Connection to the analogue outputs 4.14 Handling with Parameter-Files 4.14.1 Activating a transferred parameter file 4.14.2 Downloading an active parameter file from deltawaveVER2 to USB memory stick User-defined settings General information on software use 5.1.1 Creating a new parameter file 5.1.2 Loading a new parameter file 5.1.3...
Page 3
Weighting of the paths with filled cross-sections in accordance with ISO60041 (IEC41) 8.1.1 Path arrangement and Path weights for filled round conduits The RS232 Interface 8.2.1 General 8.2.2 Data which can be transferred via RS232 EG -Declaration of Comformity References systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Technical Reference and User’s Manual 1 deltawaveVER2 Description The deltawaveVER2 ultrasonic multimeter was developed specially for flow measurement of fluids in pipes, drains and sewers with a width of 0.2m – 100m. Measurements can take place in pressurised pipes up to 100 bar and under highly variable levels.
Technical Reference and User’s Manual 1.4 Specialised applications deltawaveVER2 can be used in spaces and areas liable to contain explosive atmospheres in compliance with the relevant regulations. The system can be converted to run on batteries. ...
Input range with 100 resistance 4 mA to 20 mA Maximum resistance 250 20V DC Maximum, relative voltage to earth Maximum voltage 240V rms Power supply for external sensors + 24V DC max. 1A systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
2.3 Cable for ultrasonic transducers Double-shielded RG58 Triaxial cable Cable connections of more than 100m in length should be clarified in advance with systec Controls. Cables should be broken into pieces only after consulting your systec dealer to avoid signal interference (echoes, attenuation).
Technical Reference and User’s Manual 2.4 Certifications 2.4.1 Safety standards and EMC guidelines The deltawaveVER2 Ultrasonic Multimeter has been designed and constructed in accordance with the following guidelines and standards. Directive 2014/35 / EU Low Voltage Directive Directive 2014/30 / EU Electromagnetic compatibility...
Page 9
Transducers for use in Ex zone 2 – name plate Picture 2 Attention: The transducers must be operated in combination with a deltawaveVER2 electronic unit and are not allowed to operate with any other device. systec Controls Mess- und Regeltechnik GmbH...
The USB interface allows the transmission of parameterization data as well as the reading out of data logger data. In addition, files for updates can be copied to the deltawaveVER2 via the USB interface. You can use a (supplied) USB cable for this purpose.
Technical Reference and User’s Manual 3 Flow calculation deltawaveVER2 can be configured with up to 8 (2) acoustic paths. The configuration of a measuring point is defined by means of a parameter set. The definition of the measuring point includes in particular its geometric and hydraulic description, the assignment of inputs and outputs and the specification of calculation settings.
Page 12
Wild streams with rough rubble, wi th m ovi ng bed load 19–22 Table 1 roughness coefficients for different channel materials Sman The energy drop (slope) can be calculated from the channel gradient. Sman = h/l systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 13
Manning Strickler coefficients for dry weather flow are calculated with a curve fit, e.g. in Excel. Please ask your systec dealer for making the curve fit. The transferability of coefficients thus calculated on to the dry weather flow is very good at many measuring points.
Page 14
Wild streams with rough rubble (head-sized stones) with resting bed load 0,40 0,13 Wild streams with rough rubble, with moving bed load 0,36 0,07 Table 2 roughness coefficients for calculation models mean-section and mid-section 1. Filled pipe in “Partly-filled pipe” mode systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
With shortened inflow routes we recommend the use of crossing paths Paths which are installed in one section at the same path height are automatically recognised by deltawaveVER2 as crossing paths. The measured velocities of two crossing paths are averaged. This reduces the influence of so called crossflow effects on the accuracy Crossflows can arise with shortened inflow routes, e.g.
The location should be chosen so that the evaluation unit is not more than 100m from the measuring point, as the cables are limited in length. (systec Controls should be consulted about larger distances.) The converter cables can be extended in principle. We recommend a RG58 (Triaxial) cable as an extension.
DeltawaveVER2 can be equipped with up to two ultrasonic boards. If more than one ultrasound board is installed in deltawaveVER2, the ultrasound paths 1 to 4 are located on the leftmost board and paths 5-8 on the right board.
NPN, PNP, push-pull, active / passive DIP 2 DO PE N Two power supply options available: alternating current 90 ... 240 V / AC, direct current 18 ... 36 V / DC V+ V- systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Technical Reference and User’s Manual 4.7 Connection hints (LEAN) To access the cable compartment of the deltawaveVER2 LEAN, please solve the two screws and remove the cover plate: Remove cover from cable compartment Please always make sure to put the correct voltage to your deltawaveVER2 LEAN.
4.9 Parameterization of the pulse output (LEAN) deltawaveVER2 LEAN is equipped with universally configurable pulse outputs. This allows a broad range of user is used. The deltawaveVER2 LEAN is used to parameterize the output form via the dip- switch on the connector board.
