Page 1
Sharky Base Size 1 to Size 2 Conversion Manual You will need the following tools and parts for this procedure: • 1x Scantling Unit Sharky size 2 6085-8081-086 • 1x Sharky shower and toilet seat unit, size 2 6082-0015-000 20230821 medifab.com...
Page 2
Remove the 3x M6 bolts holding the plate and spacer block on each side. These are the spacers and long M6 bolts that will NOT be used to reassemble/convert to the size 2 unit. Place these aside. 20230821 medifab.com solutions@medifab.com...
Page 3
Size 2 Tilt Mechanism The above image shows the original parts needed to refit to the size 1 Sharky base to convert it to a size 2, along with the new size 2 tilt mechanism and the 6x M6 x 28mm bolts.
Page 4
Now fit the original long bolts, washers, and nuts to the tilt mechanism to secure this to the main frame. Tighten the bolts and nuts to be firm, and not over tight which may impede the tilting mechanism. 20230821 medifab.com solutions@medifab.com...
Page 5
Check the tilting mechanism slides smoothly and the locking pins engage in every position. If the guide plates bind you need to add a washer between the plate and the frame. This completes the procedure, and you can now fit the size 2 seat unit to the base frame. 20230821 medifab.com solutions@medifab.com...