4.3.7. Piping for Air motor (Air motor option only) ..............15 4.4. Flow meter and Speed log ........................16 4.4.1. Flow meter general ......................16 4.4.2. BRANNSTROM standard flow transmitter............... 16 4.4.3. Speed log general ....................... 16 4.5. Paper Printer ............................17 ...
Page 3
10.9.1. Sample probes installation arrangements ½” ..............119 10.9.2. Sample probes installation arrangements 1” ..............120 10.10. Flow meter ............................122 11. Approval Certificates ......................... 125 Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 3 of 222...
Page 4
12.9. Sample Pump, Speck pump with Gast air motor ................194 13. Appendix, Work shop test and Calibration Certificate ..............221 14. Project specific drawings and data sheet ..................222 Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 4 of 222...
1. Introduction The Oil Discharge Monitoring System, Cleantrack 1000 B, has been designed to provide means of monitoring, recording and controlling the ballast discharge in accordance with the requirements in Resolution MEPC.108(49) as amended by Resolution MEPC.240(65) and is also approved for Bio- fuel blends in accordance with MEPC.1/Circ.761 as revised.
3. Specification 3.1. Description Type CleanTrack 1000 B Application Crude and Petroleum products in accordance with MEPC.108(49) as amended by Resolution MEPC.240(65) and also approved for Bio-fuel blends in accordance with MEPC.1/Circ.761 as revised. Range 0 - 1000 ppm Accuracy According to MEPC.108(49)
Page 7
50/60 Hz Consumption 120 Watt Protection IP54 Ambient Temp. 5 - 50 °C Weight ~14 kg Tolerance MEPC.108(49) Dimensions approx. 550 x 360 x 130 mm Color RAL-7035 Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 7 of 222...
Page 8
Sample flow 900 l/h Weight ~52 kg * If sample temperature is higher than the ambient temperature limit the limit becomes lower or cabinet should be ventilated. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 8 of 222...
Page 9
Ambient Temp. -25 - +55 ºC Sample Temp. min. 0 - 65 ºC Measuring range 0-16 bar (any range 0-10 bar to 0-25 bar can be used) Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 9 of 222...
4. Installation This chapter contains advice for the installation of CleanTrack 1000 B equipment on board tankers. The advice given in this chapter is of general validity and should be supplemented with a detailed installation specification for the particular ship. Additionally all applicable regulations regarding the installation standard, issued by the relevant authorities and classification society must be followed.
Keep Cb7 and other cables connected to intrinsically safe circuits separated from non-intrinsically safe circuit cables. Data communication between Converting Unit and Analyzing Unit, Cb7: Baudrate: 19200 baud. Length: ≤50 meters. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 11 of 222...
Mount a freestanding measuring cell mounted with its hood and a connection box on the backplane. Refer to drawings/chapters: Use cable diagram above for the actual sample pump used. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 12 of 222...
The sample feed pump may be damaged if run dry for more than 10 seconds. The sample valves and sample inlet filter should be located with adequate space and accessibility for servicing. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015...
The water consumption is about 250 l/h and is open about 30 seconds at start-up. It is also recommended to flush manually at closing down. This makes about 10-15 liters per start/stop. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015...
The calculation is for a straight pipe, pressure drops at bends has not been taken in account. Make own calculations and have margin to allow the pressure regulator to work properly. Pressure(final) = 5.2 bar (pipe inner diameter, [mm]) Initial pressure, [bar] Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 15 of 222...
Refer to chapter 7.8.6. Speed page 61 for software programming. The speed log signal connected to CleanTrack 1000 B should be a pulse signal form a voltage free relay or switch. The pulse frequency should be proportional to the speed.
Nominal flow of some sample pumps: Sample pump with Ex motor (SPP-100) 600 l/h Bulk head mounted (Nikuni or Matre): 900 l/h Sample pump with air driven motor: 240 l/h Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 17 of 222...
GPS input (Cb29) and Log input (Cb12) if speed log is used. 8. Connect mains to the computer unit. After a few seconds 2 different “Brannstrom Sweden” will appear on the screen before the screen goes black again and about 2 minutes later the Cleantrack1000B software appears.
The limits for fresh water is however not essential the same way as the sample limits are. 13. When finished following the relevant steps in the 5.2. Close down procedure page 34. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 19 of 222...
