CS – 100
CS – 200
Magnetic Sensor Calibration Guide
Magnetometer calibration is a process to correct for the environmental impact on the magnetometer.
This is important because the magnetometer is sensitive to magnetic fields, and even small changes in
the environment can affect its readings.

Calibration procedure

The calibration procedure for CTi Sensor devices, CS-100 and CS-200, is as follows:
1. Download the latest CTi-Sensor-Connect (CSC) software from the following link:
https://ctisensors.com/Software/CTiSensorConnect-0.13.0-amd64.msi
2. Install CSC.
3. Run CSC.
4. Select ASCII IMU data format and a data rate of 20 Hz.
5. Place the device and all associated parts/modules/systems in a magnetically clean area. (The
system will need to be rotated in later steps.)
6.
Select "Magnetometer Calibration" from the "Tool" menu bar.
08112023
Page 1 of 5
Magnetic Sensor
Calibration Guide
www.CTiSensors.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CS-100 and is the answer not in the manual?

Questions and answers

Summary of Contents for CTI CS-100

  • Page 1 This is important because the magnetometer is sensitive to magnetic fields, and even small changes in the environment can affect its readings. Calibration procedure The calibration procedure for CTi Sensor devices, CS-100 and CS-200, is as follows: 1. Download the latest CTi-Sensor-Connect (CSC) software from the following link: https://ctisensors.com/Software/CTiSensorConnect-0.13.0-amd64.msi 2.
  • Page 2 CS – 100 Magnetic Sensor Calibration Guide CS – 200 7. Rotate the device on the following surfaces. For each surface, follow the instructions on the CSC screen to rotate the device until the plot shows a clear circle. A. XY-surface (rotation around Z-axis) 08112023 Page 2 of 5 www.CTiSensors.com...
  • Page 3 CS – 100 Magnetic Sensor Calibration Guide CS – 200 When you first open the calibration panel, it will show a blank screen. Follow the instructions on the right to rotate your module on the XY-surface. Keep rotating your module on the XY-surface until the plot looks like the graph on the right.
  • Page 4 CS – 100 Magnetic Sensor Calibration Guide CS – 200 C. YZ-surface (rotation around X-axis) Follow the instructions on the right to rotate your module on the YZ-surface. Keep rotating your module on the YZ -surface until the plot looks like the graph on the right.
  • Page 5 Magnetometer calibration is an important process that can improve the accuracy of compass readings and the detection of nearby magnetic fields. If you are using a CTi Sensor device, it is important to calibrate the magnetometer regularly to ensure accurate readings.

This manual is also suitable for:

Cs-200

Table of Contents