3.04 A high limit temperature cutout is provided in the event of an empty water bath. If the sensor reads a temperature 5°C higher than 3.01 Precision Baths are widely used in research the set temperature, a high temperature cutout and quality control. Their superb temperature...
To change the temperature setpoint: OPERATION To change the set temperature, continuously CAUTION: press the /\ or \/ key until the desired set EXERCISE CARE WHEN USING ACIDIC OR temperature is displayed. When continuously CAUSTIC SOLUTIONS AS THEY WILL ATTACK pressing either the /\ or \/ keys, the display THE GALVANIZED STEEL BATH BODY IF SPILLED will change slowly at first but after 1 degree...
Troubleshooting Procedures Problem Procedure 9.03 No Heat (cont.) E. If a 5V DC power supply is not available, leave J5 connected to the control board, turn power on, and repeat step D by placing the voltmeter probes on J5, Pins 2 & 3. Check Heater.
Page 13
Troubleshooting Procedures Problem Procedure Check TRIAC (High voltage output of control board) A. Make sure power is disconnected from the unit. B. Connect an AC voltmeter across the heater capable of reading 120 or 230 volts, depending on your particular unit. C.
Page 21
HEATER HEATER HARNESS ASSEMBLY POWER HARNESS DETAIL "A" TEMPERATURE SENSOR POWER LINE CORD POWER SWITCH 120 VAC WIRING DIAGRAM NOTE: JP3 & JP4 MUST BE INSTALLED FOR 120 VAC OPERATION.