator Guide Page 12 of 40 RUO-IDV-05-14943-A...
  • Page 13: Technical Data

    Data Overall dimensions (L x W x H) 696 x 297 x 343 Maximum tilt angle 25° Nominal rotational speed 9000 Mass Inlet connection NW40 Outlet connection NW25 Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 13 of 40...
  • Page 14: Performance

    Measured A‑weighted emission sound pressure level, L at ultimate 55.4 vacuum 1 m from the pump in free space dB(A) Uncertainty, K dB(A) Values from ISO 3744: 2010 Figure 5 Performance graph SCIEX Dry Pump MSR 130 Operator Guide Page 14 of 40 RUO-IDV-05-14943-A...
  • Page 15: Materials Exposed To Gases

    16.5 for maximum 3 minutes Note: In a transient overload condition the maximum overload current drawn by the pump is 10 amps. Refer to Transient overload protection for more information. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 15 of 40...
  • Page 16: Installation

    Possible hazards on the dry pumping system include electricity, process chemicals, and Fomblin® (PFPE) oil: ▪ Contact us or see sciex.com/support/request-support for advice or assistance on installation. ▪ Do not remove the temporary covers from the system inlet and exhaust until ready to connect.
  • Page 17: Mechanical Installation

    Risk of injury or damage to equipment. Make sure that cables or pipe work attached to the pump are installed carefully to prevent a slip or trip hazard and the damage to the cable. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 17 of 40...
  • Page 18 ▪ Make sure that there is no blockage to access the pump electrical supply cable or the other controls. ▪ Make sure that there is a minimum air gap of 40 mm on all sides of the pump to allow effective air circulation. SCIEX Dry Pump MSR 130 Operator Guide Page 18 of 40 RUO-IDV-05-14943-A...
  • Page 19 ▪ We recommend that you use an exhaust extraction system appropriate for use with all pumped process gases. Make sure that the exhaust extraction system is not blocked or obstructed when the pump is in operation. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 19 of 40...
  • Page 20: Leak Test The System

    (Figure: Pump features) with an appropriate cord set. Refer to Table: Recommended mating plugs. The pump must be grounded through the earth (ground) conductor of the IEC60320 connector. SCIEX Dry Pump MSR 130 Operator Guide Page 20 of 40 RUO-IDV-05-14943-A...
  • Page 21: Connect The System

    To operate the pump with parallel or serial control, use the 15-way D-type connector. Refer to Figure: Pump features. Refer to Figure: Logic interface connections - parallel control, Figure: Logic interface connections - analogue speed control for details of the logic interface pins. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 21 of 40...
  • Page 22: Commission The Pump

    Start the pump on page 30 to shut down the system, refer to Shut down the pump on page 30. SCIEX Dry Pump MSR 130 Operator Guide Page 22 of 40 RUO-IDV-05-14943-A...
  • Page 23: Operation

    "Group 1" is defined as equipment which does not use RF energy as an intrinsic part of operation or process. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 23 of 40...
  • Page 24: Operational Modes

    Use the buttons as shown in Figure: Control panel to start and stop the pump. Note: The stop command does not isolate the pump from the electrical supply. SCIEX Dry Pump MSR 130 Operator Guide Page 24 of 40 RUO-IDV-05-14943-A...
  • Page 25: Logic Interface Data

    For Turbo controller, or Turbo and Active Gauge controller (TAG), a suitable connector mating half must be used (not supplied) to connect the pump to the customer control system. Refer Table: Logic interface pins for the electrical connections. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 25 of 40...
  • Page 26 0 V reference for all control and status sig- 0 V control reference nals listed within this table. Connect to pin 2 (0 V) to start the pump Start/Stop – control input - system. SCIEX Dry Pump MSR 130 Operator Guide Page 26 of 40 RUO-IDV-05-14943-A...
  • Page 27: Parallel Control And Monitoring

    15-Way D-Type customer interface connector. Alternatively, if the NORMAL output is connected to + 10 V reference output of the control system, i.e. pin 11 of the 15-Way D- Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 27 of 40...
  • Page 28: Analogue Speed Control

    The analogue speed input is a process control source which enables the pump to run at variable operating speeds. This speed control source is an alternative to standby speed control. SCIEX Dry Pump MSR 130 Operator Guide Page 28 of 40...
  • Page 29: Figure 8 Logic Interface Connections - Analogue Speed Control

    (125 Hz to 150 Hz) Note: 0.1 V = 1% of default run speed. Voltages below 5 V will result in a clamped speed of 50% of full speed. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 29 of 40...
  • Page 30: Hardware Configuration

    Risk of injury. Do not remove the inlet connections until the pump stops rotating and the power and nitrogen supply has been isolated. The pump can take up to 3 minutes to fully stop. SCIEX Dry Pump MSR 130 Operator Guide Page 30 of 40...
  • Page 31: Restart The Pump

    The gas ballast should continue to be applied with the pump running, for at least 1 hour after the water vapor source has been removed to enable the pump to dry. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 31 of 40...
  • Page 32: Gas Ballast Connection

    Insert the pipe and check if it rests firmly on the fitting shoulder. Use a wrench to tighten the nut. One and quarter turn is required to assure an effective seal. Refer to ballast operation. SCIEX Dry Pump MSR 130 Operator Guide Page 32 of 40 RUO-IDV-05-14943-A...
  • Page 33: Figure 11 Gas Ballast Connection

    Figure 11 Gas ballast connection Retaining ring Retaining ring Plug Plug Fitting Fitting Pipe Pipe Back ferrule Back ferrule Front ferrule Front ferrule Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 33 of 40...
  • Page 34: Maintenance And Service

    7.5. Electrical safety check Test the earth continuity and the insulation resistance of the pump system in accordance with local regulations for the periodic test of electrical equipment. SCIEX Dry Pump MSR 130 Operator Guide Page 34 of 40 RUO-IDV-05-14943-A...
  • Page 35: Return The Equipment Or Components For Service (Sciex)

    Components. The competed form tells us if any substances found in the equipment are hazardous, which is important for the safety of our employees and all other people involved in the service of your equipment. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 35 of 40...
  • Page 36: Fault Finding

    Cause There is a leak in the system. Remedy Do the leak test of the pump system in accordance with the requirements specified in Table: Performance data. SCIEX Dry Pump MSR 130 Operator Guide Page 36 of 40 RUO-IDV-05-14943-A...
  • Page 37: Alarm Indicator Codes

    If it does, contact us or the supplier. s-s-s-s-s-L Serial control mode Reactivate the serial enable and send a se- interlock rial command to clear the error code. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 37 of 40...
  • Page 38: Storage

    3. Place and secure protective covers over the inlet and outlet ports. 4. Store the pump in a clean and dry condition until required for use. When necessary, prepare and install the pump. Refer to Installation on page 16. SCIEX Dry Pump MSR 130 Operator Guide Page 38 of 40 RUO-IDV-05-14943-A...
  • Page 39: Disposal

    Take particular care with: ▪ Fluoroelastomers which may have decomposed as the result of being subjected to high temperatures. ▪ Components which have been contaminated with dangerous process substances. Operator Guide SCIEX Dry Pump MSR 130 RUO-IDV-05-14943-A Page 39 of 40...
  • Page 40 sciex.com...

Table of Contents