The maximum tested length of this cable, Cb7, is 100m. Also check the temperature class of installed intrinsically safe apparatus and if the specified zone is acceptable. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 31 of 222...
Page 32
200 pF/m and either 1 μH/m or 30 μH/Ω where the interconnection comprises two or three cores of a conventionally constructed cable (with or without screen).” Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 32 of 222...
See chapter 7.2.2. Running Settings, page 41. 13. Press the "Start Discharge" key on the Computer Units "Operate" page and confirm start on the next popup window. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 33 of 222...
10. The measuring cell can now be removed from its docking and stored in a dark, dry and tempered place. 11. If the Sample pump is powered by air motor: Close all air supply valves. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 34 of 222...
Set LCD background light intensity. A press to the left gives a darker display and to the right a brighter display. Submenu select buttons, described below. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 35 of 222...
The key can change the displayed menu, open a submenu, open a drop down list or simply change color to yellow as an indication of activation. An example of a submenu is the verification menu that open when "Manual Override Discharge Valve" key is pressed: Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 36 of 222...
Some operations need a "Password" and when pressed opens up a "Password submenu" Give the Password and select "Admin" for administrator level or "User" for lower level. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 37 of 222...
"D1 Flashing SINGLE, D2 Flashing DOUBLE" - Waiting for KB to be pressed MC-D1 MC-D2 "D1 Flashing DOUBLE, D2 Flashing SINGLE" - Waiting for KA to be pressed MC-D1 MC-D2 Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 38 of 222...
Go to "System check" page. Chapter 7.10. System Check, page 65 "System check" Password: "2" and "Admin". Go to "Approvals" page. Chapter 7.11. Approvals page 70 "Approvals" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 39 of 222...
Pressure check and Water check is further described in chapter 7.2.4. Status page 45 below. "Water 60 sec" Manually flush with fresh water for 60 seconds. Useful for testing. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 40 of 222...
Discharge line popup menu. (This popup menu will differ depending on selections made in Chapter 7.8.1. Line of discharge page 59) Select line for discharging. "#1", "#2", "#3" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 41 of 222...
Press value indication to insert manual value. "Speed" Override ship's speed with the value inserted in "Speed" "Manual Speed" Override ship's position with the "Manual GPS" value. "Manual Position" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 44 of 222...
Only to be used in case of water check malfunction. This selection need to be verified on a popup window. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 45 of 222...
Resetting alarms. Status columns indicates "ACTIVE" for not reset alarms and "Reset" "Alarm Reset" for remaining reset alarms. Go to "Operate page". "Operate" Go to "Main Menu". "MENU" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 46 of 222...
"Scroll back! ### sec" is shown if latest printout is not shown on top line. After 300 seconds of inactivity indications will go back to displaying the top line. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 47 of 222...
Recorded data can be copied to a USB-memory stick. Use a USB-stick with a Windows FAT32 file system. (Most sticks are preformatted with FAT32) 7.5.1. USB-Memory stick Indications before a USB-Memory stick is inserted. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 48 of 222...
Page 49
If the USB-stick is removed before this button is pressed, power must be switched off and on again before the Computer Unit can recognize a USB-stick again. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 49 of 222...
7.5.2. Save Printouts on USB stick If the printer file already exist on the USB-stick an overwrite selection is displayed. Select how many printouts to save on the USB-stick. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 50 of 222...
This test should be repeated separately for each sampling point; verify that alarms function correctly when a malfunction occurs external to the monitoring system, such as no sample flow, no flow meter signal, power failure, etc.; Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 51 of 222...
Page 52
30 liters per nautical mile; Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 52 of 222...
Page 53
Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 53 of 222...
Page 54
The check should be made at a flow rate of about 50% of the rated flow of the flow meter. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015...
Selection if the Zero Calibration Result is used in the concentration calculations. "Use Zero Cal. result" Graphs Displays the values from the last successful zero. "Last Successful Zero" Displays the measured values. "Is values" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 55 of 222...
Water Valve". If "No Controlled Water Valve" is selected in the Setup/Config-page this button is not visible. Press this key to start preparations for Zero Check/Calibration. "Zero Prepare" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 56 of 222...
Indicates the result of the Check/Calibration, red = fail / green = ok. "Square indication" See explanation 2 pages up when where Status is explained. "Graph indication" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 57 of 222...
