  • Page 1 Installation Guide Planmeca ClinicPatrol™ for Units & Autoclaves...
  • Page 2: Table Of Contents

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Introduction ............................3 Planmeca App download ........................3 Documents and films regarding Planmeca ClinicPatrol™ ..............3 Scope of delivery ..........................4 Installations Kits ........................... 5 Planmeca Patrol™ Filter head (Part No.: 428124201) ................. 6 Overview of the back of the unit ......................
  • Page 3: Introduction

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Introduction ClinicPatrol™ is a combination of reverse osmosis and a filter cartridge for complete Desalination of the permeate. Planmeca App download Download site for Planmeca ClinicPatrol™. Scan QR code Documents and films regarding Planmeca ClinicPatrol™...
  • Page 4: Scope Of Delivery

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Scope of delivery 1 Pcs. Planmeca ClinicPatrol™ unit. 1 Pcs. Planmeca Patrol™ 5 1 Pcs. Planmeca Patrol™ 14 1 Pcs. Power cable, Power Cord SCHUKO 1 Pcs. Elbow 8mm 2 M Black Tube, 8mm 2 M Blue Tube, 8mm 1 Pcs.
  • Page 5: Installations Kits

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Installations Kits Hydrofor 8, 25, 50L 14. 1 Pcs. 15. 1 Pcs. SPX DN8 3/4F x 3/4E 1500mm 16. 5 m. Tube foodgrade 8mm 17. 1 Pcs. Power Cable with DK ground 18.
  • Page 6: Planmeca Patrol™ Filter Head (Part No.: 428124201)

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Planmeca Patrol™ Filter head (Part No.: 428124201) The filter head is assembled like shown below. If a change or repositioning of the connectors are needed it can easily be done.
  • Page 7: Overview Of The Back Of The Unit

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Overview of the back of the unit Power Plug ( A ) Drain Outlet ( B ) Water Inlet ( C ) Pressure Tank ( D ) Product Outlet ( E ) Concentrate Restrictor ( F ) BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500...
  • Page 8: Step By Step Installation Guide

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Step by step installation guide. Mount the inlet water hose (15) to the inlet on the BWT Multiblock (21) Mount the other end of the inlet water hose (15) on the Water Inlet (C).
  • Page 9 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Be Aware!! Restrictor flow. Wrong Way Right Way BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500 9 / 24...
  • Page 10 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Insert the Concentrate Restrictor (12) in the Drain Outlet (B). The Drain elbow (5) can now be mounted on the Concentrate Restrictor (12). BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500...
  • Page 11 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Cut the 8mm hose (6) in the needed length for connecting the unit to the Siphon (19). Push the 8mm hose (6) into the drain elbow (5). Push the 8mm push-in elbow (24) into the top of the Siphon (19).
  • Page 12 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Use a pressure gauge to measure the pressure in the pressure tank (14) The factory default pressure aprox. 4 bars. The tank must be deflated to 1 bar. BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500...
  • Page 13 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Use the 8 mm hose (7) and cut it in the needed length, for connecting the unit to the pressure tank. Push one end of the 8mm hose (7) into the tank inlet (D). Remember to lock it with a locking clip (13).
  • Page 14 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Mount the Straight terminal F3/8'' 8mm (9) on the Product Outlet (E). Push the 8mm hose (16) into the adaptor (9). Remember to lock it with a locking clip (13).
  • Page 15 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Mount the Reduction FM 3/4” (22) on the right side of the BWT Multiblock (21) Push the 8mm hose (16) into the reduction (22) Remember to lock it with a locking clip (13) BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500...
  • Page 16 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Plug in the power plug (4) into the units power inlet (A). Remember to set the power button in OFF position. Insert the 230V plug (4) into a wall outlet.
  • Page 17 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Now the all the connections are done and the back look like this: BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500 17 / 24...
  • Page 18: Blindcap & Membrane Installation

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves Blindcap & Membrane installation Install the Blindcap (23) (left side) as shown in the picture: Install the membrane (3) (right side) by slightly tipping the unit back, insert the filtertop in the unit and rotate the membrane.
  • Page 19: Planmeca Patrol™ Filter Head Installation

    10 Planmeca Patrol™ Filter head installation The filter for the autoclaves needs to be installed separately. After the BWT Multiblock, a tee is needed for separating the treated water into water for the units and water to the Patrol™ filter.
  • Page 20 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves (1) Push down the bypass selector on top of the filter head to rotate it and set the bypass for 0. When the 0 appears in the slot the selector knob clicks back up.
  • Page 21: Startup

    Click the power button next to the power inlet (A) on the back. The Planmeca ClinicPatrol™ will make a beep and will now make a rinse and prepare itself for operation. BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500...
  • Page 22: Planmeca Patrol™ 5 Installation

    Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves 12 Planmeca Patrol™ 5 installation Insert the filter in the filter head as shown BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500 22 / 24...
  • Page 23 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves When the filter has been mounted in the filter head it needs to be flushed. Turn the little flush knob next to the bypass selector. The filter will start flushing out of the transparent hose at the bottom of the filter head.
  • Page 24 Installation guide Planmeca ClinicPatrol™ by BWT for Units & Autoclaves The filter can be installed both standing freely or wall mounted with the wall bracket. BWT Denmark A/S, Geminivej 24, 2670 Greve, Tel.: 43 600 500 24 / 24...