e2v CCD42-90 Manual

Back illuminated high performance ccd sensor
Hide thumbs Also See for CCD42-90:

Advertisement

Quick Links

FEATURES
*
2048 by 4608 Pixel Format
*
13.5 mm Square Pixels
*
Image Area 27.6 x 62.2 mm
*
Back Illuminated Format for High Quantum Efficiency
*
Low Noise Output Amplifiers
*
Wide Dynamic Range
*
Symmetrical Anti-static Gate Protection
*
3-side Buttable Package
*
Gated Dump Drain on Readout Register
Flatness better than 15 mm peak to valley
*
APPLICATIONS
*
Astronomy
*
Scientific Imaging
INTRODUCTION
This version of the CCD42 family of CCD sensors has full-frame
architecture. Back illumination technology, in combination with
an extremely low noise amplifier, makes the device well suited
to the most demanding applications, such as astronomy.
The output amplifier is designed to give excellent noise levels at
low pixel rates and can match the noise performance of most
conventional scientific CCDs at pixel rates as high as 1 MHz.
The low output impedance and optional FET buffer simplify the
interface with external electronics.
The readout register has a gate controlled dump-drain to allow
fast dumping of unwanted data. The register is designed to
accommodate four image pixels of charge and a summing well
is provided capable of holding six image pixels. The output
amplifier has a feature to enable the responsivity to be reduced,
allowing the reading of such large charge packets.
The device is supplied in a package designed to facilitate the
construction of large close-butted mosaics and is designed to
be used cryogenically. The design of the package will ensure
that the device flatness is maintained at the working
temperature.
The sensor is shipped in a protective container, but no
permanent window is fitted.
e2v technologies (uk) limited, Waterhouse Lane, Chelmsford, Essex CM1 2QU, UK Telephone: +44 (0)1245 493493 Facsimile: +44 (0)1245 492492
e-mail: enquiries@e2v.com Internet: www.e2v.com
e2v technologies inc. 4 Westchester Plaza, PO Box 1482, Elmsford, NY10523-1482 USA Telephone: (914) 592-6050 Facsimile: (914) 592-5148
e-mail: enquiries@e2vtechnologies-na.com
# e2v technologies (uk) limited 2006
TYPICAL PERFORMANCE (at 173 K)
Pixel readout frequency . . . . . 20 – 3000
Output amplifier sensitivity . . . . . . . 4.5
Peak signal
Spectral range
Readout noise (at 20 kHz)
QE at 500 nm
Charge transfer efficiency
GENERAL DATA
Format
Image area
Active pixels (H)
Pixel size . . . . . . . . . .
Number of output amplifiers
Number of underscan (serial) pixels
The device has a 100% fill factor.
Package
Format . . . .
Focal plane height above base . . . . .
Package size . . . . . . . . .
Package weight . . . . . . . . . .
Number of pins . . . . . . . . . .
Inactive edge spacing:
Holding Company: e2v technologies plc
CCD42-90 Back Illuminated
High Performance CCD Sensor
. . . . . . . . . . .
. . . . . . .
. . . . . . . 3
. . . . . . . . . .
. . . . . .
. . . . . . . . .
. . . . . . . .
(V)
. . . . . . . 4608 + 4
invar metal package with PGA connector
sides . . . . . . . . . . . 260 + 50
. . . . . . . . . . . 120 + 50
top
bottom (edge connections)
kHz
mV/e
7
150
ke
/pixel
200 – 1060
nm
7
e
rms
90
99.9995
27.6 x 62.2
mm
2048
13.5 x 13.5
mm
. . . . . . 2
. . .
50
14.0
mm
28.2 x 67.3
mm
150
g approx
40
mm
mm
. . . . . . 5.0
mm
A1A-100026 Issue 8, March 2006
411/9572
7
%
%

Advertisement

Table of Contents
loading

Summary of Contents for e2v CCD42-90

  • Page 1 ... 120 + 50 bottom (edge connections) ..5.0 e2v technologies (uk) limited, Waterhouse Lane, Chelmsford, Essex CM1 2QU, UK Telephone: +44 (0)1245 493493 Facsimile: +44 (0)1245 492492 e-mail: enquiries@e2v.com Internet: www.e2v.com Holding Company: e2v technologies plc e2v technologies inc.
  • Page 2: Electrical Interface Characteristics

    5. Measured using a dual-slope integrator technique (i.e. correlated double sampling) with a 10 ms integration period with OG2 = OG1 + 1 V. 6. Readout above 3000 kHz can be achieved but performance to the parameters given cannot be guaranteed. 100026, page 2 # e2v technologies...
  • Page 3: Typical Spectral Response

    (At 790 8C, measured with astronomy broadband AR coating) 7640A 1000 1100 WAVELENGTH (nm) TYPICAL OUTPUT CIRCUIT NOISE (Measured using clamp and sample, temperature range 140 - 230 K) 7639 100k 500k FREQUENCY (Hz) # e2v technologies 100026, page 3...
  • Page 4: Clock Architecture

    Column defects A column which contains at least 100 white or black defects. CLOCK ARCHITECTURE 7896 nth line Fourth line Third line Second line First line 50 elements Serial transfer phases 50 elements 3 1 2 2 1 3 100026, page 4 # e2v technologies...
  • Page 5 - maximum range specified may be required to optimise performance. Refer to the specific device test data if possible. Maximum voltage between pairs of pins: OS to OD +15 V. Maximum current through any source or drain pin: 10 mA. The CCD is not electrically connected to the metal package. # e2v technologies 100026, page 5...
  • Page 6: Device Schematic

    DEVICE SCHEMATIC Detail of FET Buffer 7885B 7907 10 kO EXTERNAL LOAD 2048 (H) x 4612 (V) PIXELS 13.5 mm SQUARE 50 BLANK 50 BLANK ELEMENTS ELEMENTS OUTPUT CIRCUIT 7641 OUTPUT EXTERNAL LOAD LS(SS) 100026, page 6 # e2v technologies...
  • Page 7 R13, SW1 NOTES 12. Clock edges are defined at mid-amplitude points. 13. Rise and fall times should be 4 overlap times. 14. Alternate patterns may be used provided sequence and minimum overlaps are maintained. # e2v technologies 100026, page 7...
  • Page 8 DETAIL OF VERTICAL LINE TRANSFER (Multiple line dump) 7991 R13, SW1 END OF CLEAR LINE START OF 1ST LINE 2ND LINE 3RD LINE PREVIOUS LINE READOUT TRANSFER LINE READOUT REGISTER INTO READOUT REGISTER DUMP MULTIPLE LINE FROM REGISTER TO DUMP DRAIN 100026, page 8 # e2v technologies...
  • Page 9 Delay time, 1R low to R13 low 0.5T 0.8T NOTES 15. No maximum other than that necessary to achieve an acceptable readout time. 16. As set by the readout period (1 ms to 100 ms is typical). # e2v technologies 100026, page 9...
  • Page 10 The studs are available in two lengths (see dimension AD). The default unless specified is the 15.00 mm stud in the offset position. 100026, page 10 # e2v technologies...
  • Page 11: Temperature Limits

    Whilst e2v technologies has taken care to ensure the accuracy of the information contained herein it accepts no responsibility for the consequences of any use thereof and also reserves the right to change the specification of goods without notice. e2v technologies accepts no liability beyond that set out in its standard conditions of sale in respect of infringement of third party patents arising from the use of tubes or other devices in accordance with information contained herein.

Table of Contents