Table Of Contents Section Section Title Page Number Description of Contents Precautions……………………………………………………………………………………..3 1.1 Precautions for All Systems 1.2 Precautions for Ultrasonic Prophylaxis Procedure Introduction…………………………………………………………………………………..4 Application………………………………………………………………………………………4 Specification…………………………………………………………………………………...4 Contraindications and Warnings…………………………………………………………5 5.1 Contraindications 5.2 Warnings Infection Control………………………………………………………………..…………….5 6.1 General Infection Control Recommendations 6.2 Water Supply Recommendations Installation Instructions…………………………………………………………………6-8 7.1 General Information...
Section 1: Safety Precautions Prior to installation and start-up of the ultrasonic scaler, carefully read the instructions provided herein! 1.1 Precautions for All Systems Do not place the ultrasonic scaler on or next to a radiator or other heat source. Excessive heat may damage the ultrasonic scaler’s electronics.
Section 2: Introduction Rolence Enterprise Inc. is an ISO 9001(2008), ISO both supragingivally and subgingivally, leaving 13485(2003) certified manufacturer of Ultrasonic crown and root surfaces clean and smooth. Scalers. All the products are FDA registered and compliant with EN standard (CE Marking).
Section 4: Specifications ® ® Rolence ELiTEDENT MS-10B Ultrasonic Scaler requires 24V AC power supply from dental unit. 4.1 Electric Voltage: 4.2 Main Control Unit Dimension: VOLTAGE: CURRENT HEIGHT : 34 mm 24V AC 50/60Hz 4.0A(MAX) WIDTH : 127 mm WATTAGE: MAX.60W DEPTH...
regarding the use of your ultrasonic scaler. 2. To maintain sterility, the insert should remain in ® Always flush your ELiTEDENT Ultrasonic Scaler the sealed pouch until it is ready for use. ‧ with highest flow before treatment. Refer to more information in section 10.
Page 7
7.5 Installation Instruction of MS-10B 1. The main control unit should be fixed inside the dental unit where is free from water. 2. Connect handpiece water supply tube to the water control valve in the dental unit. 3. Following below pin definition for the wire connection to set up the system. 4.
M3 Screw M4 Screw Important Notice: 1. This built-in scaling system is equipped with microprocessor which could be interfered and malfunctioned by outer electromagnetic noise. Most motors of dental units produce large number of electromagnetic noise while operating, such noise could result in the scaler out of order. In order to prevent the built-in scaler from shutting down (mostly known as fail to detect and lock working frequency correctly), the handpiece should be hang back to the holder where should function a safety switch to cut off the 24V AC power supply.
Handpiece Sleeve Enable to dismantle and autoclave Handpiece tubing assembly Fastening Ring Include handpiece set, tubing Hold here, rotate handpiece O-ring and fastening ring. sleeve to release. To take off the handpiece sleeve, please hold the fastening ring, and rotate the sleeve counterclockwise from handpiece tubing assembly.
Periodically check the ultrasonic insert for wear ‧ with the Insert Efficiency Indicator. 9.3 Patient Comfort Considerations Reasons for sensitivity The use of a saliva ejector or High Volume Incorrect tip placement. Point should be directed ‧ ‧ Evacuator (HVE) is recommended during all away from root surfaces.
Ultrasonic Scaler in accordance with local and national laws. Section 13: Disclaimer ROLENCE Enterprise Inc. considers itself responsible for the effects on safety, reliability and performance of this product only if: Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons ‧...
The equipment complies with the requirements in the Medical Device Directive 93/42 EEC. EU authorized representative name and address CMC Medical Devices & Drugs S.L. C/ Horacio Lengo n18, C.P 29006, Málaga-Spain Rolence Enterprise Inc. No. 18-3, Lane 231, Pu Chung Rd., Chungli, Taoyuan 32083, Taiwan S-19-01-1211-U0001 V4.3...
Need help?