Products not satisfying these requirements may only be used with special approval of the regulating body in the country where operated. All lines connected to the imc EOS devices should not be longer than 30 m and they should be shielded and the shielding must be grounded.
All available measurement ranges are now stated for the offset error. Amendments and bug-fix in manual edition 2.1 Chapter Amendment Precautions for reference for standard operating orientation of imc EOS added operation Amendments and bug-fix in manual edition 2 Chapter Amendment...
The user must also ensure that any personnel assisting in the use of the imc EOS have also read and understood the content of this document.
Page 12
We certify that imc EOS in all product configuration options corresponding to this documentation conforms to the directives in the accident prevention regulations in "Electric Installations and Industrial Equipment" (DGUV Regulation 3)*. This confirmation applies exclusively to devices of the imc EOS series, but not to all other components included in the scope of delivery.
The fast compact measurement instrument for direct connection of IEPE sensors and voltage signals The EOS U-4 model of the imc EOS series is a 4-channel measurement instrument for applications in the fields of noise and vibration, materials testing and component testing, ballistics and e-mobility. It is equipped with high bandwidth amplifiers to which any IEPE/ICP™...
3.2 Mechanical dimensions and drawings Front panel Fig. 2: Front with dimensions Fig. 3: Front side with items A through I These illustrations 2 to 4 show the imc EOS device in the standard operating orientation. Legend: Ground terminal PC / network socket...
Page 19
Notes on connecting the device 4.3.2 Power supply An imc EOS device is powered by a DC supply voltage which is supplied via a 2-pin LEMO socket, see Fig. 3 Front side with items: C LEMO-plug type designation FGG.1B.302.CLAD.52ZN with two coding keys (see Fig. 5) The permitted supply voltage range is 10 to 50 V DC (ultra wide range).
Page 20
In this regard, there is a distinction in how the Power-Handles is operated between the cases of imc CRONOS (CRFX) and imc EOS: While the pushbutton (not latching) main switch of a CRFX base unit switches the overall system consisting of the Handle and base unit on/off, with an imc EOS, an external switch connected to the...
Page 22
· If the Power Handle is connected to a imc EOS or a CRFX base unit please make sure that the LEMO supply socket is connected with the Power Handle! The supply of a block of modules always comes from the leftmost module (first module).
Page 23
4.3.3.2 Grounding with power supplied by a vehicle battery Fig. 7 imc EOS powered by an isolated DC-voltage source (e.g. battery) If the power supply (e.g. automobile battery) and the measurement device are at different voltage levels, connecting them via the power supply connection would cause a ground loop.
Chapter 4 Assembly and connection 4.4 Click mechanism imc EOS devices can be connected either together with others or with imc CRONOSflex (CRFX) modules by means of their clicking mechanism. The complete system is supplied with power from a shared voltage source.
10 seconds at most. It is not necessary to press the Power-On button constantly during this process. If no measurement is running, the switch-off process takes about one second. By contrast, deactivating imc EOS when using the Power-Handle (with or without additional UPS-function) is accomplished by means of the...
Chapter 8 imc EOS properties 8.2 Operating software imc EOS is integrated into the operating software imc STUDIO under Microsoft Windows, and can alternatively be controlled using an app (in preparation) under either Android or IOS from Apple. 8.2.1 imc STUDIO (PC) imc EOS, imc BUSDAQflex, imc SPARTAN, imc C-SERIES and devices belonging to the imc CRONOS series are...
Chapter 8 imc EOS properties 8.3 Starting measurement imc EOS comes with two configurable triggers, Trigger_01 and Trigger_02. The "BaseTrigger" is of the type "immediate trigger", which starts and stops the experiment and can not be configured. In the user interface it is hidden by default.
Chapter 8 imc EOS properties 8.4 imc EOS in a compound assembly imc EOS can be combined with other EOS and with devices of the CRONOSflex (CRFX) family of devices. See also Installation instructions Fig. 16 imc EOS with USV, SEN-SUPPLY-4 and NET-SWITCH-5.
For the purpose of synchronizing measurements, a SYNC socket (BNC, isolated) is provided. It is to be connected with other imc devices in order to synchronize with them. Using the software it is possible to configure one device as the IRIG-B Master. Alternatively, imc EOS can be synchronized via LAN with NTP. Notes An exact description of the manner of functioning is provided in the chapter Synchronization in the imc Software User's Manual.