The instructions on the 2 pages below describes how the Measuring Cell is replaced and the tools needed. Before replacing the measuring cell, study the chapter 4.10. Measuring Cell Instructions and Replacement page 25. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015...
ZF2 input on Zener Barrier PCB is used. "ZF2" CVF input on Converting Unit PCB is used. "CVF" ZF1 input on Zener Barrier PCB is used but calculations are setup in ZF2-page. "ZF1->ZF2" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 59 of 222...
NO if current is linear to flow, YES if current is linear to pressure. "Square root" 7.8.3. ZF2 Same as ZF1 above. 7.8.4. CVF Same as ZF1 above. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 60 of 222...
“Overboard Valve Output” Activating both functions above. “Override and OBV Valve Outputs” Contact information can be displayed in the "Start page" and in the "Contact Information" "Alarm table". Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 62 of 222...
7.9.2. Real Time Clock Setting of the Real Time Clock. Clock shall be set to GMT-time. The clock can be set manually or to the time read from the GPS. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 63 of 222...
Normally the touch screen calibration should be in effect directly after exit from the calibration but sometimes a power cycle of the computer unit is also needed. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 64 of 222...
Press to activate. When activated, relay outputs of other printed circuit boards "Test Mode" in the system can be manually activated. Take caution when starting the sample pump, pressure alarm are disabled. Password: "8378" and "Admin". Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 65 of 222...
Item d2 on drawing: CTB10001p sheet 3, Computer Unit with open door (Printer version) page 83. Communication status with the paper printer module. "PowerCommStatus" Activate the paper rewinder. "ActivateWinder" Makes a test printout to the paper printer. "Print test line" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 66 of 222...
Relay outputs of the power supply "PowerOut" Software version, temperature and flags status of the power supply. "PowerFlags" Transmit data to the power supply. "PowerTxData" Received data from the power supply. "PowerRxData" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 67 of 222...
Software version, digital inputs and current input of the CVCT PCB. "CVCT_In" Transmit data to the CVCT PCB. "CVCT_TxData" Received data from the CVCT PCB. "CVCT_RxData" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 68 of 222...
Indicates software version, serial number and more communication status. "MC_Info" Communication status of reading calibration from with the MC. "CalRd_Com" Internal use. "GainIn" Internal use. "MC_IR" Internal use. "MC_Ain" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 69 of 222...
Note that the communication to the ZBCT must work before this communication failure can be solved. For light emitting diode indications, see chapter 8.8.3. Zener barrier PCB indications page 77 "13 System - Comm. error Computer Power PCB" Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 72 of 222...
If source is GPS, check that the NMEA communication from the GPS works. If source is Pulse log, check the connection and the programming of pulses / hour. See chapter 7.8.6. Speed, page 61. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 73 of 222...
Indicates a problem on the outlet side of the analyzing unit. Running is stopped. See chapter 7.8.5. Pressure, page 61 and 7.8.7. System Configuration Setup, page 62. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 74 of 222...
Printer buffer has overflow. Printer busy do not work, check cable connections to the printer. If this alarm remains the printer and/or the CUIO PCB connected to the printer need to be replaced. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015...
Receive. Transmit. Indicate Power and Modbus Modbus Lit if Measuring Cell running program with a communication: communication: communication is ok. flashing diode, 0.5 Hz. Transmit. Receive. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 77 of 222...
CTB10010 sheet 4 of 5, Analyzing unit with air motor sample pump page 100. 9.4. Verification of accuracy and access to restricted parts. User access to the Measuring cell is restricted with seals, which only Brannstrom Sweden AB can replace, in order to comply with clause 5.1 of MEPC. 108(49).
2 pcs cleaning brush for cleaning of glass pipe in Measuring Cell. 9.9. Consumables Lithium Battery The Computer unit CPU printed circuit board contains a Lithium battery, type CR1220 3V. See chapter 9.1. Battery replacement page 78. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 79 of 222...
Paper roll diameter: max 50 mm. Cleaning Brush Cleaning brush for CleanTrack 1000 B (cleaning of glass pipe in Measuring Cell). 9.10. Storage before installation Prior to installation the unit should be stored in a tempered and dry location protected from sunlight.
This chapter contains standard drawing alternatives and options. For project specific drawings, refer to chapter 14. Project specific drawings and data sheet page 222. The rest of the page is intentionally left blank. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 81 of 222...
