Invictus
Power T Handle
®
INSTRUCTIONS FOR USE
INDICATIONS FOR USE
The Invictus Power T Handle is a single-use battery powered driver intended for use during
surgical procedures to power various orthopedic accessories or attachments. The Invictus Power
T Handle is also intended for use as a non-powered hand-held device for manual use with
orthopedic instruments.
PRINCIPLE OF OPERATION
The Invictus Power T Handle motor is powered by 2 lithium metal-oxide batteries, 4.0 volts each.
The batteries are located in the grip portion of the driver. The motor and reducing gears are
located in the body of the driver. The output shaft connects to orthopedic accessories through a
standard ¼ inch square connection. There is an automatic output shaft locking component that
allows the driver to also function as a manual T handle.
WARNINGS/CAUTIONS/PRECAUTIONS:
1. This device has been designed for single use only. Reuse, reprocessing, resterilization,
or repackaging may compromise the structural integrity and/or essential material and
design characteristics that are critical to the overall performance of the device and may
lead to device failure which may result in injury to the patient.
2. Reuse, reprocessing, resterilization, or repackaging may also create a risk of
contamination of the device and/or cause patient infection or cross infection, including,
but not limited to, the transmission of infectious disease from one patient to another.
Contamination of the device may lead to injury, illness or death of the patient or end user.
3. Device is not intended to be modified, serviced, or repaired.
4. Do not use in an oxygen-enriched atmosphere.
5. Do not incinerate the batteries as this can cause a risk to the environment.
6. Do not recharge batteries, put in backwards, or mix with used or other battery types as
these actions may cause explosion or leakage leading to personal injury or can cause a
risk to the environment.
7. After use, this product may be a potential biohazard. Handle and dispose of in
accordance with accepted medical practice and applicable local, state, and federal laws
and regulations.
8. Follow Occupation Safety and Health Administration (OSHA) standards or Universal
Precautions for blood borne pathogens.
9. In order to limit contact with infectious agents from splashing, personnel using or patient
exposed to the Invictus Power T Handle must wear personal protective equipment.
10. If device becomes objectionably warm during normal operation, discontinue use of device.
11. Soft tissue contact with rotating output shaft should be avoided to prevent soft tissue
damage.
12. Federal law restricts this device for sale by or on the order of a medical practitioner licensed
to do so.
13. If package is damaged or open, do not use product.
14. Read all instructions prior to use.
15. Do not submerge handle in liquid as this may compromise the battery and motor for the
driver.
16. Power T Handle operates at a continuous amperage which does not allow for speed control
and is not torque-limiting. Do not use Power T Handle with torque-sensitive devices.
INS-136 C
12/2023
Page 1 of 7