Elimko E-680 Series User Manual

Universal data loggers / scanners
Hide thumbs Also See for E-680 Series:

Advertisement

Manufacturer / Technical Support
Elimko Elektronik İmalat ve Kontrol Ltd. Şti.
8. Cadde 21. Sokak No:16 Emek 06510 Ankara / TÜRKİYE
Telefon: + 90 312 212 64 50 Faks: + 90 312 212 41 43
www.elimko.com.tr • e-mail:elimko@elimko.com.tr
E-680 scanners / data loggers are designed for panel mounting and should be used in an industrial environment.
The package of E-680 device contains;
Device
2 pieces of mounting clamps
User manual
Guarantee certificate
After opening the package, please check the contents with the above list. If the delivered product is wrong type, any
item is missing or there are visible defects, contact the vendor from which you purchased the product.
Before installing and operating the device, please read the user manual thoroughly.
The installation and configuration of the controller must only be performed by a person qualified in instrumentation.
Keep the unit away from flamable gases, that could cause explosion.
Do not use alcohol or other solvents to clean the device. Use a clean cloth soaked in water tightly squeezed to
gently wipe the outer surface of the device.
It is not used in medical applications.
EU DIRECTIVE COMPLIANCE
Low Voltage Directive
EN 61010-1
EMC Directive
EN 61326-1
1. Introduction....................................................................................................................................................................
1.1. Technical Specifications............................................................................................................................................
1.2. Type Coding..................................................................................................................................................................
1.3. Dimensions...................................................................................................................................................................
1.4. Panel Mounting.............................................................................................................................................................
2. Usage................................................................................................................................................................................
2.1. General.....................................................................................................................................................................
2.2. Configuration Pages...................................................................................................................................................
2.2.1. General Configuration Page............................................................................................................................
2.2.2. Alarm Configuration Page................................................................................................................................
2.2.3. Input Configuration Page.................................................................................................................................
2.2.4. Security Configuration Page................................................................................................................................
2.2.5. Calibration Page.............................................................................................................................................
2.3. Warning Messages.....................................................................................................................................................
2.4. E-680 Connection Diagrams.................................................................................................................................
2.5. E-680 Jumper Settings..........................................................................................................................................
2.6. E-680 Communication Connection.............................................................................................................................
E-680 Series
Universal Data Loggers / Scanners
CHANNEL
MEASUREMENT
32 680. 0
Program
Manual
Set1
Set2
Elimko
KY-680-1123-1
Index
1
ALARM
c#
SET2
SET1
A/M
680
User Manual
E-680
E-680
2-3
4-5
6
7
8
9
9-12
13
14-16
17-19
20-25
26
27-30
31
32-35
36-37
38

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the E-680 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Elimko E-680 Series

  • Page 1 Elimko Manufacturer / Technical Support User Manual Elimko Elektronik İmalat ve Kontrol Ltd. Şti. 8. Cadde 21. Sokak No:16 Emek 06510 Ankara / TÜRKİYE Telefon: + 90 312 212 64 50 Faks: + 90 312 212 41 43 www.elimko.com.tr • e-mail:elimko@elimko.com.tr E-680 E-680 scanners / data loggers are designed for panel mounting and should be used in an industrial environment.
  • Page 2: Technical Specification

    96×192 mm compatible with IEC/TR 60668 standards. Universal inputs and outputs of the device can be programmed easily by the user. E-680 series indicate measurements from 32 different points on instrument display and determines alarm conditions according to the result of comparison of two set points for each channel. The alarm conditions can be directed to the common alarm relays and / or to the independent relays.
  • Page 3: Type Coding

    Number of Inputs (01-32) E-680 1.3. Dimensions 3 mm 11 mm CHANNEL MEASUREMENT ALARM SET2 32 680. 0 SET1 Program Manual Set1 Set2 Elimko 93 mm 192 mm Note: Drawings are not in real scale. Do not use for scaling.
  • Page 4: Panel Mounting

    Manual Set 1: Lights during Set 1 Set1 Set2 common alarm. Elimko Set2 : Lights during Set 2 common alarm.  and  buttons : Used to change channel number during manual  button : Used to access mode, page and parameter values ...
  • Page 5 E-680 2.1 General When the device is in normal operation state, the  button can be used Normal Operation State to switch to the totalizer and the alarm pages. When the device is in Auto. / Manual manual mode and the  button is pressed, totalizer value is displayed. 01 100.
  • Page 6: General Configuration Page

    E-680 2.2.1 General Configuration Page GENERAL CONFIGURATION PAGE PAGE GCNF   Display Interval (1 - 99 sec) : Channel information display time in automatic mode.   0. 5 Scan Time (0. 2 - 9. 9 sec) : The sampling time of the channels. ...
  • Page 7: Alarm Configuration Page

    E-680 2.2.2 Alarm Configuration Page PAGE = ACNF ALARM CONFIGURATION PAGE   CHNO Channel Number (1 - KS) : Determines the channel number which will be configured.   Alarm 1 Set Point (ZERO / SPAN) A1sP   A1HY Alarm 1 Hysteresis (0 - 9999) ...
  • Page 8: Input Configuration Page

    E-680 2.2.3 Input Configuration Page INPUT CONFIGURATION PAGE PAGE ICNF   CHNO Channel Number (1 - KS) : Determines the channel number which will be configured..  Scan (ON / OFF)  SCAN ON : Channel is scanned, OFF : Channel is not scanned ...
  • Page 9 E-680 2.2.3 Input Configuration Page  0. 0 Coefficient 3 (-1999 - 9999)   Channel 4 (1 - KS )   0. 0 Coefficient 4 (-1999 - 9999)  Chno=Chno+1 (continue on next page) E-680 2.2.3 Input Configuration Page Table 2.1.
  • Page 10: Security Configuration Page

    E-680 2.2.4 Security Configuration Page PAGE = PRTC SECURITY CONFIGURATION PAGE  Display Parameter (0 - 4) : Determines the number of pages can be accessed with false  DISP security code. 0 : None of the pages. 1 : GCNF. 2 : GCNF and ACNF. 3 : GCNF, ACNF and ICNF. 4 : GCNF, A CNF, ICNF ve PRTC (SCOD excluded).
  • Page 11: Warning Messages

    E-680 2.2.5 Calibration Page Determines the lower value of 2. Analog output. Connect a ampermeter to  AO2L 2. Analog output of the device. While this parameter is selected, adjust 1200 the parameter value until the meter reading is equal to the desired lower value. ...
  • Page 12 E-680 2.4. E-680 Connection Diagram Back panel view of E-680 device is shown in Figure 2.3.. There are 3 connection blocks X, Y and Z are located on the back panel. Connection group X is used for the processor and the power supply card. Connection group Y is used for multiplexer or relay card.
  • Page 13 E-680 2.4. E-680 Connection Diagram -01+ -02+ -03+ -04+ -05+ -06+ -07+ -08+ -09+ -10+ -11+ -12+ -13+ -14+ -15+ -16+ Figure 2.8. mA Source Connection Diagram -01+ -02+ -03+ -04+ -05+ -06+ -07+ -08+ -09+ -10+ -11+ -12+ -13+ -14+ -15+ -16+...
  • Page 14 Manual More than one E-680 device can be connected to the same line and all Set1 Set2 Elimko information from devices can be collected in one center (PC, PLC). RS- ADRS=2 DEVICE-2 485 is used for communication. When a PC is used as a master, E-IB-10...

Table of Contents