Exiting From The Test Program; Setting Initial Settings For The Factory State - Nikon OT-8A Repair Manual

Auto-optester
Hide thumbs Also See for OT-8A:
Table of Contents

Advertisement

I. OT-8A TEST PROGRAM
3-6. Setting the Initial Settings for the Factory State
(1) Setting the factory state
© Open the DIP switch cover on the bottom of the operation unit. Set No. 7 on DIP
switch S1 and No. 1 on DIP switch S2 to ON.
®
Holding down the [ADD] and [PRIS] keys together, turn on the power to enter the
test program.
(D Press the [NEA1] and [FAR1] keys to turn on the LED
© Press the secret key between the [ADD] key and the [XCOF] key, then press the
[ENTE] key. The factory state initial settings are set to export specifications.
© Turn off the power, set No. 7 on DIP switch SI and No. 1 on DIP switch S2 to OFF,
then close the cover.
(2) Printing out the initial settings
Press the [PRIN] key to print out the initial settings in the operation unit CPU
board memory.
** Sample printout
dnitial settings> [Export model]
1.1 print form
1
1.2 sign of SPH -
1.3 sign of CYL -
1.4 auto SPH +
on
1.6 data select YMD 1.7 time select 24h
2.1 recos reti
1.5 2.2 prism set
on
2.3 va sel lect
0.1 2.4 X. C va I ue
.5
2.5 chart auto
on
2.6 PD select
R.L
3.1 RS232C CHI
nnk 3.2 RS232C CH2
off
3.3 entry Idata on
3.4 out RS232C
off
4. Exiting from the Test Program
The only method of exiting from the test program is to turn OFF the power switch.
1-9
Repair Manual