10.1. Computer Unit Drawing: CTB10030 sheet 1, Computer Unit mounting Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 82 of 222...
Page 83
Drawing: CTB10001p sheet 3, Computer Unit with open door (Printer version). Note: printer is optional. Standard version is without items d2 and d3. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 83 of 222...
Page 84
Drawing: P3715090 sheet 1, Computer Unit panel mount instructions. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 84 of 222...
10.3. Analyzing Unit 10.3.1. Analyzing Unit skid with Ex. motor Sample pump Drawing: CTB10015 sheet 1 of 3, Analyzing unit skid with electrical Ex. motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 88 of 222...
Page 89
Drawing: CTB10015 sheet 2 of 3, Analyzing unit skid with electrical Ex. motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 89 of 222...
Page 90
Drawing: CTB10015 sheet 3 of 3, Analyzing unit skid with electrical Ex. motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 90 of 222...
10.3.2. Analyzing Unit with external Sample pump Drawing: CTB10014 sheet 1 of 5, Analyzing unit with external sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 91 of 222...
Page 92
Drawing: CTB10014 sheet 2 of 5, Analyzing unit with external sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 92 of 222...
Page 93
Drawing: CTB10014 sheet 3 of 5, Analyzing unit with external sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 93 of 222...
Page 94
Drawing: CTB10014 sheet 4 of 5, Analyzing unit with external sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 94 of 222...
Page 95
Drawing: CTB10014 sheet 5 of 5, Analyzing unit with external sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 95 of 222...
10.3.3. Analyzing Unit with Freestanding items for External Sample pump (To be completed with more detailed drawings) Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 96 of 222...
10.3.4. Analyzing Unit with Air motor Sample pump Drawing: CTB10010 sheet 1 of 5, Analyzing unit with air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 97 of 222...
Page 98
Drawing: CTB10010 sheet 2 of 5, Analyzing unit with air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 98 of 222...
Page 99
Drawing: CTB10010 sheet 3 of 5, Analyzing unit with air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 99 of 222...
Page 100
Drawing: CTB10010 sheet 4 of 5, Analyzing unit with air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 100 of 222...
Page 101
Drawing: CTB10010 sheet 5 of 5, Analyzing unit with air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 101 of 222...
10.7. GA-plans 10.7.1. Partnames of typical arrangement Drawing: CTB10601 sheet 1 of 1, Partnames of typical arrangement. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 108 of 222...
10.7.2. GA-plan with skid Ex. motor Sample pump Drawing: CTB10907 sheet 1 of 1, GA-plan with skid Ex. motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 109 of 222...
Page 110
Drawing: CTB10917 sheet 1 of 1, GA-plan with skid Ex. motor sample pump and flushing. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 110 of 222...
10.7.3. GA-plan with bulkhead penetrating Sample pump Drawing: CTB10906 sheet 1 of 1, GA-plan with bulkhead penetrating Sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 111 of 222...
Page 112
Drawing: CTB10916 sheet 1 of 1, GA-plan with bulkhead penetrating Sample pump and flushing. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 112 of 222...
10.7.4. GA-plan with Air motor Sample pump Drawing: CTB10901 sheet 1 of 1, GA-plan with air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 113 of 222...
Page 114
Drawing: CTB10911 sheet 1 of 1, GA-plan with air motor Sample pump and flushing. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 114 of 222...
10.7.5. GA-plan with pilot controlled Air motor Sample pump Drawing: CTB10903 sheet 1 of 1, GA-plan with pilot controlled air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 115 of 222...
10.8.3. Electrical cable diagram for air motor sample pump Drawing: CTB110204.1pn, electrical cable diagram for air motor sample pump. Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 118 of 222...
12.8. Sample Pump, SPP-100 with Elprom explosion proof motor 12.8.1. Elprom instructions Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 173 of 222...
Page 174
Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 174 of 222...
Page 175
Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 175 of 222...
Page 176
Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 176 of 222...
13. Appendix, Work shop test and Calibration Certificate The 4 following pages contains: 4 pages Work shop test: 1 page Calibration Certificate: Brannstrom Sweden AB CleanTrack 1000B - Rev 2.12, 2015 Page 221 of 222...
Need help?
Do you have a question about the CleanTrack 1000 B and is the answer not in the manual?
Questions and answers