As a guideline value (test with voltmeter), we recommend 4.7 kΩ for the pull DOWN resistor. In practice, the resistance value may deviate from the recommended value depending on the internal resistance or maximum permissible current. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
As a guideline (test with voltmeter) we recommend for the pull UP resistor 4.7 kOhm. In practice, the resistance value may deviate from the recommended value depending on the internal resistance or maximum permissible current. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Technical Reference and User’s Manual 4.11 Connecting the Ultrasonic Transducers systec Controls produces a variety of different sensors which are fitted in different ways. Please observe the separate instructions provided when connecting sensors Before the cables are connected the sensors and cables should be checked as follows.
(external 24VDC supply). The I/O board is delivered from the factory in active mode, i.e. deltawaveVER2 provides a 24VDC power supply. There are small switches on the I/O board with which the deltawaveVER2 auxiliary energy can be switched on (active) or off (passive) WARNING! IF deltawaveVER2 is operated actively, i.e.
After transferring of the parameter-file the new values will be automatically activate. 4.15.2 Downloading an active parameter file from deltawaveVER2 to USB memory stick The current hardware state (4GB SD memory) allows the recording of data in a secondary interval.
5 User-defined settings deltawaveVER2 parametrisation is performed comfortably with the DeltawaveParam software. This is supplied with deltawaveVER2 and runs with all current Windows versions from WIN7. DeltawaveParam permits the creation, loading and storing of parameter files. When a record has been created it can be stored on a memory stick as param.par and uploaded by...
The following settings must be selected under system configuration: 5.2.1 Definition of existing hardware Select which hardware is in use in your deltawaveVER2 from the four listboxes. Please indicate what is fitted in the four expansion slots. PATH means you have an ultrasound board in this expansion slot.
Technical Reference and User’s Manual 5.3 Section configuration In the section configuration, all settings for the section (measuring point) are made OPTIONS – options for all configured sections systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 31
There are reasonable default settings and you normally do not have to change them. Please change carefully or ask your systec dealer. Just for the sake of completeness the option parameters are described shortly in this manual.
Page 32
-> Parameter K2 of each measurement has to be greater than 80000 in order to be used as valid measurement. The parameter K2 is not shown on deltawaveVER2 display but is continuously stored in the data logger files. Reading out data logger files gives you an idea of the size of K2. This gives you an indication of setting value of K2 (e.g.
5.3.1 Geometric definition of your conduit (drain) To calculate the correct flow rate deltawaveVER2 has to know what your conduit looks like. The geometric form of the conduit is defined in the channel window. For the channel type, first select the kind of application.
If you selected the rectangular channel type, deltawaveVER2 connects the sampling points with straight lines. If you selected the Round over spline channel type, deltawaveVER2 draws an organic curve through the individual sampling points (see the following example).
Page 35
With the “Show Drain form“ button you can check your entries graphically. Please bear in mind that the display always shows “rectangular symmetrical channels”, the depiction of spline-connected sampling points is not possible in the parameterising software. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
We recommend using the expert menu, but only for trained service personnel. Sensor Type XU-W2 250 kHz XU-W5 500 kHz XU-W10 1 MHz / 1MHz HD Table 3 Selection of Transducers systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 37
If crossed paths are installed (see chapter 3.1.2) you have to enter equal values for path elevation in order to make sure that paths can be detected from deltawaveVER2 as crossed paths (even when real elevations are slightly different). If you do not enter equal path elevations deltawaveVER2 will not calculate with crossed paths but with two single paths.
Further path-specific settings can be made here. Several of the parameters shown here have a large influence on system stability. All expert settings are initialised with sensible default values. We recommend changing the parameters in the expert menu only after consultation with our trained service personnel. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 39
5.4.1.1 Auto Window To receive the ultrasound signals deltawaveVER2 opens the predicted signal reception time according to an internally calculated signal run time. This depends amongst other things on the mean temperature and consistence. Since these parameters can change under certain circumstances it is possible to automatically adapt the time of reception to the mean conditions.
Page 40
(gain) can cause noise which influences the measurement. Gain values higher than 200-220 indicates a principle problem with signal transfer (e.g. echo’s caused by gas bubbles, misalignment of transducers,etc…) and do not help to get better measurement. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 41
5.4.1.18 Samples before Cut Off Here the number of sample values is determined which lie before the CutOff triggering and are not “cut off”. A default value of 50 is set here. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
The error value will be set in case (input) current signal is <3.84mA or >22.8mA. Values between 3.84….22.8mA will be recognized as valid signals and will be displayed as regular level value. 5.6 Configuration of external speed sensor systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 43
If there is a second (e.g. for redundancy) flow meter (speed sensor) installed you can connect it to your deltawaveVER2 using one of the analogue inputs. The value (flow speed) of the external sensor can then be used in case of malfunction of deltawaveVER2 or in case a defined range of flow speed has exceeded.
Page 44
The signal of the external sensor (value of speed) will be used only when signal meets the defined conditions (Upper / Lower) It will not be used in case of malfunction of deltawaveVER2. Upper 0 = Do not use parameter Upper Please a value for flow speed (m/s).
4...20mA. Finally you can define a fault current which should be issued in case of error. 120 % (=23,2mA) or –10%(=2.8mA) are typical values. deltawaveVER2 is able to detect back flows. It is possible to set a negative flow value for the 4mA parameter.
You can use digital outputs to count flow. You can define how many m3 a impulse should represent by putting this value into the input field counter value for Type 2. If you put e.g. 10 into that field you get one impulse (at your output) for every 10m3. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Min-Switch Value – Unit selected Type and Max-Switch Value – Unit selected Typ.e The following Min, Max, Min-Max-Alarms are available: Flow Level Temperature Speed systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
In the Medium Definition menu area settings for the medium can be made, namely the definition of the media sonic velocity and the viscosity. The following table shows typical values for a series of media. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
All flows below this level are set to zero. This prevents very small and therefore possibly erroneous level values being included in the calculation (e.g. “dirt effects” when using pressure sensors installed on the channel floor) 5.11.4 Temperature offset systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Technical Reference and User’s Manual deltawaveVER2 calculates the temperature of the medium from the measured sound velocity. This calculated temperature can deviate from the actual temperature through calibration errors or natural transducer imprecision. This offset can be corrected by means of this input value. The calculated temperature is an additional value which is calculated and can used e.g.
In order to provide a flow value also in case of dry weather flow (no paths surcharged properly) deltawaveVER2 is able to calculate the flow depending on the current water level. For that you can define an individual discharge curve which you can get e.g. from your customer or you can use the well-known formula “Manning Strickler”...
In the ”Manning Strickler” sub menu the necessary constants for the q-h flow equation are entered. (rawness coefficient and energy drop (slope)) are to be found in the chapter 2 and Table 1 deltawaveVER2 uses the q-h equation according to Manning Strickler if the following conditions are fulfilled (see also chapter 3.1 Calculation algorithms): ...
Page 53
Sampling point L1 must always be at the lowest level, L2 and the following must then have rising level values. After entering values please always press the save button. Flow values between two sampling points will be calculated using linear interpolation. See Picture 6 for example. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
5.14.1 Min. and max. velocity deltawaveVER2 calculates the sound velocity (vs) of the medium with every run from the signal runtimes. In the minimum and maximum sound velocity input fields limit values are defined within which the calculated sound velocity must lie if the single measurement is to be regarded as valid.
5.15 Creating the parameter file After finishing your entries confirm them by clicking the Save Configuration button. A familiar window for saving files appears to save the parameters. Please note that in deltawaveVER2 only files named param.par can be loaded.
Section in measurement mode, flow calculated by manning strickler equation via level meter Section in measurement mode, flow calculated by q-h-curve via level meter Section in measurement mode, flow calculated by activated speed sensor Invalid parameter set 6.2.2 Sektion alarm systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
The header indicates the actual section flow and level and the path status: Path status Path setup incorrect Path not used in the acatual measurement cycle Path in OSZ-operation, currently not used for flow metering Path working fine Path is deactivated (e.g. not submerged) systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
The oscilloscope screen should show a signal with an acceptable signal to noise ratio (better 5:1) and a signal with a steep slope. There should be no or low visible reflections or disturbing random peaks. slope Noise Noise Signal systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
In the analog calibration screen you will be guided through the calibration process step by step. First you have to select which analog output (for this you need a calibrated current meter) or which analog input (for this you need a calibrated current source) you want to calibrate. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
In the service window the counters (totalizer) can be reset (RESET TOT), new parameters - if available - can be loaded (LOAD PARAM) the current parameters can be saved on the SD card (SAVE PARAM) and various system settings (SETUP SYST.) systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
System Reset: The system can be reset in this menu, in which case all settings and parameters are lost. Logger setup: The logger can be activated and the log interval set in this menu. Modbus setup: The settings for the (optional) Modbus are set here. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 62
PC. So that removing the SD carde does not collide with the storage activities of your deltawave2, you should deactivate the SD storage in this menu before removing the card. After an SD card is reinserted, saving will start automatically again. systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
But there are other Arrangements with more or less than 4 measurement planes possible. DeltawaveVER2 offers up to 8 path configuration, therefore the weights and positions (sketch below) of the transducers are listed in the table 12 below.
8.2 The RS232 Interface 8.2.1 General deltawaveVER2 has an RS232 interface, which output the measured data as an ASCII string. 8.2.2 Data which can be transferred via RS232 The following data can be transferred via RS232. You can transfer the following data: ...
DIN EN 61000-4-11 VDE 0847-4-11: 2005-02: Voltage dip, short-term interruption, voltage fluctuation Conformity to the Directive on the restriction of the use of certain hazardous substances in electrical and electronic equipment (ROHS) is fulfilled ROHS Directive (2011/65 / EC) systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Page 68
Technical Reference and User’s Manual The conformity of the device is ensured Manufacturer: systec Controls Mess- und Regeltechnik GmbH Lindberghstraße 4 82178 Puchheim systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
[Describes the flow measurement in filled conduits] ASME PTC 18: Hydraulic Turbines and Pump-Turbines [Describes the flow measurement in filled conduits, mainly used in UK and USA] deltawaveVER2 product brochure Installation guide for ultrasonic transducers systec Controls Mess- und Regeltechnik GmbH Rev. 1.2...
Need help?
Do you have a question about the deltawaveVER2 and is the answer not in the manual?
Questions and answers